Development and experimental implementation of new approaches to measure MTF and NPS

  • C. Brunner
  • A. Schegerer
  • B. Renger
  • C. Hoeschen
Conference paper
Part of the IFMBE Proceedings book series (IFMBE, volume 25/2)

Abstract

In order to compare novel CT-geometries with existing ones, it is necessary to revise conventional quality assurance methods or develop new approaches. In a first step, this has been done for a determination of the Modulation Transfer Function (MTF) and the Noise Power Spectrum (NPS) parallel for reconstructed images and raw data. The analysis of the raw data received from a scan of a tungsten wire allowed to calculate the MTF of the detecting system independent of the reconstruction algorithm. Another new approach allowed the determination of the NPS out of the raw data independent of the number of slices of the detector.

Keywords

Modulation Transfer Function Noise Power Spectrum novel CT-methods 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Copyright information

© International Federation for Medical and Biological Engineering 2009

Authors and Affiliations

  • C. Brunner
    • 1
  • A. Schegerer
    • 1
  • B. Renger
    • 1
    • 2
  • C. Hoeschen
    • 1
  1. 1.Helmholtz Zentrum MünchenInstitute of Radiation ProtectionNeuherbergGermany
  2. 2.Klinikum rechts der IsarInstitute of Radiation DiagnosticsMünchenGermany

Personalised recommendations