Inaccurate measurement of preoperative axial length is the main reason for postoperative refractive error in axial high myopia. When axial length is measured by optical biometry, the adoption of the original formula to transform the optical path length into a geometrical path length leads to overestimation of axial length in myopic eyes. The group refractive index adopted to calibrate optical biometry and immersion ultrasound biometry, in fact, does not take the different relative lengths of the ocular segments into account. Segmented axial length, which is obtained by adding the individual geometrical path length of cornea, aqueous depth, lens thickness, and vitreous chamber depth, provides more accurate axial length measurements in long eyes.
Axial length adjustment methods (Wang-Koch adjustment) have been developed and refined to improve the refractive outcome of traditional formulas (Holladay 1, Holladay 2 and SRK/T). Similar adjustments are included in newer formulas and lead to better results compared to the original axial length measurements.
You have full access to this open access chapter, Download chapter PDF
Keywords
- Axial length
- Myopia
- Optical biometry
- Immersion ultrasound biometry
- Segmented axial length
- Sum-of-segments axial length
- Group refractive index
In long axial length (AL) eyes, traditional intraocular lens (IOL) power formulas tend to select IOLs of insufficient power, leaving patients with postoperative hyperopia. To reduce the chances for hyperopic surprises, surgeons used to empirically aim for a more myopic postoperative outcome by targeting a postoperative refraction of −1.00 to −2.00 diopter (D). Norrby [1] reported that the largest contributor of error in IOL power calculation was the estimation of effective lens position (ELP) (35%), followed by the postoperative refraction determination (27%) and AL measurement (17%). In long eyes, the required IOL powers are low, and errors in ELP produce low refractive effect. This indicates that accuracy of ELP estimation in long eyes is not as important as in normal and short eyes in which higher IOL powers are required. In this chapter, we will discuss the factors contributing to challenges in IOL power prediction in long eyes, the formulas appropriate for use in these eyes, and the refractive outcomes with these formulas.
Factors Contributing to Challenges in IOL Power Calculation in Long Eyes
Inaccurate measurement of preoperative AL has been reported to be the main reason for postoperative refractive error in axial high myopia [2]. There are three factors that primarily contribute to challenges in IOL power calculations in long eyes.
Posterior Staphyloma
The incidence of posterior staphyloma increases with increasing AL, and it is likely that nearly all eyes with pathologic myopia have some form of posterior staphyloma. Ultrasonic biometric methods can produce errors in the presence of a posterior staphyloma by giving falsely long AL.
An immersion A/B-scan approach for AL measurement has been described in the setting of posterior staphyloma [3]. Using a horizontal axial B-scan, an immersion echogram through the posterior fundus is obtained with the cornea and lens echoes centered while simultaneously displaying the optic nerve void. The A-scan vector is then adjusted to pass through the middle of the cornea as well as the anterior and posterior lens echoes to ensure that the vector will intersect the retina in the region of the fovea. Optical biometry with appropriate patient fixation may solve this problem of identifying the fovea and has improved outcomes in long eyes with a posterior staphyloma.
Calculation of Axial Length Matching US Data: The IOL Master Calibration
Theoretically, optical biometry permits more accurate measurements when a posterior staphyloma is present. However, in a study investigating the accuracy of SRK/T formula in eyes with negative and zero-powered IOLs, MacLaren and colleagues [4] reported consistent hyperopic errors across all three methods of biometry (A-scan, B-scan, and optical). This indicates that eliminating or minimizing the adverse impact of posterior staphylomata on IOL calculations does not prevent hyperopic surprises in long eyes.
During the development of the first optical biometer (Carl Zeiss Meditec, Jena, Germany), first, the OPL data were transformed to geometrical path length (GPL) data using a group refractive index calculated theoretically from the Gullstand eye model (n = 1.3549). Then regression analysis produced the definitive conversion formula programmed in the commercial version of the first IOL Master model: GPL (OPL/1.3549) = ALGBS × 0.9571 + 1.3033, where ALGBS is the AXL measured by immersion US. There were two main reasons for this transformation: (1) to avoid the bias produced in the extremes of the AXL range that would have occurred if only an average refractive index was used for the calculation from measured OPL and (2) to adjust the retinal plane reference from PCI to the internal limiting membrane for US. In the PCI instrument, the main retinal signal is produced in the retinal pigment epithelium, which produces systematically longer measurements.
