Abstract
The ongoing development of the Pentacam, from its origins in 2002 to the latest version (the Pentacam AXL Wave), is described. Today, the Pentacam AXL Wave combines Scheimpflug imaging, partial coherence interferometry, a Hartmann–Shack wavefront sensor and retroillumination. It enables an assessment of the total eye visual performance, including objective refraction and high-order aberration analysis, and a true separation of the internal wavefront from the total corneal wavefront (not possible with Placido technology).
The many features of the Pentacam family are illustrated, with specific attention to the IOL power calculation formulas available for unoperated and post-corneal refractive surgery eyes, as well as for eyes requiring toric IOLs to correct corneal astigmatism.
You have full access to this open access chapter, Download chapter PDF
Keywords
- Scheimpflug camera
- Tomography
- Topography
- Aberrometry
- Wavefront analysis
- Axial length
- Optical biometry
- IOL power calculation
- Tomography
- Total eye wavefront
- Retroillumination
- Objective refraction
Introduction
Cataract surgery is the most frequently performed eye surgery today—and IOL power calculation is a fascinating discipline in Ophthalmology!
May we introduce, your partners:
The use of the Pentacam family (Fig. 26.1) in modern cataract surgery can be described like a continues process (Fig. 26.2):
A few topics as listed below should be touched in this chapter:
-
Pentacam history and basic principle.
-
Some basic questions.
-
Every patient is an individuum = customization.
-
IOL power calculation formulas in the Pentacam.
-
Post-op visual assessment.
History and Basic Principle
The Pentacam family was born in 2002 and further expanded by the Pentacam HR in 2006. Both devices are based on a rotating Scheimpflug camera (Fig. 26.3), capturing high-resolution pictures of the anterior eye segment, from the cornea, down to the crystalline lens. The benefits of this technology are the snapshot-capturing of the single images, highest density in the corneal center, full cornea and scleral coverage, and a minimum of nose shadow.
The Pentacam contains since day one, a second camera, the iris camera, detecting eye motions during the scan process. The captured Scheimpflug images, up to 100, are composed to a three-dimensional model of the anterior eye segment in which the eye motions are corrected. A quality specification informs the user regarding the quality of the exam (Table 26.1).
This proven concept is reflected by its highest repeatability of keratometry [1,2,3,4], the most influential component in IOL power calculation. This might be due its tear film independency since Scheimpflug tomography does not require an intact tear film to reflect Placido rings or keratometry LEDs. No artificial tears should be applied before the measurements since this might change the normal conditions of the cornea. Moreover, objective crystalline lens density analysis [5] and grading of the nucleus [6] are possible.
In 2015, the Pentacam AXL was launched, combining the proven Scheimpflug tomography with optical biometry based on PCI technology and its comparability to the gold standard was proven [7]. This model includes the IOL calculator, containing IOL power calculation formulas for almost every cornea status, including IOL constant optimization. The IOL database is included, so no time-wasting collection of whatever IOLs are necessary. It contains up to 500 different IOL models from up to 35 manufactures who provided all details. Moreover, the IOL geometries are included for many IOLs, allowing total ray-traced-based IOL power calculation using the Olsen [8] formula.
The Pentacam AXL Wave launched in 2019 contains, besides the Scheimpflug tomography and optical biometry, a Hartmann–Shack wavefront sensor and retroillumination. These two features allow an assessment of the total eye visual performance, including objective refraction and high-order aberration analysis, and a post-op assessment of the IOL position in the human eye. The true separation of the internal wavefront from the total corneal wavefront (not possible with Placido technology) is the basis for a better understanding of individual visual quality and possible reasons for visual disturbance.
Some Basic Questions
Is Pure Keratometry and Axial Length Enough for IOL Power Calculation Today?
The most often used IOL power formulas like SRK/T [9], Haigis [10], Hoffer Q [8], and Holladay 1 [10] use axial length and keratometry for the calculation of the IOL power and for the prediction of the position of the IOL in the pseudophakic eye—whereby the Haigis formula uses the anterior chamber depth, measured from the epithelium as well. Every IOL formula has at least two components, the calculation of the power and the prediction of its position in the pseudophakic eye, and the second component is of highest interest and the biggest source of errors today. To improve this, many more factors are taken into account like HCD (corneal diameter), thickness of the human lens, and others are necessary. To achieve low prediction errors and less post-op surprises, more parameters have to be considered like the Barrett Universal 2 [11].
