Abstract
The introduction at INOR, Havana, Cuba of new treatment and verification technologies for radiotherapy has enabled the use of kilovoltage Cone Beam Computed Tomography (kV-CBCT) imaging for patient positioning. In this paper setup margin from initial clinical experience of first patients positioned with this technique is estimated. Setup margins ranging between 5 and 14 mm for different directions depending on the immobilization system and anatomical site were obtained using the simplified expression proposed by van Herk. The causes of these high values were analyzed in order to minimize these margins for further patients’ setup.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2015 Springer International Publishing Switzerland
About this paper
Cite this paper
Argota, R., Larrinaga, E., GarcĂa, F. (2015). Estimating Setup Margins using IGRT Techniques. Preliminary results in Havana. In: Jaffray, D. (eds) World Congress on Medical Physics and Biomedical Engineering, June 7-12, 2015, Toronto, Canada. IFMBE Proceedings, vol 51. Springer, Cham. https://doi.org/10.1007/978-3-319-19387-8_137