Abstract
To verify the fluence distribution and number of monitor units for dynamic treatment techniques each patient treatment plan is applied to a measuring device before first treatment. We verify Rapid Arc treatment plans patient oriented using the IMRT-MatriXX (IBA, Schwarzenbruck, Germany). Therefore we generated a verification protocol. 3 patients with cerebral metastasis, 2 patients with head and neck cancer, and 5 patients with prostate cancer receiving Rapid Arc radiotherapy were included in this study. It is poss-ible to verify the 2D fluence distribution and the number of monitor units of Rapid Arc treatment plans using IMRT-MatriXX within a few minutes. Therefore we recommend the use of the IMRT-MatriXX for Rapid Arc treatment plans in the clinical routine.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2009 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Wagner, D.M., Vorwerk, H. (2009). First experience with patient oriented Rapid Arc treatment plan verification. In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/1. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03474-9_95