With the IOL Master calibration, the same IOL constants could be used when surgeons moved from immersion US to optical biometry, which, of course, made this transition easier. All optical biometers developed thereafter, except the Argos biometer (Movu, Komaki, Japan), are all calibrated to provide an AL equivalent or similar to the first optical biometer. However, in long eyes, the relative lengths of the ocular segments may differ from those in eyes with normal ALs, and the use of a fixed group refractive index for the entire eye may yield incorrect values for AL.
We proposed a segmented AL that is calculated by summing the GPL of individual ocular segments converted from their respective OPLs using specific refractive indices for each ocular medium: cornea, aqueous depth (AD), lens thickness (LT), and vitreous chamber depth [6]. Theoretically, the segmented AL may provide more accurate AL measurements in eyes with unusual ocular segment proportions. We found that the segmented ALs were shorter in long eyes compared with the AL calculated with the IOL Master calibration in an OLCR instrument. The refractive accuracy with segmented ALs was improved in long eyes with the Barrett, Haigis, Hoffer Q, Holladay 1, and SRK/T formulas [ 6]. Cooke and Cooke [7] compared prediction accuracy with the AL calculation method of the Lenstar biometer (transitional AL) and that of the Argos biometer (sum-of-segments AL). They found that using sum-of-segments AL, instead of traditional AL, improved predictions for formulas designed on ultrasound data (SRK/T, Holladay 1, Holladay 2, Hoffer Q, and Haigis), although it worsened the Barrett and Olsen formulas. Further studies are desirable in this regard.
Extrapolation Issue in Extreme Long Eyes
The dataset used in the study by Haigis et al that developed regression formula by converting the OPL data to GPL in millimeters, eyes with AL up to 27.45 mm were included [5]. When this conversion method is used in eyes longer than 27.45 mm, extrapolation is introduced and errors may presumably occur.
Principal Plane Shift in Negative-Power IOLs
There are differences in geometries of positive-diopter IOLs and negative-diopter IOLs. The optic principle plane shifts in negative-power IOLs, compared to the principle plane in positive-power IOLs. Petermeier and colleagues [8] proposed seperate constants optimization for eyes with negative IOL powers.
IOL Power Calculation Formulas for Long Eyes
Axial Length Adjustment Methods (Wang-Koch Adjustment)
We assume that the hyperopic error seen in long eyes is in the measurement of AL or in the way that formulas use this value. In a previous study, we proposed a method of optimizing AL in long eyes (Wang-Koch adjustment) [9]. Our results showed that this method significantly improved the accuracy of IOL power calculation in eyes with IOL powers ≤5 D, and significantly reduced the percentage of eyes that would be left hyperopic.
In a more recent study [10], we modified the original AL adjustment formulas by using ULIB (User Group for Laser Interference Biometry) lens constants and manifested refraction converted to 6 meters. The modified AL adjustment formulas are less aggressive (less myopic outcomes) than the original AL adjustment formulas. AL adjustment is required in eyes with an AL > 26.5 mm for the modified Holladay 1 formula and AL > 27.0 mm for the modified SRK/T formula. The modified equations for optimizing the AL are as follows:
-
Modified Holladay 1 optimized AL = 0.817 × (measured AL) + 4.7013.
-
Modified SRK/T optimized AL = 0.8453 × (measured AL) + 4.0773.
Based on the formula, the optimized AL is calculated from the measured optical or ultrasonic AL. Then, the optimized AL is entered into the IOLMaster or Lenstar, and the calculation is performed again. We recommend selecting the IOL power that predicts a minus prediction error close to zero (−0.1 to −0.2 D), since slight myopic results may occur with this approach of optimizing AL. Figure 63.1 shows that an 8.0 D SN6ATT was suggested using the Holladay 1 with original AL of 28.41 mm. Recalculation with the optimized AL of 27.81 mm produced a 9.5 D IOL with predicted refraction of −0.06 D. A 9.5 D SN6AT3 was implanted and, at 3 weeks postoperatively, the uncorrected visual acuity was 20/15 and the manifest refraction was plano.