Does One Formula for Every Purpose Exist?
One formula for every purpose which should achieve best results no matter what the cornea look like may still not exist today. Many formulas exist for corneas after laser refractive interventions [11,12,13,14,15], for keratoconus [8], for the correction of astigmatism [16, 17] and ray tracing formulas for every corneal shape, including corneal transplants and all the odd corneal shapes like corneal transplants and others.
Every IOL for Every Patient?
The development of different IOL designs to improve our patients’ visual performance is a blessing but requires careful patients selection. Our patients are entitled to understand about the possibilities and limitations to adjust expectations and avoid disappointments after surgery.
Every Patient Is an Individuum = Customization
Considering the fact that just keratometry and axial length is not enough to achieve top- outcome, that individual formulas might be necessary and patients selection is key to success [18], more than just a pure standard optical biometer is necessary.
The OCULUS Pentacam addresses this in particular (Fig. 26.4):
Corneal Morphology Assessment
Corneal Tomography = total cornea assessment has its benefit over pure corneal topography [19]. The total cornea is analyzed and described like a thick lens: anterior and posterior surface and its thickness at every single position are known. Scheimpflug tomography analyzes the cornea in almost every detail and provides important information to detect abnormalities and diseases:
-
The keratometry and the assessment of the corneal astigmatism is basic knowledge. Moreover, the low astigmatism and the consideration of the posterior surface providing Total Corneal Refractive Astigmatism without any assumptions is a step forward [15, 16].
-
The topography maps of the anterior and posterior surface highlights irregular corneal shape [20].
-
The analysis regarding possible laser refractive or other surgical interventions using tomography, the B/F-ratio (back to front ratio) of the cornea, which is for normal eyes around 82% [21], plays an important role too. This factor is lower for post-myopic and higher for post-hyperopic laser surgery and for post-RK (radial keratotomy).
-
The Belin/Ambrosio Enhanced Ectasia separates normal from abnormal patients and supports in the detection of corneal ectasia while having a final color-coded parameter, the “D”-value [20].
-
The early detection of Endothelium Fuchs Dystrophy became more important. It is a progressive disease which requires sooner or later a posterior cornea transplant (DMEK or DSEK). The Pentacam supports in the early detection [22]. This often results in a post-op hyperopic shift. Arising questions are first, the best surgical planning, combined or in two steps and second the formula which should be used after the corneal transplant. A good option could be the Olsen ray tracing formula.
The More Complex Corneas: How to Deal with It?
These corneas, often after refractive surgery or other surgical interventions, are always a challenge in IOL power calculation. The corneal power distribution display is a powerful assessment tool for these cases. But not only these challenging cases might be of interest, it just starts with the assessment of the astigmatisms (Fig. 26.5):
-
Is the magnitude and the axis in the central zone same or different compared to the periphery?
-
Does it change and if yes how much?
-
How is this related to the pupil diameter, does it matter?
-
What about the orientation of the axis of the astigmatism, for a WTR (with the rule), ATR (against the rule), or oblique astigmatism?
-
What about the influence of the posterior corneal surface in terms of possible axis shifts? Does it matter and if yes, which IOL formula approach should be used?
The example below shows a patient after LASIK with a homogenous ablation zone and a small corneal power distribution (Fig. 26.6).
On the other hand, an example of a post-LASIK patients with a decentered ablation and flap problems (Fig. 26.7).
Corneal Optical and Anterior Chamber Properties Assessment
The human eye is not an optical bench. Hence, it requires individual assessment of corneal optical and anterior chamber properties which are “solid-factors” for IOL selection.
The Pentacam tomography calculates the total corneal wavefront, considering the posterior surface.
-
The total spherical aberrations, which are often associated with halos, starburst, ghost images, and loss of contrast sensitivity, are important to measure. This supports the selection of an aspherical or an aberration-neutral IOL design. Normative values are provided [23].
-
Increased coma, which causes an optical effect like a comet tail and may result in double-vision, might be contraindication for multifocal IOLs.
