Advertisement

Mechanical and Geometrical Properties Assessment of Thermoplastic Materials for Biomedical Application

  • Sandro BARONE
  • Alessandro PAOLI
  • Paolo NERI
  • Armando Viviano RAZIONALEEmail author
  • Michele GIANNESE
Chapter
Part of the Lecture Notes in Mechanical Engineering book series (LNME)

Abstract

Clear thermoplastic aligners are nowadays widely used in orthodontics for the correction of malocclusion or teeth misalignment defects. The treatment is virtually designed with a planning software that allows for a definition of a sequence of little movement steps from the initial tooth position to the final desired one. Every single step is transformed into a physical device, the aligner, by the use of a 3D printed model on which a thin foil of plastic material is thermoformed. Manufactured aligners could have inherent limitations such as dimensional instability, low strength, and poor wear resistance. These issues could be associated with material characteristics and/or with the manufacturing processes. The present work aims at the characterization of the manufactured orthodontic devices. Firstly, mechanical properties of different materials have been assessed through a set of tensile tests under different experimental conditions. The tests have the purpose of analyzing the effect that the forming process and the normal use of the aligner may have on mechanical properties of the material. The manufacturing process could also introduce unexpected limitations in the resulting aligners. This would be a critical element to control in order to establish resulting forces on teeth. Several studies show that resulting forces could be greatly influenced by the aligner thickness. A method to easily measure the actual thickness of the manufactured aligner is proposed. The analysis of a number of real cases shows as the thickness is far to be uniform and could vary strongly along the surface of the tooth.

Keywords

3D Human Modeling Virtual Design Clear Aligner Thermoforming Process Mechanical Properties Assessment Optical 3D Scanner 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Simon M., Keilig L., Schwarze J., Jung B.A., Bourauel C. Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization, Am. J. Orthod. Dentofacial Orthop., 2014,145:728-736.Google Scholar
  2. 2.
    Miller K.B., McGorray S.P., Womack R., et al. A comparison of treatment impacts between Invisalign aligner and fixed appliance therapy during the first week of treatment, Am. J. Orthod. Dentofacial Orthop., 2007,131 :302.el-302.e9.Google Scholar
  3. 3.
    Zhang N., Bai Y., Ding X., Zhang Y. Preparation and characterization of thermoplastic materials for invisible orthodontics, Dent. Mater. J., 2011,30:954–959.Google Scholar
  4. 4.
    Ma Y.S., Fang D.Y., Zhang N., Ding X.J., Zhang K.Y., Bai Y.X. Mechanical Properties of Orthodontic Thermoplastics PETG, The Chinese journal of dental research, 2016, 19 (1), pp. 43-48.Google Scholar
  5. 5.
    Hahn W., Dathe H., Fialka-Fricke J., et al. Influence of thermoplastic appliance thickness on the magnitude of force delivered to a maxillary central incisor during tipping, Am. J. Orthod. Dentofacial Orthop, 2009,136:12.e1–12.e7.Google Scholar
  6. 6.
    Martorelli M., Gerbino S., Giudice M., Ausiello P. A comparison between customized clear and removable orthodontic appliances manufactured using RP and CNC techniques, Dental Materials, 2013, 29(2), pp. 1-10Google Scholar
  7. 7.
    Hahn W., Engelke B., Jung K., et al. Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor, Angle Orthod., 2010, 80:239–246.Google Scholar
  8. 8.
    Savignano R., Viecilli R.F., Paoli A., Razionale A.V., Barone S. Nonlinear dependency of tooth movement on force system directions, American Journal of Orthodontics and Dentofacial Orthopedics, 2016, 149 (6), pp. 838-846.Google Scholar
  9. 9.
    Barone S., Paoli A., Razionale A.V., Savignano R. Design of customised orthodontic devices by digital imaging and CAD/FEM modelling. In BIOIMAGING 2016 - 3rd International Conference on Bioimaging, Proceedings, Part of 9th International Joint Conference on Biomedical Engineering Systems and Technologies, BIOSTEC 2016, pp. 44-549.Google Scholar
  10. 10.
    Duffó G. S., Castillo E. Q. Development of an artificial saliva solution for studying the corrosion behavior of dental alloys, Corrosion 60.6, 2004, 594-602.Google Scholar
  11. 11.
    Porcayo-Calderon J., et al. Corrosion Performance of Fe-Cr-Ni Alloys in Artificial Saliva and Mouthwash Solution, Bioinorganic chemistry and applications, 2015Google Scholar
  12. 12.
    ISO 527-1:2012(en) Plastics – Determination of tensile properties. International Organization for Standardization, Geneva, Switzerland, 2012Google Scholar
  13. 13.
    Ryokawa H., et al. The mechanical properties of dental thermoplastic materials in a simulated intraoral environment. Orthodontic Waves 65.2, 2006, 64-72.Google Scholar
  14. 14.
    Barone S., Paoli A., Razionale A.V. Computer-aided modelling of three-dimensional maxillofacial tissues through multi-modal imaging, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2013, 227 (2), pp. 89-104.Google Scholar
  15. 15.
    Barone S., Paoli A., Razionale, A.V. Multiple alignments of range maps by active stereo imaging and global marker framing, Optics and Lasers in Engineering, 2013, 51 (2), pp. 116-127.Google Scholar

Copyright information

© Springer International Publishing AG 2017

Authors and Affiliations

  • Sandro BARONE
    • 1
  • Alessandro PAOLI
    • 1
  • Paolo NERI
    • 1
  • Armando Viviano RAZIONALE
    • 1
    Email author
  • Michele GIANNESE
    • 1
  1. 1.DICI—Department of Civil and Industrial EngineeringUniversity of PisaPisaItaly

Personalised recommendations