Advertisement

Digit III – Palm, Digit III – Wrist

Surface Recording Technique, Orthodromic Study
  • Giuliano Gentili
  • Mario Di Napoli
Chapter

Original Settings

Sensitivity was 20 μV/division, low-frequency filter was 10 Hz, high-frequency filter was 5 kHz, sweep speed was 2 ms/division, duration of pulse was 0.1 ms (0.05 for digit III to reduce stimulus artifact), and the machine used was a Medelec Sapphire 4ME. Rate of pulse was not specified.

Position

This study was performed in the supine position, with the hands and forearms placed on an arm support with the wrist in a neutral position and the digits relaxed.

Recording

Following the orthodromic method [ 1], the sensory nerve action potentials (SNAPs) were recorded using recording electrodes mounted on a plastic bipolar bar, placed on the palm and wrist (Fig. 1). The interelectrode distance was 10 mm and the active electrode (A) was in the distal position; the reference (R) was placed proximally. Ground (G) electrode was attached to the distal region of the wrist (on the distal wrist crease).

References

  1. 1.
    Padua L, Lo Monaco M, Valente EM et al (1996) A useful electrophysiologic parameter for diagnosis of carpal tunnel syndrome. Muscle Nerve 19:48–53PubMedCrossRefGoogle Scholar
  2. 2.
    Lee KY, Lee YJ, Koh SH (2009) Usefulness of the median terminal latency ratio in the diagnosis of carpal tunnel syndrome. Clin Neurophysiol 120:765–769PubMedCrossRefGoogle Scholar
  3. 3.
    Buchthal F, Rosenfalck A (1971) Sensory conduction from digit to palm and from palm to wrist in the carpal tunnel syndrome. J Neurol Neurosurg Psychiatry 34:243–252PubMedCentralPubMedCrossRefGoogle Scholar
  4. 4.
    Melvin JL, Schuchmann JA, Lanese RR (1973) Diagnostic specificity of motor and sensory nerve conduction variables in the carpal tunnel syndrome. Arch Phys Med Rehabil 54:69–74PubMedGoogle Scholar
  5. 5.
    Buchthal F, Rosenfalck A, Trojaborg W (1974) Electrophysiological findings in entrapment of the median nerve at wrist and elbow. J Neurol Neurosurg Psychiatry 37:340–360PubMedCentralPubMedCrossRefGoogle Scholar
  6. 6.
    Kimura J (1978) A method for determining median nerve conduction velocity across the carpal tunnel. J Neurol Sci 38:1–10PubMedCrossRefGoogle Scholar
  7. 7.
    Kimura J (1979) The carpal tunnel syndrome: localization of conduction abnormalities within the distal segment of the median nerve. Brain 102:619–635PubMedCrossRefGoogle Scholar
  8. 8.
    Felsenthal G, Spindler H (1979) Palmar conduction time of median and ulnar nerves of normal subjects and patients with carpal tunnel syndrome. Am J Phys Med Rehabil 58:131–138Google Scholar
  9. 9.
    Mills KR (1985) Orthodromic sensory action potentials from palmar stimulation in the diagnosis of carpal tunnel syndrome. J Neurol Neurosurg Psychiatry 48:250–255PubMedCentralPubMedCrossRefGoogle Scholar
  10. 10.
    Bland JDP (2000) A neurophysiological grading scale for carpal tunnel syndrome. Muscle Nerve 23:1280–1283PubMedCrossRefGoogle Scholar

Copyright information

© Springer International Publishing Switzerland 2015

Authors and Affiliations

  • Giuliano Gentili
    • 1
  • Mario Di Napoli
    • 1
  1. 1.Neurological ServiceS. Camillo de’ Lellis General HospitalRietiItaly

Personalised recommendations