Abstract
Patients’ telemetry deals with remote follow-up of vital parameters and clinical conditions without continuous direct contact between physician and patient. This study presents a telemetry system for real-time monitoring of cardiac patients and investigates its accuracy, reliability and applicability. The system includes personal analyzer (PA) for continuous acquisition and analysis of single-channel electrocardiogram (ECG), central unit and a console for the cardiologist. PA transfers to the central unit 10 s ECG segments on regular basis (5/10/15 min), on demand or when an alarm based on rhythm disorders and/or heartbeat abnormalities is generated.
The ECG telemetry system is tested on 85 hospitalized patients with cardiovascular diseases. In parallel, the patients’ ECGs have been recorded by a Holter device Signa-Lyzer SD. An experienced cardiologist has compared the ECG segments that have provoked an alarm to the respective ECG segments in the Holter recording in order to judge about the correctness (accuracy) of the alarm, as well as has checked the Holter ECG recording for events that are missed by the PA (reliability).
Full correspondence between the PA alarms and the rhythms recorded via the Holter has been observed (100% accuracy). Considering the witnessed Holter events, PA reliably warns for 80.6% of the rhythm disorders. In the remaining 19.4% an alarm is not generated mainly due to the fact that the patients have been out of the telemetry range.
The observed results show that the tested PA could support timely diagnosis, prompt reaction and adequate therapeutic treatment in potentially life-threatening clinical conditions.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Similar content being viewed by others
References
Komalasari, R., Nurjanah, N., Yoche, M.: Quality of life of people with cardiovascular disease: a descriptive study. Asian/Pacific Island Nurs. J. 4(2), 92–96 (2019). https://doi.org/10.31372/20190402.1045
Kvedar, J., Coye, M., Everett, W.: Connected health: a review of technologies and strategies to improve patient care with telemedicine and telehealth. Health Aff (Millwood) 33, 194–199 (2014)
Global ECG Telemetry Devices Market - Growth, Trends, and Forecast (2019 - 2024), November (2019). https://www.researchandmarkets.com/r/fwhusd
Marouf, M., Vukomanovic, G., Saranovac, L., Bozic, M.: Multipurpose ECG telemetry system. Biomed. Eng. Online 2017, 1–20 (2017). https://doi.org/10.1186/s12938-017-0371-6
Linzer, M., Pritchett, E., Pontinen, M., McCarthy, E., Divine, G.: Incremental diagnostic yield of loop electrocardiographic recorders in unexplained syncope. Am. J. Cardiol. 66(2), 214–219 (1990)
Zimetbaum, P., Kim, K., Josephson, M., Goldberger, A., Cohen, D.: Diagnostic yield and optimal duration of continuous-loop event monitoring for the diagnosis of palpitations: a cost-effectiveness analysis. Ann. Intern. Med. 128(11), 890–895 (1998)
Kohno, R., Abe, H., Benditt, D.: Ambulatory electrocardiogram monitoring devices for evaluating transient loss of consciousness or other related symptoms. J Arrhythm. 33, 583–589 (2017)
Wu, C., et al.: Utility of patient-activated cardiac event recorders in the detection of cardiac arrhythmias. J. Interv. Card. Electrophysiol. 8(2), 117–120 (2003). https://doi.org/10.1023/A:1023604816368
Kaleschke, G., et al.: Prospective, multicenter validation of a simple, patient-operated electrocardiographic system for the detection of arrhythmias and electrocardiographic changes. Europace 11(10), 1362–1368 (2009)
Park, M., de Asmundis, C., Chierchia, G., Sarkozy, A., Benatar, A., Brugada, P.: First experience of monitoring with cardiac event recorder electrocardiography omron system in childhood population for sporadic, potentially arrhythmia related symptoms. Europace 13(9), 1335–1339 (2011)
Attanasio, P., et al.: Use of a patient-activated event recording system in patients with tachycardic palpitations: how long to follow up? Ann Noninvasive Electrocardiol. 20, 566–569 (2015)
Turakhia, M., et al.: Latency of ECG displays of hospital telemetry systems: a science advisory from the american heart association. Circulation 126(13), 1665–1669 (2012). https://doi.org/10.1161/CIR.0b013e31826ae459
Steinberg, J., Varma, N., Cygankiewicz, I., et al.: 2017 ISHNE-HRS expert consensus statement on ambulatory ECG and external cardiac monitoring/telemetry. Heart Rhythm 14(7), e55–e96 (2017). https://doi.org/10.1016/j.hrthm.2017.03.038
Bocchiardo, M., Asteggiano, R.: ECG portable devices: example of e-Health strength and threats. e-J. Cardiol. Pract. 18(25) (2020). https://www.escardio.org/Journals/E-Journal-of-Cardiology-Practice/Volume-18/ecg-portable-devices-example-of-e-health-strength-and-threats
Ramkumar, S., Nerlekar, N., D’Souza, D., Pol, D., Kalman, J., Marwick, T.: Atrial fibrillation detection using single lead portable electrocardiographic monitoring: a systematic review and meta-analysis. BMJ Open. 8, e024178 (2018)
Emergency Cardiac Care Committee members: Recommended guidelines for in-hospital cardiac monitoring of adults for detection of arrhythmia. J. Am. Coll. Cardiol. 18(6), 1431–1433 (1991)
Dhillon, S., et al.: Effectiveness of telemetry guidelines in predicting clinically significant arrhythmias in hospitalized patients. Cardiol. Res. 3(1), 16–22 (2012)
United Healthcare Oxford Clinical Policy (2018). https://www.uhcprovider.com/en/policiesprotocols/commercial-policies/oxford-policies.html
Tabakov, S., Iliev, I., Krasteva, V.: Online digital filter and QRS detector applicable in low resource ECG monitoring systems. Ann. Biomed. Eng. 36(11), 1805–1815 (2008). https://doi.org/10.1007/s10439-008-9553-5
Iliev, I., Krasteva, V., Tabakov, S.: Real-time detection of pathological cardiac events in the electrocardiogram. Physiol. Meas. 28, 259–276 (2007)
Acknowledgement
This work was supported by the Bulgarian National Science Fund under Grant Ref. No. КП-06-H37/9 “Audio transformation (sonification) of the electrocardiogram – a new approach for remote monitoring of the heart activity” and the European Regional Development Fund within the Operational Programme “Science and Education for Smart Growth 2014–2020” under the Project CoE “National center of mechatronics and clean technologies“ BG05M2OP001-1.001-0008.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2022 The Author(s), under exclusive license to Springer Nature Switzerland AG
About this paper
Cite this paper
Iliev, I., Jekova, I., Tabakov, S., Koshtikova, K., Runev, N., Manov, E. (2022). High-Risk Cardiac Patients’ Follow-Up via Portable Telemonitoring Personal Analyzer: Applicability, Reliability and Accuracy. In: Sotirov, S.S., Pencheva, T., Kacprzyk, J., Atanassov, K.T., Sotirova, E., Staneva, G. (eds) Contemporary Methods in Bioinformatics and Biomedicine and Their Applications. BioInfoMed 2020. Lecture Notes in Networks and Systems, vol 374. Springer, Cham. https://doi.org/10.1007/978-3-030-96638-6_33
Download citation
DOI: https://doi.org/10.1007/978-3-030-96638-6_33
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-030-96637-9
Online ISBN: 978-3-030-96638-6