Abstract
The OA-2000 is an optical biometer that automatically measures all parameters needed for IOL power calculation. Axial measurements are obtained by means of swept-source optical coherence tomography and corneal measurements by means of a Placido ring corneal topographer. Long coherence length allows high penetration and high success rate on long axial length myopic eyes. Different technical features provide a high success rate in measuring dense cataract eyes. Corneal keratometry and astigmatism are measured with accuracy and precision. The device includes different kinds of IOL formulae: standard, toric, and post-Lasik. In addition, software for optimization of IOL constants is included in OA-2000.
You have full access to this open access chapter, Download chapter PDF
Keywords
- Optical biometry
- Swept-source
- Optical coherence tomography
- Corneal topography
- Placido disc
- IOL power
- IOL formula
- Toric IOL
- Cataract
- Axial length
- Lens thickness
- Keratometry
- Anterior chamber depth
Introduction
Optical biometer OA-2000 (Tomey Corporation, Nagoya Japan) can measure all biometric parameters needed for pre-cataract surgery: axial length, anterior chamber depth, lens thickness, corneal radius of curvature over a 2.5 mm and 3.0 mm diameter, corneal topography, central corneal thickness, pupil diameter, and Corneal Diameter (Fig. 21.1 and Table 21.1). It is done almost automatically by one measurement in short time. The touch screen enables intuitive operation, measurement, data checking, IOL calculation, and data output within one compact unit. The OA-2000 supports doctors to reduce stress on patients and accurate and smooth preoperative examination and planning for high-quality cataract surgery by these features on daily practice [1].
Axial Length Measurement
The touch-alignment system is employed on OA-2000. It can provide a stable measurement and a high reproducibility between operators because OA-2000 aligns itself automatically to patient eye once an operator only touches the center of the pupil on the screen. Operators can measure all data easily without any special skills (Fig. 21.2) [2].
Touch alignment, easy operation. http://vimeo.com/259629303?width=640&height=480
OA-2000 achieves a high signal–noise ratio (SNR) and high transmitting on opacity parts by using swept-source coherence tomography method. In addition, long coherence length allows high penetration and high success rate on long axial length myopic eyes (Fig. 21.3) [3]. Additionally, OA-2000 obtains not only A-scan wave form but also B-scan image of 1 mm width on retina by 2D scanning. On a measurement result of normal eye (Fig. 21.4), SNR is high and standard deviation (SD) is very low because the measurement light isn’t reduced as the crystalline lens is clear. It means the measurement is very stable.
On the other hand, there are some cases where it is difficult to obtain the signal around the center of the retina as shown in Fig. 21.5a. In such case, OA-2000 measures the axial length by detecting a stronger signal from peripheral area automatically. When the retinal signal cannot be detected even with that way, V scan (Fig. 21.6) is performed instead of horizontal scan to find less opacity position so that the retinal signal can be detected. High success rate is achieved by these techniques in combination.
In the case of dense cataract with low SNR, multi-peaks retinal waveforms may appear as Fig. 21.5b shows. OA-2000 detects retinal pigment epithelium (RPE) automatically by an original algorithm based on signal appearance ratio and peak analysis results. It is recommended to judge it comprehensively including the results of other examinations, when it is suspected if the detected retinal peak is correct or not, for example, miss-detecting epiretinal membrane as RPE. Caliper function (Fig. 21.7) allows operators to select RPE position manually if the auto-detected RPE is not correct.
For converting optical path lengths (OPLs) to geometrical distances, an original conversion formula was established by clinical dataset of ultrasound biometer and OA-1000, how to be performed based on the way by Haigis [4, 5]. OPLs of OA-2000 are in good agreement with OA-1000, which is the previous model; OA-2000 also uses the same conversion formula.
Clinical Cases
Clinical cases of cataract eyes are introduced in Fig. 21.8.
Clinical cases of cataract eyes, Provided by Dr. Hitoshi Tabuchi, Chief director of Department of Ophthalmology, Tsukazaki Hospital. (a) Hypermature cataract eye. The nuclear sclerosis is Emery class V. SDs for axial length, ACD, and lens show small numbers even with relatively low SNRs; measurement result is stable. (b) Cortex and posterior subcapsular cataract eye. The nuclear sclerosis is Emery class III. SDs for axial length, ACD, and lens show small numbers even with relatively low SNRs; measurement result is stable.. (c) Cortex and posterior subcapsular cataract eye. The nuclear sclerosis is Emery class II-III. The retinal signal is very low because the opacity of the cortex and the posterior subcapsular is strong. However, SDs are small enough and the measurement result is stable
OA-2000 includes the function of wireless connection with ultrasound biometer AL-4000 (Fig. 21.9). It is useful when measurement is difficult by an optical biometer due to subject eyes conditions. The measurement results of AL-4000 can be transferred to OA-2000, and it can be used for IOL calculation on OA-2000.
