Advertisement

Catheter Velocimeters

  • Michael D. Stern
Part of the Developments in Cardiovascular Medicine book series (DICM, volume 107)

Abstract

The laser-Doppler velocimeter is potentially a very powerful device for measuring intravascular blood flow velocity. Because light can be delivered and collected through optical fibers of very small diameter, the technique lends itself naturally to use through a catheter passed through the blood vessel. The small size of optical fibers also permits high spatial resolution, so that flow velocity at a single point in the cross section of a blood vessel can be measured. However, because catheters disturb the flow of blood in their neighborhood, and because light is multiply scattered in passing through blood, combining the high spatial and temporal resolution of (LDV) laserDoppler velocimetry with catheter delivery is not straightforward. At the present time, the development of instruments for the practical measurement of blood flow velocity through a catheter is an area of active research. In this chapter, we will present the principles behind this work, and will indicate the present state of the art and the potential uses of catheter LDV systems.

Keywords

Doppler Shift Single Fiber Laser Doppler Anemometry Doppler Spectrum Bare Fiber 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Tanaka, T., and G.B. Benedek. 1975. Measurement of the velocity of blood flow (in vivo) using a fiber optic catheter and optical mixing spectroscopy. Appl Optics 14:189–196.Google Scholar
  2. 2.
    Kilpatrick, D., T. Linderer, R. E. Sievers, and J.V. Tyberg. 1982. Measurement of coronary sinus blood flow by fiber-optic laser Doppler anemometry. Am J Physiol 242:H1111–H1114.PubMedGoogle Scholar
  3. 3.
    Kilpatrick, D., J.V. Tyberg, and W.W. Parmley. 1982. Blood velocity measurement by fiber optic laser Doppler anemometry. IEEE Trans Biomed Eng 29:142–145.PubMedCrossRefGoogle Scholar
  4. 4.
    Nishihara, H., J. Koyama, N. Hoki, F. Kajiya, M. Hironaga, and M. Kano. 1982. Opticalfiber laser Doppler velocimeter for high-resolution measurement of pulsatile blood flows. Appl Optics 21:1785–1790.CrossRefGoogle Scholar
  5. 5.
    Kajiya, F., G. Tomonaga, K. Tsujioka, Y. Ogasawara, and H. Nishihara. 1985. Evaluation of local blood flow velocity in proximal and distal coronary arteries by laser Doppler method. J Biomed Eng 107:10–14.Google Scholar
  6. 6.
    Stern, M.D. 1985. Laser Doppler velocimetry in blood and multiply scattering fluids: theory. Appl Optics 24:1968–1986.CrossRefGoogle Scholar
  7. 7.
    Kajiya, F., O. Hiramatsu, Y. Ogasawara, K. Mito, and K. Tsujioka. 1988. Dual-fiber laser Doppler velocimeter and its application to the measurement of coronary blood velocity. Biorheology 25:227–235.PubMedGoogle Scholar