Advertisement

Possibility to Use Finapres Signal for Augmentation Index Estimation

  • K. Pilt
  • K. Meigas
  • M. Viigimaa
  • K. Temitski
Part of the IFMBE Proceedings book series (IFMBE, volume 34)

Abstract

The possible method for augmentation index estimation from a Finapres signal is described. The experiments were carried out on volunteers. The augmentation indices were calculated from the Finapres signal by using the proposed method and compared with the Sphygmocor reference device. As result the correlation between Sphygmocor and Finapres augmentation indices was found to be r=0.82 (p<0.001) and the regression model was constructed.

Keywords

Augmentation index pressure wave derivatives Sphygmocor Finapres 