Three patients with normal-pressure hydrocephalus, who were treated by installation of a low-pressure ventriculo-peritoneal (V-P) shunt system equipped with anti-siphon device, did not show improvement of clinical signs and ventricular size on CT. While, the lumbar infusion manometric test in horizontal patient position showed good functioning of the shunting system in all 3 patients. However, the clinical signs and ventricular size on CT were remarkably improved by replacing the shunt system to the low pressure V-P shunt system which was not equipped with anti-siphon device. The shunt malfunction seemed to be caused by the anti-siphon function itself of the previous shunt system which was equipped with anti-siphon device, especially when the patients were kept sitting and/or standing during the daily activity.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Similar content being viewed by others
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 1990 Springer-Verlag Berlin Heidelberg
About this chapter
Cite this chapter
Seida, M., Ito, U., Tomida, S., Yamazaki, S., Inaba, Y. (1990). Malfunction of V-P Shunt System Equipped with Antisiphon Device in Normal-pressure Hydrocephalus. In: Matsumoto, S., Sato, K., Tamaki, N., Oi, S. (eds) Annual Review of Hydrocephalus. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-662-11152-9_54
Download citation
DOI: https://doi.org/10.1007/978-3-662-11152-9_54
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-662-11154-3
Online ISBN: 978-3-662-11152-9
eBook Packages: Springer Book Archive