Abstract
The aim of this chapter is to describe the commercial electrical capacitance method for measuring skin hydration, the Corneometer 825 with the new digital probe, in comparison with the older analog system. In vitro calibration and validation experiments were considered on cellulose and filter pads using different solvents in order to estimate the influence of electrolytes and the evaluation of water desorption kinetics. Measurements over different layers of low dielectric plastic foils were used in order to estimate the detection depth of the hydration measurements.
In vivo measurements were considered on various skin sites covering a large range of hydration status going from very dry to well-hydrated skin areas. The intraperson variability was lower for the digital probe (7.2 %) compared with the analog probe (8.7 %). Sensitivity of the capacitance method was very good at low and very low hydration levels while sensitivity decreases at very high levels of hydration. The influence of external factors such as pressure of application and environmental factors (temperature, relative humidity) and seasonal variation is discussed.
Keywords
- Stratum Corneum
- Skin Surface
- Capacitance Measurement
- Analog Probe
- Capacitance Method
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.
This is a preview of subscription content, access via your institution.
Buying options



References
Blank IH (1952) Factors which influence the water content of the S.C. J Invest Dermatol 18:433–440
Schwan HP (1957) Dielectric properties of living tissues. Adv Biol Med Phys V:147–163
Tagami H, Oki M, Iwatsuki K, Kanamura Y, Yamada M, Ichijo B (1980) Evaluation of skin surface hydration in vivo by electrical measurement. J Invest Dermatol 75:500–507
Tagami H, Kanamaru Y, Inoue K et al (1982) Water sorption–desorption test of the skin in vivo for functional assessment of the stratum corneum. J Invest Dermatol 78:425–428
Lévêque JL, de Rigal J (1983) Impedance methods for studying skin moisturization. J Soc Cosmet Chem 34:419–428
Dikstein S, Bercovici PG (1985) Measurement of skin surface capacitance at 16 Hz and at other frequencies. Bioeng Skin 1:357
Moseley H, English JS, Coghill GM, Mackie RM (1985) Assessment and use of a new skin hygrometer. Bioeng Skin 1:177–192
Salter DC (1987) Instrumental methods for assessing skin moisturization. Cosmet Toilet 102:103–109
Lévêque JL, Grove G, de Rigal J, Corcuff P, Kligman AM, Saint Léger D (1987) Biophysical characterization of dry facial skin. J Soc Cosmet Chem 82:171–177
Tagami H (1988) Impedance measurements. In: Lévêque JL (ed) Cutaneous investigation in health and disease. Marcel Dekker, New York, pp 79–111
Blichman CW, Serup J (1988) Assessment of skin moisture. Acta Dermatol Venereol (Stockholm) 68:284–290
Barel AO, Clarys P, Wessels B, de Romsée A (1991) Non invasive electrical measurement for evaluating the water content of the horny layer: comparison between the capacitance and the conductance measurements. In: Scott RC, Guy RH, Hadgraft J, Bodde HE (eds) Prediction of percutaneous penetration – methods, measurements, modelling. IBC Technical Services, London, pp 238–247
Van Neste D (1991) Comparative study of normal and rough human skin hydration in vivo: evaluation with four different instruments. J Dermatol Sci 2:119–124
Gabard B (1994) Testing the efficacy of moisturizers. In: Elsner P, Berardesca E, Maibach HI (eds) Bioengineering of the skin: water and the stratum corneum. CRC Press, Boca Raton, pp 147–170
Salter DC (1994) Further hardware and measurement approaches for studying water in the stratum corneum. In: Elsner P, Berardesca E, Maibach HI (eds) Bioengineering of the skin: water and the stratum corneum. CRC Press, Boca Raton, pp 205–215
Loden M (1995) Biophysical methods of providing objective documentation of the effects of moisturizing creams. Skin Res Technol 1:101–108
Tagami H (1995) Measurements of electrical conductance and impedance. In: Serup J (ed) Handbook of non-invasive methods and the skin. CRC Press, London, pp 159–164
Barel AO, Clarys P (1995) Measurement of epidermal capacitance. In: Serup J (ed) Handbook of non-invasive methods and the skin. CRC Press, London, pp 165–170
Courage W (1995) Hardware and measuring principle. In: Elsner P, Berardesca E, Maibach HI (eds) Bioengineering and the skin: water and stratum corneum. CRC Press, Boca Raton, pp 171–175
Moss J (1996) The effect of 3 moisturizers on skin surface hydration. Electrical conductance (Skicon 200), capacitance (Corneometer CM420) and transepidermal water loss (TEWL). Skin Res Technol 2:32
Piérard GE (1997) Charactérisation des peaux sèches, la biométrologie complète la clinique. Cosmétologie 14:48–51
Clarys P, Barel AO, Gabard B (1999) Non-invasive electrical measurements for the evaluation of the hydration state of the skin: comparison between three commercial instruments: the Corneometer, the Skicon and the Nova DPM. Skin Res Technol 5:14–20
Barel AO, Clarys P, Gabard B (1999) In vivo evaluation of the hydration state of the skin: measurements and methods for claim support. In: Elsner P, Merk HF, Maibach HI (eds) Cosmetics: controlled efficacy studies and regulation. Springer, Berlin, pp 57–80
Bernengo JC, de Rigal J (2000) Techniques physiques de mesure de l’hydratation du stratum corneum in vivo. In: Agache P (ed) Physiologie de la peau et explorations fonctionnelles cutanées. Editions Médicales Internationales, Cachan, pp 117–162