-
Increased trefoil, which spreads the light in three directions, is important to quantify as well.
They do not occur individually, and they are limiting factors for the visual performance per se. The Pentacam provides all these values, including cut-off suggestions, supporting in the selection of multifocal IOLs.
The Pentacam provides the anterior chamber depth, measured form the epithelium as well as the anterior chamber depth measured from the endothelium. angle is calculated in every Scheimpflug image and is used for the selection if a patient might be suitable for a pIOL implantation. Please note that for pseudophakic eyes the anterior chamber depth should be double-checked.
Centration of Optical Elements and Pupil Diameter
Pentacam tomography provides parameters associated with the optical path of the individual eye.
The vectorial distance between the vertex normal—the reference for all Pentacam measurements—and the pupil center, called chord μ and chord α which is the distance between vertex normal and the corneal geometric center. If they are high, there might be a risk for reduced visual performance.
The Pentacam AXL Wave provides the pupil diameter under day and night conditions. In combination with the corresponding refraction, additional support for cataract refractive surgery is provided.
Total Eye Visual Performance
The Pentacam AXL Wave with its built-in Hartmann–Shack sensor for total eye wavefront has the ability to display the source, or the reasons for visual impairments. The example below (Fig. 26.8) shows an early presbyopia case of a female aged 47 with a previous myopic LASIK. The reason for her typical problems, like driving at night or when it is rainy and foggy, is the crystal lens. This picture helps her understand immediately the reason.
The example below shows a patient with previous RK (Fig. 26.9) having high expectations in the cataract surgery. No matter which lens you are going to implant, the visual quality will never be as good as expected. The patient understood—the image told more than 1000 words.
IOL Power Calculation Formulas in the Pentacam
The improvement of IOL power calculation formulas is a process. The IOL calculator built into the Pentacam AXL and Pentacam AXL Wave includes IOL power calculation formulas for almost every purpose and the IOL database—ready to use. No online calculators have to be assessed.
The Pentacam keratometry was proven to be most accurate, for normal and abnormal corneas [2, 4, 12]. Combined with precise axial length and other required parameter, the basis is made to achieve very good refractive outcomes [21].
Every single surgeon in a bigger clinic can have his/her own profile with individual combinations of IOLs with IOL power calculation formulas. For the calculation of toric IOLs, the SIA (surgical-induced astigmatism) has to be entered and is considered in the respective formulas. The IOL calculator displays the standard parameters as well as total corneal spherical and high-order aberrations. Abnormal values are highlighted to inform the user (Fig. 26.10).
Monofocal IOL Formulas for Virgin Corneas
The IOL formulas for monofocal IOLs are intuitively organized and contain the most common standard and modern IOL formulas (Figs. 26.11 and 26.12).
Toric IOL Formulas for Virgin Corneas
The IOL power calculation for toric IOLs offers formulas (Fig. 26.13) with measured and with estimated posterior surface. The estimated post-op refraction as well as the orientation of the toric implant are shown (Fig. 26.14).
IOL Formulas for Patients After Corneal Laser Refractive Surgery and RK (Radial Keratotomy)
This is still a challenge today. The IOL calculator offer customized formulas [13, 14] for the Pentacam (Fig. 26.15) as well as the Barrett True K and the double-K formulas [17]. Latest study has shown very good results using the Barrett True K formula with increasing precision the more information prior history are available [24]. On the other hand, the Olsen ray tracing formula is fully independent of any information prior refractive surgery [25] (Fig. 26.16).
The Pentacam with its rotating Scheimpflug tomography allow to measure even the very irregular corneas. For patients having had previous corneal refractive surgery with a remaining high astigmatism as well as for patients having had PKP (penetrating keratoplasty), the Olsen formula can be used (Fig. 26.17).
The Post-op Visual Assessment
The post-op visual assessment for documentation, quality assessment, and continuous improvement is a must today in modern cataract surgery.
The subjective refraction is one parameter combined with the slit-lamp exam, and a final short talk to the patients is routine anyway. But, what to do and how to handle unhappy patients? We all heard about the “20/20 unhappy patients.” Here, the Pentacam can be of help again.
