Micro Total Analysis Systems 2002 pp 73-75 | Cite as
Optimal Design of the Micromixer using Nonhomogeneous Multilayer Laminar Flow
Conference paper
Abstract
An efficient micromixer using nonhomogeneous multilayer in the laminar flow regime is described. In order to increase the mixing efficiency, the CFD simulation was used to optimize the width of the multilayers. It is estimated that the reagents mixing is proceeded more than 10 times faster compared to previous multilayer micromixers. The optimized micromixer was fabricated, and its mixing efficiency was evaluated.
keywords
micromixer multilayer laminar flow diffusion mixing CFD simulationPreview
Unable to display preview. Download preview PDF.
References
- [1]A. Manz, F. Bessoth and M. U. Kopp, MictoTAS’98, pp235–240Google Scholar
- [2]T. M. Floyd, M. A. Schmidt and K. F. Jensen, MicroTAS’Ol, pp. 277–279Google Scholar
- [3]K. B. Greiner, et. al, MicroTAS’OO, pp. 87–90Google Scholar
Copyright information