Advertisement

Galvanic-Coulometric Detectors in Gas Chromatography

  • Paul Hersch

Abstract

Among the most sensitive detectors for gas chromatography are those based on coulometry. They owe their sensitivity to the high value of the faraday and to the ease with which minute electric currents are measurable. Electrometers with a discrimination for less than 10-15 amp are available, a current corresponding to about 10-20 equivalents per sec, or 6000 univalent ions per sec. For example, take anthracene, C14H10. A coulometrically quantitative oxidation of one molecule means stripping it of 14 × 4 + 10 × 1 = 66 electrons. Thus, a stream of fewer than 100 molecules per sec of this compound is detectable in principle. In the case of compounds with more carbon and hydrogen atoms in the molecule, the sensitivity in these terms is higher still, compensating for their lesser volatility.

Keywords

Cavity Cell Hydrogen Halide Electrolytic Manganese Dioxide Manganous Oxide Loop Cell 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    A.P. Altshuiler and I.R. Cohen, “Application of Diffusion Cells to the Production of Known Concentrations of Gaseous Hydrocarbons,” Anal. Chem. 32:802, 1960.CrossRefGoogle Scholar
  2. 2.
    Anonymous “CO on the Highway,” Sci. Am., pp. 52 and 57, May, 1965.Google Scholar
  3. 3.
    R.R. Austin, “An Olefin Recorder for Atmospheric Monitoring Service,” Trans. Instr. Soc. Am. 1:211, 1962.Google Scholar
  4. 4.
    Beckman Instruments, “Sulfur Dioxide Analyzer,” Model 906, Bulletin 4072, 1966.Google Scholar
  5. 5.
    R. Belcher, L. Ottendorfer, and T.S. West, “Acid Chlorides of Substituted Succinic and Glutaric Acids as Hydrolytic Reagents for the Determination of Water,” Talanta 4:166, 1960.CrossRefGoogle Scholar
  6. 6.
    A. Berton, “Application des Osmopiles Galvaniques à la Détection et au Dosage de Produits Toxiques Volatils,” in: J.A.Gautier,ed., Mises au Point de Chimie Pure et Appliquée et d’Analyse Bromatologique, 10e Série, Masson et Cie, Paris, 1962, pp. 6–22.Google Scholar
  7. 7.
    A. Berton, “Gas Chromatography and Selective Galvanic Detection,” Chim. Anal. 45:585, 1963.Google Scholar
  8. 8.
    H.P. Burchfield, D.E. Johnson, J.W. Rhoades, and R.J. Wheeler, “Selective Detection Systems in the Analysis of Drugs and Pesticides,” in: Lectures on Gas Chromatography-1964, ed.Mattick and Szymanski, Plenum Press, New York, 1964, pp. 129–146.Google Scholar
  9. 9.
    K. Cammann, “Analytical Fuel Cells,” Z. Anal. Chemie 216:294, 1965.Google Scholar
  10. 10.
    E.R. Gilliland, “Diffusion Coefficients in Gaseous Systems,” Ind. Eng. Chem. 26:681. 1934.CrossRefGoogle Scholar
  11. 11.
    M. Guillot and A. Berton, “Analysis of Chlorinated Solvents by Gas Chromatography at Low Temperature with Electrochemical Detection,” Comp. Rend. 250:1857, 1960;Google Scholar
  12. 11a.
    M. Guillot and A. Berton, “Analysis of Chlorinated Solvents by Gas Chromatography at Low Temperature with Electrochemical Detection,” Chem. Abstr. 55:12156f, 1961.Google Scholar
  13. 12.
    A.J. Haagen-Smit and T.W. Latham, “Carbon Monoxide Levels in City Driving,” Clean Air Quart. 8:(4)8, 1964.Google Scholar
  14. 13.
    P. Hersch, “Trace Monitoring in Gases Using Galvanic Systems,” Anal. Chem. 32:1030, 1960.CrossRefGoogle Scholar
  15. 14.
    P. Hersch, Gas Chromatographic Detector, Brit. Pat. 880, 965, filed February, 1959, published October, 1961.Google Scholar
  16. 15.
    P. Hersch and R. Deuringer, “Galvanic Monitoring of Ozone in Air,” Anal. Chem. 35:897, 1963.CrossRefGoogle Scholar
  17. 16.
    P. Hersch and C. J. Sambucetti, “Galvanic Coulometric Monitoring of Carbon Monoxide,” Paper at Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, 1963.Google Scholar
  18. 17.
    P. Hersch, C.J. Sambucetti, and R. Deuringer, “Electrolytic Calibration of Gas Monitors, Analysis Instrumentation, 1963,” Proceedings of the Ninth National ISA Analysis Instrumentation Symposium, Houston, Texas, Plenum Press, New York, 1963, pp. 65–71; Chim. Anal. 46:31, 1964.Google Scholar
  19. 18.
    P. Hersch and C.J. Sambucetti, “A Galvanic-Coulometric Monitor of Acid Vapors,” Paper at Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, 1964.Google Scholar
  20. 19.
    P. Hersch, “Galvanic Analysis,” in Advan. Anal. Chem. Instr. Vol. 3 (C.N. Reilly, ed.), Wiley, New York, 1964, p 183–249.Google Scholar
  21. 20.
    P. Hersch and R. Deuringer, “Galvanic Monitoring of Moisture in Gases,” Paper at Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy, 1965.Google Scholar
  22. 21.
    P. Hersch and R. Deuringer, “Galvanic-Coulometric Monitoring of Sulfur Dioxide in Air,” Paper at 149th National Meeting of the American Chemical Society, Detroit, 1965.Google Scholar
  23. 22.
    P. Hersch, “Method and Means for Oxygen Analysis of Gases,” U. S. Pat. 3,223,597, issued December 14, 1965; Brit. Pat. 913,473, filed March, 1960, published 1962; German Pat. 1, 191, 984.Google Scholar
  24. 23.
    G.E. Hillman and J. Lightwood, “Determination of Small Amounts of Oxygen Using a Hersch Cell as a Gas Chromatography Detector,” Anal. Chem. 38:1430, 1966.CrossRefGoogle Scholar
  25. 24.
    W.W. Jacobi, “Batteries, Secondary,” in: Encyclopedia of Chemical Technology, Vol. 3, ed. Kirk-Othmen, Wiley, New York, 1963, p. 197.Google Scholar
  26. 25.
    V.A. Klimova and G.F. Anisimova. Izv. Akad. Nauk USSR. Otd. Khim. Nauk 9:2088, 1961;Google Scholar
  27. 25a. V.A. Klimova and G.F. Anisimova. Chem. Abstr. 56:8003g, 1962.Google Scholar
  28. 26.
    R.L. Martin, “Fast and Sensitive Method for Determination of Nitrogen,” Anal. Chem. 38:1209, 1966.CrossRefGoogle Scholar
  29. 27.
    T.R. Phillips, E.G. Johnson, and H. Woodward, “The Use of a Hersch Cell as a Detector in Gas Chromatography,” Anal. Chem. 36:450, 1964.CrossRefGoogle Scholar
  30. 28.
    M. Picon, “The Extreme Sensitivity of the Determination of Carbon Monoxide, Ethylene, and Acetylene by Iodic Anhydride,” Mém. Soc. Chim. France, Sér. 5, p. 370, 1956; Chem. Abstr. 50:7664a.Google Scholar

Copyright information

© Plenum Press 1967

Authors and Affiliations

  • Paul Hersch
    • 1
  1. 1.Gould-National Batteries, Inc.MinneapolisUSA

Personalised recommendations