Der Minimal Invasive Arteria Hepatica-Katheter (MIAH). Ein Revisionsverfahren bei Defekten konventionell implantierter Therapiekatheter zur regionalen Chemotherapie bei inoperablen Lebermetastasen / The Minimally Invasive Hepatic Artery Catheter (MIAH). A Revision Procedure for Defective Conventionally Implanted Therapy Catheters in Regional Chemotherapy of Nonresectable Liver Metastases

  • U. Pohlen
  • A. Wagner
  • F. Wacker
  • G. Berger
  • H. J. Buhr
Conference paper
Part of the Deutsche Gesellschaft für Chirurgie book series (DTGESCHIR, volume 2001)

Zusammenfassung

Eine Revisionsmöglichkeit bei Defekten operativ implantierter A. hepatica-Katheter ist der MIAH-Katheter, welcher via A. subklavia in die A. hepatica eingebracht wird. Ziel dieser Arbeit war der Vergleich dreier unterschiedlicher MIAH-Katheter und derer Komplikationen. Bei 47 Patienten mit defekten Kathetersystem kamen bei einem Revisionseingriff 3 unterschiedliche MIAH-Katheter zum Einsatz. Gruppe 1 (n=15 SOS-Katheter mit Konnektion an das Portsystem mittels Ligatur und Histoacrylkleber) Poblem: 6 Patienten (40%) mit Diskonnektion der Verbindung zwischen Katheter und Portsystem. Gruppe 2 (n=15 PIPS-Katheter mit direkter Konnektion an den Port) Problem: 4 Patienten (27%) mit Katheterdislokation in die abdominelle Aorta. Gruppe 3 (n=17 SOS-Katheter mit dem Port direkt über einen compression/boot connector) keine Probleme. Der SOS-Katheter mit compression/boot connector ist eine sichere und minimal invasive Möglichkeit eine regionale Therapie durchzuführen/weiterzuführen.

Schlüsselwörter

Regionale Chemotherapie Minimal Invasiver Arteria Hepatica Katheter Lebermetastasen 

Summary

The MIAH catheter implanted into the hepatic artery via the subclavian artery offers the possibility for revision of defective surgically implanted catheters. The aim of his study was to compare 3 different MIAH catheters and their complications. Three different MIAH catheters were implanted during a revision procedure in 47 patients with defective catheter systems. Group 1: (n= 15 SOS catheters connected to the port system by ligature and histoacrylic glue). Problem: 6 patients (40%) with disconnection of the catheter and port system. Group 2: (n=5 PIPS catheters directly connected to the port system). Problem: 4 patients (27%) with dislocation of the catheter in the abdominal aorta. Group 3: (n= 17 SOS catheters directly connected to the port system via a compression/boot connector). No problems. Implantation of an SOS catheter with compression/boot connector is a safe and minimally invasive procedure for regional chemotherapy of nonresectable liver metastases.

key words

Hepatic arterial infusion Minimally invasive hepatic artery catheter Liver metastases 

Copyright information

© Springer-Verlag Berlin Heidelberg 2001

Authors and Affiliations

  • U. Pohlen
    • 1
  • A. Wagner
    • 1
  • F. Wacker
    • 1
  • G. Berger
    • 1
  • H. J. Buhr
    • 1
  1. 1.Chirurgische Klinik IUniversitätsklinikum Benjamin FranklinBerlinDeutschland

Personalised recommendations