The Influence of Patient Miscentering on Patient Dose and Image Noise in Two Commercial ct Scanners

  • M. A. Habibzadeh
  • M. R. Ay
  • A. R. Kamali asl
  • H. Ghadiri
  • H. Zaidi
Part of the IFMBE Proceedings book series (IFMBE, volume 29)

Abstract

The clinical influences of patient miscentering on patient dose and image noise were investigated for two models of commercial CT scanners. Several phantoms were scanned on 4-slice GE Lightspeed and 64-slice GE Lightspeed VCT. Regression models of surface dose and image noise were generated as a function of phantom size and the value of miscentering. 64 scout images of patients from the first scanner and 113 from the second one were analyzed to assess the possible amount of increasing in dose and noise. For the first scanner the average amount of miscentering was 3 cm below the isocenter which leads to 25.8% increase in dose and 8.3% increase in noise. These values for the second scanner were 1.6 cm below the center, 19.8% and 6.2%, respectively. The results clearly demonstrate that patient miscentering may substantially increase dose and image noise. Therefore, technologists are strongly encouraged to pay greater attention to patient centering.

Keywords

Bowtie Filter Computed Tomography Dose Noise 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Copyright information

© International Federation for Medical and Biological Engineering 2010

Authors and Affiliations

  • M. A. Habibzadeh
    • 1
    • 2
  • M. R. Ay
    • 2
    • 3
    • 4
  • A. R. Kamali asl
    • 1
  • H. Ghadiri
    • 4
    • 5
  • H. Zaidi
    • 6
    • 7
  1. 1.Faculty of Medical Radiation EngineeringShahid Beheshti UniversityTehranIran
  2. 2.Research Center for Science and Technology in MedicineTehran University of Medical SciencesTehranIran
  3. 3.Department of Medical Physics and Biomedical EngineeringTehran University of Medical SciencesTehranIran
  4. 4.Research Institute for Nuclear MedicineTehran University of Medical SciencesTehranIran
  5. 5.Department of Medical PhysicsIran University of Medical SciencesTehranIran
  6. 6.Division of Nuclear MedicineGeneva University HospitalGenevaSwitzerland
  7. 7.Geneva Neuroscience CenterGeneva UniversityGenevaSwitzerland

Personalised recommendations