Lee CM, Maibach HI (2002) Bioengineering analysis of water hydration: an overview. Exog Dermatol 1:269–275
Barel AO (2002) Product testing: moisturizers. In: Elsner P, Berardesca E, Wilhelm KP, Maibach HI (eds) Bioengineering of the skin: skin biomechanics. CRC Press, Boca Raton, p 241
Tagami H (2006) Epidermal hydration: measurement of high frequency electrical conductance. In: Serup J, Jemec GB, Grove GL (eds) Handbook of non-invasive methods and the skin, 2nd edn. CRC Press, Boca Raton, pp 329–336
Clarys P, Barel AO (1995) Quantitative evaluation of skin surface lipids. Clin Dermatol 13:307–3021
Clarys P, Barel AO, Gabard B (2011) Fonction sébacée et métrologie de l’excrétion sébacée. In: Revuz J (ed) Cosmétologie et Dermatologie Esthétique. Elsevier Masson, Paris, pp 1–8
Barel AO, Clarys P (2006) Measurement of epidermal capacitance. In: Serup J, Jemec GB, Grove GL (eds) Handbook of non-invasive methods and the skin, 2nd edn. CRC Press, Boca Raton, pp 337–344
Gabard B, Clarys P, Barel AO (2006) Comparison of commercial electrical measurement instruments for assessing the hydration state of the stratum corneum. In: Serup J, Jemec GB, Grove GL (eds) Handbook of non-invasive methods and the skin, 2nd edn. CRC Press, Boca Raton, pp 351–358
Khazaka G (2008) Improvement of digital sensor technology of the Corneometer CM 825 probe. Available from http://www.courage-khazaka.de/index.php/en/all-downloads/downloads-en/file/6-brochcm825sm815e
Khazaka D (2008) Nominal pressure of application of the analogic and digital hydration measuring probe of the Corneometer CM 825 and MPA. Available from http://www.courage-khazaka.de/index.php/en/all-downloads/downloads-en/file/28-brochmpae
Khazaka G (2008) Technical information of the Courage-Khazaka calibration filter pad and the in vitro calibration of the Corneometer CM 825 (analogic and digital probes). Available from http://www.courage-khazaka.de/index.php/en/faq-en/faq-scientific-devices/60-general-questions-on-the-scientific-measurements#faqsc1
Barel AB, Clarys P (1997) In vitro calibration of the capacitance method (Corneometer CM 825)- and conductance method (Skicon 200) for the evaluation of the hydration state of the skin. Skin Res Technol 3:107–133
Clarys P, Clijsen R, Taeymans J, Barel AO (2012) Hydration measurements of the stratum corneum: comparison between the capacitance method (digital version of the Corneometer 8525) and the impedance method (Skicon 200EX). Skin Res Technol 18(3):316–323
Courage-Khazaka technical information of the Corneometer Cm 825 (2008) Available from http://www.courage-khazaka.de/index.php/en/faq-en/faq-scientific-devices/61-corneometer
Fluhr JW, Gloor M, Lazzerinin SL, Kleesz ZP, Grieshaber R, Berardesca E (1999) Comparative study of five instruments measuring stratum corneum hydration (Corneometer CM 820 and CM 825, Skicon 200, Nova DPM 9003 and Dermalab). Part I. In vitro. Part II. In vivo. Skin Res Technol 5:156–178
Bielfeldt S, Brandt M, Gerstenkorn A, Wilhelm KP (2010) Capacitance measurement of skin moisture: sophisticated calibration of instruments. Available from http://www.courage-khazaka.de/index.php/de/alle-downloads/downloads-de/file/221-litcm-multicenter
Heinrich U, Koop U, Leveneu-Duchemin MC, Osterrieder K, Bielfeldt S, Charcaut C et al (2003) Members of the DGK task Force « Skin Hydration », multicenter comparison of skin hydration in terms of physical, physiological and product dependant parameters by the capacitive method (Corneometer CM 825). J Cosmet Sci 25:31–53
André T, De Wan M, Lefèvre P, Thonnard JL (2008) Moisture Evaluator: a direct measure of fingertip skin hydration during object manipulation. Skin Res Technol 14:385–389
Clarys P, Clijsen R, Barel AO (2011) Influence of probe application on in vitro and in vivo capacitance (Corneometer CM 825) and conductance (Skicon 200 EX) measurements. Skin Res Technol 17:445–450
Wilhelm KP (1989) Possible pitfalls in hydration measurements. In: Elsner P, Barel AO, Berardesca E, Gabard B, Serup J (eds) Skin bioengineering: techniques and applications in dermatology and cosmetology. Karger, Basel, pp 223–234
Rogiers V, Derde MP, Verleye G, Roseeuw D (1990) Standardized conditions needed for skin surface hydration measurements. Cosmet Toilet 105:73–82
Berardesca E, European Group for Efficacy Measurements on Cosmetics and Other Topical Products (EEMCO) (1997) EEMCO guidance for the assessment of stratum corneum hydration: electrical methods. Skin Res Technol 3:126–132
O’ Goshi KI, Serup J (2005) Inter-instrumental variation of skin capacitance measured with the Corneometer. Skin Res Technol 11:107–109
Acknowledgments
The authors thank Mr. Gabriel Khazaka and Mrs. Diana Khazaka, Courage-Khazaka Electronic Köln, for technical information concerning the Corneometer CM 825 and MPA.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2014 Springer Berlin Heidelberg
About this chapter
Cite this chapter
Barel, A.O., Clarys, P. (2014). Skin Capacitance. In: Berardesca, E., Maibach, H., Wilhelm, KP. (eds) Non Invasive Diagnostic Techniques in Clinical Dermatology. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-32109-2_33
Download citation
DOI: https://doi.org/10.1007/978-3-642-32109-2_33
Published:
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-32108-5
Online ISBN: 978-3-642-32109-2
eBook Packages: MedicineMedicine (R0)