Accurate On-line Estimation of Delivered Dialysis Dose by Dialysis Adequacy Monitor (DIAMON)

  • Conference paper
11th Mediterranean Conference on Medical and Biomedical Engineering and Computing 2007

Part of the book series: IFMBE Proceedings ((IFMBE,volume 16))

  • 66 Accesses

Abstract

The aim of this study was to compare equilibrated urea Kt/V (eKt/V) from the slope of the logarithmic online UV-absorbance measurements measured by the Dialysis Adequacy Monitor (DIAMON), by eKt/V obtained from a new algorithm that can be implemented into the DIAMON prototype. As the reference the urea eKt/V was utilized obtained from the blood samples according to the rate adjustment method. The mean value of equilibrated Kt/V obtained with UV-absorbance (eKt/Va) was 1.06 ± 0.21, using the new algorithm (eKt/Vn) was 1.09 ± 0.18, and eKt/V from blood-urea (eKt/Vb) 1.09 ± 0.20 (N = 21). The mean values of eKt/V were not statistically different comparing different methods. However, both the systematic and the random error were diminished by the new algorithm. The systematic error was decreased from 2.19% to 0.312%, and the random error was decreased from 14.37% to 7.30% using the new algorithm. In summary, the DIAMON prototype can accurately and on-line estimate the dialysis dose.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 259.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 329.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Similar content being viewed by others

References

  1. NKF K/DOQI guidelines. CLINICAL PRACTICE GUIDELINES FOR HEMODIALYSIS ADEQUACY, UPDATE 2006. at http://www.kidney.org/professionals/KDOQI/guideline_upH D_PD_VA/index.htm.

    Google Scholar 

  2. Port F K, Ashby V B, Dhingra R K, Roys E C and Wolfe R A (2002) Dialysis dose and body mass index are strongly associated with survival in hemodialysis patients. Journal of the American Society of Nephrology 13:1061–1066

    Google Scholar 

  3. Locatelli F, Buoncristiani U, Canaud B, Khler H, Petitclerc T and Zucchelli P (2005) Haemodialysis with on-line monitoring equipment: tools or toys? Nephrology Dialysis Transplantation 20:22–33

    Article  Google Scholar 

  4. Fridolin I, Magnusson M and Lindberg L-G (2002) On-line monitoring of solutes in dialysate using absorption of ultraviolet radiation: technique description. The International Journal of Artificial Organs 25:748–761

    Google Scholar 

  5. Uhlin F, Fridolin I, Lindberg L-G and Magnusson M (2003) Estimation of delivered dialysis dose by on-line monitoring of the UV-absorbance in the spent dialysate. American Journal of Kidney Diseases 41:1026–1036

    Article  Google Scholar 

  6. Fridolin I, Uhlin F, Magnusson M and Lindberg L-G (2006) Accurate Estimation of Delivered Dialysis Dose by On-Line Ultra Violet Absorbance in the Spent Dialysate. Nephrol Dial Transplant. Vol 21:ERA/EDTA XLIII Congress (abstract), Glasgow

    Google Scholar 

  7. Daugirdas J T (1995) Simplified equations for monitoring Kt/V, PCRn, eKt/V, and ePCRn. Advances in Renal Replacement Therapy 2:295–304

    Google Scholar 

  8. Fridolin I and Uhlin F, 'Device for Dialysis Quality Parameters, Utility Model nr. EE 00620 U1. Estonia, 2006.

    Google Scholar 

  9. Fresenius, OCM. Online Clearance Monitor. Operating instructions for 4008 H7S Dialysis Machines, Fresenius Medical Care 2000.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2007 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Fridolin, I., Jerotskaja, J., Lauri, K., Scherbakov, A., Luman, M. (2007). Accurate On-line Estimation of Delivered Dialysis Dose by Dialysis Adequacy Monitor (DIAMON). In: Jarm, T., Kramar, P., Zupanic, A. (eds) 11th Mediterranean Conference on Medical and Biomedical Engineering and Computing 2007. IFMBE Proceedings, vol 16. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-540-73044-6_89

Download citation

  • DOI: https://doi.org/10.1007/978-3-540-73044-6_89

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-540-73043-9

  • Online ISBN: 978-3-540-73044-6

  • eBook Packages: EngineeringEngineering (R0)

Publish with us

Policies and ethics