We also developed an AL adjustment equation for Holladay 2 formula [11]. The polynomial optimization equation is as follows:
-
Holladay 2 optimized AL = 0.0001154786 × (measured AL)3 + 0.0032939472 × (measured AL)2 + 1.001040305 × (measured AL) − 0.3270056564.
With the Holladay IOL Consultant Software, users have the option to select the AL adjustment method, and IOL power calculations will be performed automatically using the optimized AL in long eyes.
It should be noted that the AL adjustment method should be used with combination of the Holladay 1, Holladay 2, and SRK/T formulas. The newer IOL power calculation formulas already have the AL optimized or adjusted empirically by their authors and the AL adjustment method should not be used.
Super Formula
This formula is a combination of the Hoffer Q, Holladay 1, Holladay 2, and SRK/T formulas and also has a small component of artificial intelligence [12]. In long eyes, the Wang-Koch AL adjustment is used. In 2019, the formula was revised using the postoperative data as component of artificial intelligence. It is available at www.iolcalc.com.
Barrett Universal II Formula
The Barrett Universal II (BUII) formula is the evolution of the Barrett Universal I, which was published in 1987 as a thick-lens paraxial formula [13]. It uses AL, keratometry, anterior chamber depth (ACD), LT, and corneal diameter (CD) values. The detailed prediction approach for effective lens position (ELP) is not published. This formula has been refined to improve outcomes in long eyes.
Hill-RBF Formula
The Hill-RBF (Radial Basis Function) calculator is an artificial intelligence-based, self-validating method for IOL power selection employing pattern recognition and a sophisticated form of data interpolation [14]. Based on artificial intelligence, this methodology is entirely data-driven. This approach also employs a validating boundary model, indicating to the user when it is performing within a defined area of accuracy. The Hill-RBF 3.0 was recently released based on significantly expanded datasets for short and long eyes. Additionally, it has increased the number of parameters used for IOL power selection by adding central corneal thickness (CCT), LT, CD, and gender to the existing parameters of AL, keratometry, ACD, and the desired postoperative spherical equivalent refraction.
Olsen Formula
With the Olsen formula, IOL power is calculated based on exact ray tracing (Snell’s law of refraction) and paraxial ray tracing (Gaussian Optics). This formula incorporates the latest generation ACD prediction algorithms based on the complex relationship between the preoperative ocular dimensions (in particular ACD and LT) and the postoperative position of the IOL (the postoperative ACD) [15]. Measurements of the anterior and posterior corneal curvatures as well as conic coefficients (Q-values) obtained by modern anterior segment imaging systems can be used directly by the PhacoOptics program developed by Thomas Olsen (www.phacooptics.net).
Kane Formula
The Kane formula was developed by Jack X. Kane. It uses theoretical optics with artificial intelligence and regression-based components to refine the predictions (www.iolformula.com). It utilizes K, AL, ACD, and gender to predict the IOL position, with LT and CCT being optional factors.
EVO Formula
The Emmetropia Verifying Optical (EVO) formula was developed by Tun Kuan Yeo in Singapore (www.evoiolcalculator.com). It is a thick lens formula based on the theory of emmetropization. It uses AL, K, and ACD as the predictors, and LT and CCT are optional.
Panacea IOL Calculator
The Panacea IOL calculator was developed by David Flikier (www.panaceaiolandtoriccalculator.com). It is a vergence formula. In addition to the AL, keratometry, ACD, and LT, it also uses additional variables, such as ratio of posterior to anterior corneal radius of curvature for corneal value adjustment, and corneal asphericity in the IOL power calculation.
Pearl-DGS Calculator
The PEARL stands for Prediction Enhanced by ARtificial Intelligence and output Linearization, and DGS is named after the formula developers: Debellemanière, Gatinel, and Saad. This formula is based on artificial intelligence and optics (www.iolsolver.com). It uses several machine learning models that are selected according to the inputs entered by the surgeon and can adjust its prediction using the postoperative data of the contralateral eye if it is available.
Refractive Accuracy of IOL Formulas in Long Eyes
Table 63.1 shows the refractive accuracy of IOL power prediction in long eyes using different formulas reported in the literature over the past 10 years.
