Abstract
This work compares the time series obtained by measuring simultaneously the RR time series with a Polar Sensor and the Pulse to Pulse (PP) time series by using the camera of smartphones. The measured differences between time series are modulated by the breathing rate and have little influence from the measuring device. The standard deviation of the differences between time series depends mostly on the subject under measurement and can change from values of around 10 ms order to values near 100 ms. These differences affect the quantification of HRV. On average, a breathing rate around 0.2 Hz is advised to minimize these errors.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2015 Springer International Publishing Switzerland
About this paper
Cite this paper
Mateu-Mateus, M., Guede-Fernández, F., García-González, M.A. (2015). RR Time Series Comparison Obtained by H7 Polar Sensors or by Photoplethysmography Using Smartphones: Breathing and Devices Influences. In: Lacković, I., Vasic, D. (eds) 6th European Conference of the International Federation for Medical and Biological Engineering. IFMBE Proceedings, vol 45. Springer, Cham. https://doi.org/10.1007/978-3-319-11128-5_66