Measurement of Corneal Radius of Curvature and Topography
OA-2000 employs a Placido disc for measuring corneal radii and topography, which has been proven being an accurate measurement method and used for many years [6]. The Placido-based topography can capture image in a single shot; therefore, it is hard to be affected by eye motion, when performed at the same time as the axial length measurement. Figure 21.10 shows examples of comparison of topography map between TMS-4N and OA-2000 for same eyes. The measurement range of OA-2000 is 5.5 mm which is narrower than TMS-4N. However, it is enough for evaluating visual function, and they have good consistency in that range.
Important Indices of Cornea for Cataract Surgery
OA-2000 provides not only the corneal radius of curvature results but also important indices of KAI (Kerato-Asymmetry Index) and KRI (Kerato-Regularity Index) (Fig. 21.11). These indices are ranked with A (Low)/B (Slightly high)/C (High) which show possibility to be irregular corneal astigmatism. When KAI shows high value in B or C, the eye is suspected to be a deformed cornea such as in keratoconus. When KRI shows high value, the eye is suspected to be a corneal transplant or with CL-induced problems, etc. It can call to attention checking the topographic maps so that doctors can prevent postoperative troubles. In the case that these values are high, there is higher risk of insufficient visual recovery and it should be carefully considered to choose multifocal IOLs.
CEI (Corneal Eccentricity Index) is shown in the topographic screen (Fig. 21.12). When CEI indicates positive number, the corneal shape is prolate which is normal eye. On the other hand, when CEI indicates negative number, the corneal shape is oblate which can be observed in typical post-LASIK eyes. By checking this value, even in the case that the surgical history is unknown, doctors can realize the possibility being post-LASIK eye and adopt post-LASIK IOL formula so that they can avoid refractive surprise after surgery.
IOL Calculation and Toric Calculation
OA-2000 includes different kinds of IOL formulae, which are standard IOL formulae, toric IOL formulae, and post-Lasik IOL formulae (Table 21.1 and Fig. 21.13). In addition, function for optimization of IOL constants is included in OA-2000 and it can support to calculate personal lens constant for each surgeon (Fig. 21.14).
Toric planning function to support with axis registration is available (Fig. 21.15). It allows doctors to mark the target axis based on the reference axis of iris pattern or conjunctival blood vessels.
References
Savini G, Hoffer KJ, Shammas HJ, Aramberri J, Huang J, Barboni P. Accuracy of a new swept-source optical coherence tomography biometer for IOL power calculation and comparison to IOLMaster. J Refract Surg. 2017;33(10):690–5.
Wang W, Miao Y, Savini G, McAlinden C, Chen H, Hu Q, Wang Q, Huang J. Precision of a new ocular biometer in eyes with cataract using swept source optical coherence tomography combined with Placido-disk corneal topography. Sci Rep. 2017;7(1):13736.
McAlinden C, Wang Q, Gao R, Zhao W, Yu A, Li Y, Guo Y, Huang J. Axial length measurement failure rates with biometers using swept-source optical coherence tomography compared to partial-coherence interferometry and optical low-coherence interferometry. Am J Ophthalmol. 2017;173:64–9.
Mizushima Y, Kawana K, Suto C, Shimamura E, Fukuyama M, Oshika T. Evaluation of axial length measurement with new partial coherence interferometry OA-1000. Jpn J Ophthalmic Surg. 2010;23(3):453–7.
Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238(9):765–73.
Guilbert E, Saad A, Grise-Dulac A, Gatinel D. Corneal thickness, curvature, and elevation readings in normal corneas: combined Placido-Scheimpflug system versus combined Placido-scanning-slit system. J Cataract Refract Surg. 2012;38(7):1198–206.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Open Access This chapter is licensed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license and indicate if changes were made.
The images or other third party material in this chapter are included in the chapter's Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the chapter's Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.
Copyright information
© 2024 The Author(s)
About this chapter
Cite this chapter
Hara, N., Benedikt, K., Owaki, H. (2024). Optical Biometer OA-2000. In: Aramberri, J., Hoffer, K.J., Olsen, T., Savini, G., Shammas, H.J. (eds) Intraocular Lens Calculations. Essentials in Ophthalmology. Springer, Cham. https://doi.org/10.1007/978-3-031-50666-6_21
Download citation
DOI: https://doi.org/10.1007/978-3-031-50666-6_21
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-031-50665-9
Online ISBN: 978-3-031-50666-6
eBook Packages: MedicineMedicine (R0)