IOL Power Prediction Accuracy
Axial length adjustment methods ((Wang-Koch Adjustment): With the AL adjustment method, 64–82.4% of eyes have accuracy of refractive prediction errors ±0.5 D using Holladay 1 formula, 60–76.22% using SRK/T formula, and 71–84.21% using Holladay 2 formula. In an independent dataset of 1664 eyes with AL ≥ 25 mm used for the Hill-RBF formula development from Dr. Warren Hill, 93% of eyes had prediction errors of ±0.5 D using the modified AL adjustment Holladay 1 formula (unpublished data).
Super formula: With the Super formula, 55.3–83.33% of eyes have accuracy of refractive prediction errors ±0.5 D.
Barrett Universal II formula: The BUII formula is refined/optimized constantly. There are many studies that evaluated the accuracy of the BUII formula in long eyes, and 57.14–89.5% of eyes have accuracy of refractive prediction errors ±0.5 D.
Hill-RBF formula: The Hill-RBF 3.0 version was just released recently, and no study has yet reported its outcomes in long eyes. Several studies evaluated its accuracy in long eyes using the Hill-RBF 2.0 version, and 51.79–94.74% % of eyes had accuracy of refractive prediction errors ±0.5 D.
Olsen formula: Several studies evaluated the accuracy of Olsen formula in long eyes, and 65–89.47% of eyes had refractive prediction errors ±0.5 D.
Kane formula: A few studies evaluated the accuracy of Kane formula in long eyes, and 72–94.74% of eyes had refractive prediction errors ±0.5 D.
EVO formula: A few studies evaluated the accuracy of EVO formula in long eyes, and 79.27–89.47% of eyes had refractive prediction errors ±0.5 D.
Panacea IOL Calculator: One study evaluated the accuracy of Panacea IOL calculator in long eyes, and 63.16% of eyes had refractive prediction errors ±0.5 D.
Pearl-DGS Calculator: The Pearl-DGS calculator was introduced recently and there is no report of its accuracy in long eyes yet.
Comparison of Refractive Accuracy Among Formulas
The majority of studies reported that the performances of the above formulas were comparable in long eyes [16,17,18]. In general, studies have reported that the BUII, Kane, Hill-RBF, and Olsen formulas produced the best results or the lowest prediction errors, with no significant differences among those formulas.
The incidence of hyperopic outcomes (hyperopic relative to the predicted refraction) with the AL adjustment formulas was significantly lower than the BUII and Hill-RBF 2.0 (15–33% vs. 36–54%) [19]. Hill and colleagues [20] reported that the AL adjusted Holladay 1 produced less eyes with hyperopic outcomes (47.4–49%) than did the BUII and Hill-RBF 1.0 formulas (69.7–76.1%). Using the ULIB lens constants, Cheng et al. [21] found that the original AL adjustment Holladay 1 and SRK/T formulas produced significantly lower percentages of eyes (25.9–27.8%) with hyperopic outcomes than did the Kane, Hill-EBF 2.0, EVO, and BUII formulas (54.1–72.4%).
Conclusion
Due to the low IOL powers required in long eyes, accuracy of ELP estimation is not as important as in normal and short eyes. By adjusting the AL values used in Holladay 1, Holladay 2, and SRK/T formulas, excellent outcomes can be achieved. The refractive accuracy can be improved in long eyes with segmented ALs using specific refractive indices for each ocular medium. For IOL power calculation, based on the findings in the literature, any of the following formulas is a reasonable choice in long eyes: modified AL adjustment Holladay 1, BUII, Hill-RBF, Olsen, Kane, and EVO formulas.
References
Norrby S. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 2008;34(3):368–76.
Kora Y, Koike M, Suzuki Y, Inatomi M, Fukado Y, Ozawa T. Errors in IOL power calculations for axial high myopia. Ophthalmic Surg. 1991;22(2):78–81.
Zaldivar R, Shultz MC, Davidorf JM, et al. Intraocular lens power calculations in patients with extreme myopia. J Cataract Refract Surg. 2000;26(5):668–74.
MacLaren RE, Sagoo MS, Restori M, Allan BD. Biometry accuracy using zero- and negative-powered intraocular lenses. J Cataract Refract Surg. 2005;31(2):280–90.
Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238(9):765–73.
