Non-Invasive Measurements in Hyperthermia: Radiometry and Previsional Thermometry

  • G. L. Sannazzari
  • P. Gabbriele
  • E. Madon
Part of the Advances in Experimental Medicine and Biology book series (AEMB, volume 267)

Abstract

The development of Hyperthermia as a therapeutic method is closely related to the possibilty to control and quantify hyperthermic treatment in the portion of a biological body heated. For this purpose, the thermometer would have resolution and precision between 0.1°C and 0.2°C, because of the great dependence of the biological response to the temperature increase; suitable spatial resolution, between 5 and 10 mm, cause of the necessity to work in large temperature gradients; ability to follow the temporal variations of temperature with a temporal resolution of 1 second and a dimensions below 16 gauge to minimize trauma for patients. Actually the thermometry system used in the routine consist of thermistors, thermocouples and optical fibers. All these system are invasive: they give punctual temperature measurements in the whole tissue area and are traumatic for the patient.To avoid these disadvantages in our institute two non invasive methods have been tested: they’re a microwave radiometer, which measures the emitted intensity of bodies, that is directly related to their temperature and a simulation program, which calculates the temperature behavior of an heated tissue by the solution of the bio-heat equation. The results of our tests show that the thermometric technique, based on radiometer, gives temperature measurements that are to be considered as a mean among the ones obtained in the whole tissue area and not global at all; on the other hand the calculation of the heat produced by metabolism and over all of the heat taken away by the blood flow. With the intent to verify the goodness of the program we have simulated 22 hyperthermic treatments comparing the real temperature, obtained by optical fibers inserted in the tissue area during heating, with the simulated one. In parallel it was made a comparison between invasive measurements, with optical fibers, and non invasive ones, revealed by radiometer.

Keywords

Simulation Program Good Correspondence Tissue Area Microwave Radiometer Heated Tissue 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Reference

  1. 1).
    Audone B, Bolla L, Delogou A and coll: Dosimetry and applicators in RF and microwave hyperthermia. Strahlentherapie: 159–367, 1983.Google Scholar
  2. 2).
    Audone B, Bolla L, and coll: Microwave and RF computerized system for hyperthermic oncology. Proc 4Th Int Symp on Hyp Onc: 715–717, Aahrus, 1984.Google Scholar
  3. 3).
    Cerri G, De Leo R: Waveguide applicators for biomedical application: an accurate analysis of power coupling in homogeneous biomedical media. Int Wroc Symp on El Comp, 1984.Google Scholar
  4. 4).
    Hand JW: Thermometry in Hyperthermia. Prooc 4Th Int Symp on Hyp Onc: 299–308, Aahrus, 1984.Google Scholar
  5. 5).
    Plancot M, Prevost B, Chive M and coll: a new method for thermal dosimetry in microwave hyperthermia using microwave radiometry for temperature control. Int J Hyp 1, 9–19, 1984Google Scholar
  6. 6).
    Strohben JW, Roemer RB: A survey of computer simulation of hyperthermia treatment. IEEE Trans, vol BME-31: 136–149, 1984.Google Scholar

Copyright information

© Plenum Press, New York 1990

Authors and Affiliations

  • G. L. Sannazzari
    • 1
  • P. Gabbriele
    • 1
  • E. Madon
    • 1
  1. 1.Radiology Institute, Radiotherapy DivisionUniversity of TurinItaly

Personalised recommendations