Backscattered Electron Imaging

  • Charles E. Lyman
  • Joseph I. Goldstein
  • Alton D. RomigJr.
  • Patrick Echlin
  • David C. Joy
  • Dale E. Newbury
  • David B. Williams
  • John T. Armstrong
  • Charles E. Fiori
  • Eric Lifshin
  • Klaus-Ruediger Peters

Abstract

Experiment 8.1: E-T Detector Collection Efficiency. The solid angle and efficiency of a specific E-T detector for direct collection of BSEs is:
  1. (a)

    Area, A, of scintillator (cm2) = 1.5 cm2.

     
  2. (b)

    Distance, r, from specimen to scintillator (cm) = 4 cm.

     
  3. (c)

    Solid angle, Ω = A/r2 = 1.5/16 = 0.094 steradians.

     
  4. (e)

    For a specimen set normal to the beam (0° tilt), approximate take-off angle = 45°.

     

Keywords

Solid Angle Backscatter Electron Image Difference Mode Raney Nickel Superplastic Alloy 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Copyright information

© Plenum Press, New York 1990

Authors and Affiliations

  • Charles E. Lyman
    • 1
  • Joseph I. Goldstein
    • 1
  • Alton D. RomigJr.
    • 2
  • Patrick Echlin
    • 3
  • David C. Joy
    • 4
  • Dale E. Newbury
    • 5
  • David B. Williams
    • 1
  • John T. Armstrong
    • 6
  • Charles E. Fiori
    • 5
  • Eric Lifshin
    • 7
  • Klaus-Ruediger Peters
    • 8
  1. 1.Lehigh UniversityBethlehemUSA
  2. 2.Sandia National LaboratoriesAlbuquerqueUSA
  3. 3.University of CambridgeCambridgeEngland
  4. 4.University of TennesseeKnoxvilleUSA
  5. 5.National Institute of Standards and TechnologyGaithersburgUSA
  6. 6.California Institute of TechnologyPasadenaUSA
  7. 7.GE Corporate Research and DevelopmentSchenectadyUSA
  8. 8.The University of Connecticut Health CenterFarmingtonUSA

Personalised recommendations