Wang L, Cao D, Weikert MP, Koch DD. Calculation of axial length using a single group refractive index versus using different refractive indices for each ocular segment: theoretical study and refractive outcomes. Ophthalmology. 2019;126(5):663–70.
Cooke DL, Cooke TL. A comparison of two methods to calculate axial length. J Cataract Refract Surg. 2019;45(3):284–92.
Petermeier K, Gekeler F, Messias A, Spitzer MS, Haigis W, Szurman P. Intraocular lens power calculation and optimized constants for highly myopic eyes. J Cataract Refract Surg. 2009;35(9):1575–81.
Wang L, Shirayama M, Ma XJ, Kohnen T, Koch DD. Optimizing intraocular lens power calculations in eyes with axial lengths above 25.0 mm. J Cataract Refract Surg. 2011;37(11):2018–27.
Wang L, Koch DD. Modified axial length adjustment formulas in long eyes. J Cataract Refract Surg. 2018;44(11):1396–7.
Wang L, Holladay JT, Koch DD. Wang-Koch axial length adjustment for the Holladay 2 formula in long eyes. J Cataract Refract Surg. 2018;44(10):1291–2.
Ladas JG, Siddiqui AA, Devgan U, Jun AS. A 3-D “super surface” combining modern intraocular lens formulas to generate a “super formula” and maximize accuracy. JAMA Ophthalmol. 2015;133(12):1431–6.
Barrett GD. Intraocular lens calculation formulas for new intraocular lens implants. J Cataract Refract Surg. 1987;13:389–96.
Hill WE. IOL power selection: think different. In: 11th annual Charles D. Kelman Lecture AAO, Las Vegas, Nevada; 2015.
Olsen T, Hoffmann P. C constant: new concept for ray tracing-assisted intraocular lens power calculation. J Cataract Refract Surg. 2014;40(5):764–73.
Melles RB, Holladay JT, Chang WJ. Accuracy of intraocular lens calculation formulas. Ophthalmology. 2018;125:169–78.
Melles RB, Kane JX, Olsen T, Chang WJ. Update on intraocular lens calculation formulas. Ophthalmology. 2019;126(9):1335.
Kane JX, Chang DF. Intraocular lens power formulas, biometry, and intraoperative aberrometry: a review. Ophthalmology. 2020;128(11):e94–e114.
Liu J, Wang L, Chai F, Han Y, Qian S, Koch DD, Weikert MP. Comparison of intraocular lens power calculation formulas in Chinese eyes with axial myopia. J Cataract Refract Surg. 2019;45(6):725–31.
Hill DC, Sudhakar S, Hill CS, King TS, Scott IU, Ernst BB, Pantanelli SM. Intraoperative aberrometry versus preoperative biometry for intraocular lens power selection in axial myopia. J Cataract Refract Surg. 2017;43(4):505–10.
Cheng H, Wang L, Kane JX, Li J, Liu L, Wu M. Accuracy of artificial intelligence formulas and axial length adjustments for highly myopic eyes. Am J Ophthalmol. 2020;223:100–7.
Zhang JQ, Zou XY, Zheng DY, Chen WR, Sun A, Luo LX. Effect of lens constants optimization on the accuracy of intraocular lens power calculation formulas for highly myopic eyes. Int J Ophthalmol. 2019;12(6):943–8.
Popovic M, Schlenker MB, Campos-Möller X, Pereira A, Ahmed IIK. Wang-Koch formula for optimization of intraocular lens power calculation: evaluation at a Canadian center. J Cataract Refract Surg. 2018;44(1):17–22.
Cooke DL, Cooke TL. Comparison of 9 intraocular lens power calculation formulas. J Cataract Refract Surg. 2016;42(8):1157–64.
Abulafia A, Barrett GD, Rotenberg M, Kleinmann G, Levy A, Reitblat O, Koch DD, Wang L, Assia EI. Intraocular lens power calculation for eyes with an axial length greater than 26.0 mm: comparison of formulas and methods. Cataract Refract Surg. 2015;41(3):548–56.
Zhang J, Tan X, Wang W, Yang G, Xu J, Ruan X, Gu X, Luo L. Effect of axial length adjustment methods on intraocular lens power calculation in highly myopic eyes. Am J Ophthalmol. 2020;214:110–8.