The Pentacam AXL Wave performs total eye wavefront, objective refraction, biometry, and tomography, providing a solid basis for further diagnosis—before the physicians starts the conversation with the patient.
The first example shows a happy patient after multifocal-toric implantation (Fig. 26.18). The refraction is almost plano, the Total Visual Performance is very good, and the IOL is on axis.
The example below is an example of an unhappy patient with bad visual quality after cataract surgery (Fig. 26.19). The Pentacam AXL Wave shows the Total Visual Performance and the refraction at a glance.
The retroillumination image below gives the answer, and it is a decentered IOL (Fig. 26.20).
Summary and Take-Home Message
The Pentacam AXL as well as the Pentacam AXL Wave offer the full-capacity performing IOL power calculation on the highest level. Besides this, it offers so many other clinical applications, making it the “swiss-army-knife” for every eye clinic.
References
Rozema JJ, Wouters K, Mathysen DGP, Tassignon M-J. Overview of the repeatability, reproducibility, and agreement of the biometry values provided by various ophthalmic devices. Am J Ophthalmol. 2014;158(6):1111–1120.e1. http://www.ncbi.nlm.nih.gov/pubmed/25128596
Shetty R, Arora V, Jayadev C, Nuijts RMMA, Kumar M, Puttaiah NK, Kummelil MK. Repeatability and agreement of three Scheimpflug-based imaging systems for measuring anterior segment parameters in keratoconus. Invest Ophthalmol Vis Sci. 2014;55(8):5263–8. http://www.ncbi.nlm.nih.gov/pubmed/25074774
Visser N, Berendschot TTJM, Verbakel F, de Brabander J, Nuijts RMMA. Comparability and repeatability of corneal astigmatism measurements using different measurement technologies. J Cataract Refract Surg. 2012;38(10):1764–70. http://www.ncbi.nlm.nih.gov/pubmed/22999600
Fityo S, Bühren J, Shajari M, Kohnen T. Keratometry versus total corneal refractive power: analysis of measurement repeatability with 5 different devices in normal eyes with low astigmatism. J Cataract Refract Surg. 2016;42(4):569–76. http://www.ncbi.nlm.nih.gov/pubmed/27113880
Weiner X, Baumeister M, Kohnen T, Bühren J. Repeatability of lens densitometry using Scheimpflug imaging. J Cataract Refract Surg. 2014;40(5):756–63. http://www.ncbi.nlm.nih.gov/pubmed/24767909
Nixon DR. Preoperative cataract grading by Scheimpflug imaging and effect on operative fluidics and phacoemulsification energy. J Cataract Refract Surg. 2010;36(2):242–6. http://www.ncbi.nlm.nih.gov/pubmed/20152604
Shajari M, Cremonese C, Petermann K, Singh P, Müller M, Kohnen T. Comparison of axial length, corneal curvature, and anterior chamber depth measurements of 2 recently introduced devices to a known biometer. Am J Ophthalmol. 2017;178:58–64. http://www.ncbi.nlm.nih.gov/pubmed/28263734
Olsen T, Jeppesen P. Ray-tracing analysis of the corneal power from Scheimpflug data. J Refract Surg. 2018;34(1):45–50. http://www.ncbi.nlm.nih.gov/pubmed/29315441
Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238:765–73.
Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19(11):700–12. Errata: 1994;20(6):677 and 2007;33(1):2–3.
Aramberri J. Intraocular lens power calculation after corneal refractive surgery: Double-K method. J Cataract Refract Surg. 2003;29(11):2063–8. http://www.sciencedirect.com/science/article/pii/S088633500300957X
Camellin M, Savini G, Hoffer KJ, Carbonelli M, Barboni P. Scheimpflug camera measurement of anterior and posterior corneal curvature in eyes with previous radial keratotomy. J Refract Surg. 2012;28(4):275–9. http://www.ncbi.nlm.nih.gov/pubmed/22386371
Savini G, Hoffer KJ, Barrett GD. Results of the Barrett True-K formula for IOL power calculation based on Scheimpflug camera measurements in eyes with previous myopic excimer laser surgery. J Cataract Refract Surg. 2020;46(7):1016–9. https://pubmed.ncbi.nlm.nih.gov/32271267/
Potvin R, Hill W. New algorithm for post-radial keratotomy intraocular lens power calculations based on rotating Scheimpflug camera data. J Cataract Refract Surg. 2013;39(3):358–65. http://www.ncbi.nlm.nih.gov/pubmed/23337527
Potvin R, Hill W. New algorithm for intraocular lens power calculations after myopic laser in situ keratomileusis based on rotating Scheimpflug camera data. J Cataract Refract Surg. 2015;41(2):339–47. http://www.ncbi.nlm.nih.gov/pubmed/25661127
Savini G, Næser K. An analysis of the factors influencing the residual refractive astigmatism after cataract surgery with toric intraocular lenses. Invest Ophthalmol Vis Sci. 2015;56(2):827–35. https://doi.org/10.1167/iovs.14-15903.
