Energy Dispersive X-ray Spectrometry: Physical Principles and User-Selected Parameters

  • Chapter
  • First Online:
Scanning Electron Microscopy and X-Ray Microanalysis

Abstract

As illustrated in Fig. 16.1, the physical basis of energy dispersive X-ray spectrometry (EDS) with a semiconductor detector begins with photoelectric absorption of an X-ray photon in the active volume of the semiconductor (Si). The entire energy of the photon is transferred to a bound inner shell atomic electron, which is ejected with kinetic energy equal to the photon energy minus the shell ionization energy (binding energy), 1.838 keV for the Si K-shell and 0.098 keV for the Si L-shell. The ejected photoelectron undergoes inelastic scattering within the Si crystal. One of the consequences of the energy loss is the promotion of bound outer shell valence electrons to the conduction band of the semiconductor, leaving behind positively charged “holes” in the valence band. In the conduction band, the free electrons can move in response to a potential applied between the entrance surface electrode and the back surface electrode across the thickness of the Si crystal, while the positive holes in the conduction band drift in the opposite direction, resulting in the collection of electrons at the anode on the back surface of the EDS detector. This charge generation process requires approximately 3.6 eV per electron hole pair, so that the number of charge carriers is proportional to the original photon energy, E p:

$$ n={E}_p/3.6\ eV $$

For a Mn K-L3 photon with an energy of 5.895 keV, approximately 1638 electron–hole pairs are created, comprising a charge of 2.6 × 10−16 coulombs. Because the detector can respond to any photon energy from a threshold of approximately 50 eV to 30 keV or more, the process has been named “energy dispersive,” although in the spectrometry sense there is no actual dispersion such as occurs in a diffraction element spectrometer.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Institutional subscriptions

References

  • Fiori C, Newbury D (1978) Artifacts in energy dispersive X-ray spectrometry. Scan Electron Microsc 1:401

    Google Scholar 

  • Fitzgerald R, Keil K, Heinrich KFJ (1968) Solid-state energy-dispersive spectrometer for electron-microprobe X-ray analysis. Science 159:528

    Article  CAS  Google Scholar 

  • Gatti E, Rehak P (1984) Semiconductor drift chamber – an application of a novel charge transport scheme. Nucl Instr Meth A 225:608

    Article  CAS  Google Scholar 

  • Struder L, Fiorini C, Gatti E, Hartmann R, Holl P, Krause N, Lechner P, Longini A, Lutz G, Kemmer J, Meidinger N, Popp M, Soltau H, Van Zanthier C (1998) High-resolution high-count-rate X-ray spectroscopy with state-of-the-art silicon detectors. Mikrochim Acta Suppl 15:11

    CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

Copyright information

© 2018 Springer Science+Business Media LLC

About this chapter

Check for updates. Verify currency and authenticity via CrossMark

Cite this chapter

Goldstein, J.I., Newbury, D.E., Michael, J.R., Ritchie, N.W.M., Scott, J.H.J., Joy, D.C. (2018). Energy Dispersive X-ray Spectrometry: Physical Principles and User-Selected Parameters. In: Scanning Electron Microscopy and X-Ray Microanalysis. Springer, New York, NY. https://doi.org/10.1007/978-1-4939-6676-9_16

Download citation

Publish with us

Policies and ethics