Savini G, Hoffer KJ, Balducci N, Barboni P, Schiano-Lomoriello D. Comparison of formula accuracy for intraocular lens power calculation based on measurements by a swept-source optical coherence tomography optical biometer. J Cataract Refract Surg. 2020;46(1):27–33.
Darcy K, Gunn D, Tavassoli S, Sparrow J, Kane JX. Assessment of the accuracy of new and updated intraocular lens power calculation formulas in 10 930 eyes from the UK National Health Service. J Cataract Refract Surg. 2020;46(1):2–7.
Carmona-González D, Castillo-Gómez A, Palomino-Bautista C, Romero-Domínguez M, Gutiérrez-Moreno MÁ. Comparison of the accuracy of 11 intraocular lens power calculation formulas. Eur J Ophthalmol. 2020;31(5):2370–6.
Kane JX, Van Heerden A, Atik A, Petsoglou C. Accuracy of 3 new methods for intraocular lens power selection. J Cataract Refract Surg. 2017;43(3):333–9.
Ji J, Liu Y, Zhang J, Wu X, Shao W, Ma B, Luo M. Comparison of six methods for the intraocular lens power calculation in high myopic eyes. Eur J Ophthalmol. 2021;31(1):96–102.
Omoto MK, Torii H, Hayashi K, Ayaki M, Tsubota K, Negishi K. Ratio of axial length to corneal radius in Japanese patients and accuracy of intraocular lens power calculation based on biometric data. Am J Ophthalmol. 2020;218:320–9.
Tang KS, Tran EM, Chen AJ, Rivera DR, Rivera JJ, Greenberg PB. Accuracy of biometric formulae for intraocular lens power calculation in a teaching hospital. Int J Ophthalmol. 2020;13(1):61–5.
Hipólito-Fernandes D, Elisa Luís M, Gil P, Maduro V, Feijão J, Yeo TK, Voytsekhivskyy O, Alves N. VRF-G, a new intraocular lens power calculation formula: a 13-formulas comparison study. Clin Ophthalmol. 2020;14:4395–402.
Zhou D, Sun Z, Deng G. Accuracy of the refractive prediction determined by intraocular lens power calculation formulas in high myopia. Indian J Ophthalmol. 2019;67(4):484–9.
Wan KH, Lam TCH, Yu MCY, Chan TCY. Accuracy and precision of intraocular lens calculations using the New Hill-RBF version 2.0 in eyes with high axial myopia. Am J Ophthalmol. 2019;205:66–73.
Rong X, He W, Zhu Q, Qian D, Lu Y, Zhu X. Intraocular lens power calculation in eyes with extreme myopia: comparison of Barrett universal II, Haigis, and Olsen formulas. J Cataract Refract Surg. 2019;45(6):732–7.
Roberts TV, Hodge C, Sutton G, Lawless M, contributors to the Vision Eye Institute IOL outcomes registry. Comparison of Hill-radial basis function, Barrett universal and current third generation formulas for the calculation of intraocular lens power during cataract surgery. Clin Exp Ophthalmol. 2018;46(3):240–6.
Connell BJ, Kane JX. Comparison of the Kane formula with existing formulas for intraocular lens power selection. BMJ Open Ophthalmol. 2019;4(1):e000251.
Kane JX, Van Heerden A, Atik A, Petsoglou C. Intraocular lens power formula accuracy: comparison of 7 formulas. J Cataract Refract Surg. 2016;42(10):1490–500.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Open Access This chapter is licensed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license and indicate if changes were made.
The images or other third party material in this chapter are included in the chapter's Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the chapter's Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.
Copyright information
© 2024 The Author(s)
About this chapter
Cite this chapter
Wang, L., Bezerra, R.L.F., Koch, D.D. (2024). IOL Power Calculation in Long Eye. In: Aramberri, J., Hoffer, K.J., Olsen, T., Savini, G., Shammas, H.J. (eds) Intraocular Lens Calculations. Essentials in Ophthalmology. Springer, Cham. https://doi.org/10.1007/978-3-031-50666-6_63
Download citation
DOI: https://doi.org/10.1007/978-3-031-50666-6_63
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-031-50665-9
Online ISBN: 978-3-031-50666-6
eBook Packages: MedicineMedicine (R0)