Savini G, Næser K, Schiano-Lomoriello D, Ducoli P. Optimized keratometry and total corneal astigmatism for toric intraocular lens calculation. J Cataract Refract Surg. 2017;43(9):1140–8. http://www.ncbi.nlm.nih.gov/pubmed/28991609
Donaldson K, Fernández-Vega-Cueto L, Davidson R, Dhaliwal D, Hamilton R, Jackson M, et al. Perioperative assessment for refractive cataract surgery. J Cataract Refract Surg. 2018;44(5):642–53. http://www.ncbi.nlm.nih.gov/pubmed/29891157
Tonn B, Klaproth OK, Kohnen T. Anterior surface–based keratometry compared with Scheimpflug tomography–based total corneal astigmatism. 2015. http://iovs.arvojournals.org/article.aspx?articleid=2212759
Villavicencio OF, Gilani F, Henriquez MA, Izquierdo L, Ambrósio RR. Independent population validation of the Belin/Ambrósio enhanced ectasia display: implications for keratoconus studies and screening. Int J Keratoconus Ectatic Corneal Dis. 2014;3(1):1–8. https://doi.org/10.5005/jp-journals-10025-1069.
Ho J-D, Tsai C-Y, Tsai RJ-F, Kuo L-L, Tsai I-L, Liou S-W. Validity of the keratometric index: evaluation by the Pentacam rotating Scheimpflug camera. J Cataract Refract Surg. 2008;34(1):137–45. https://pubmed.ncbi.nlm.nih.gov/18165094/
Ní Dhubhghaill S, Rozema JJ, Jongenelen S, Ruiz Hidalgo I, Zakaria N, Tassignon MJ. Normative values for corneal densitometry analysis by Scheimpflug Optical Assessment. Invest Ophthalmol Vis Sci. 2014;55:162–8. http://iovs.arvojournals.org/article.aspx?articleid=2128026
Klaproth OK, Buehren J, Otto K, Kohnen T. Repeatability of the corneal wavefront measurements with Pentacam HR. 2011. https://iovs.arvojournals.org/article.aspx?articleid=2356309
Taroni L, Hoffer KJ, Barboni P, Schiano-Lomoriello D, Savini G. Outcomes of IOL power calculation using measurements by a rotating Scheimpflug camera combined with partial coherence interferometry. J Cataract Refract Surg. 2020;46(12):1618–23. https://pubmed.ncbi.nlm.nih.gov/32818357/
Barrett GD. An improved universal theoretical formula for intraocular lens power prediction. J Cataract Refract Surg. 1993;19(6):713–20. https://doi.org/10.1016/S0886-3350(13)80339-2.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Open Access This chapter is licensed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license and indicate if changes were made.
The images or other third party material in this chapter are included in the chapter's Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the chapter's Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.
Copyright information
© 2024 The Author(s)
About this chapter
Cite this chapter
Iwanczuk, J. (2024). The Pentacam Family. In: Aramberri, J., Hoffer, K.J., Olsen, T., Savini, G., Shammas, H.J. (eds) Intraocular Lens Calculations. Essentials in Ophthalmology. Springer, Cham. https://doi.org/10.1007/978-3-031-50666-6_26
Download citation
DOI: https://doi.org/10.1007/978-3-031-50666-6_26
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-031-50665-9
Online ISBN: 978-3-031-50666-6
eBook Packages: MedicineMedicine (R0)