Abstract
With its newest linear accelerator, the ARTISTETM, Siemens Healthcare also has introduced the 160 MLCTM. Because of the specific design of the 160 MLCTM featuring tilted leaves, the exact position of the linac’s focal spot plays an important role in the leakage performance of the MLC. Upon commissioning of a dual photon energy ARTISTETM, capable of generating 6 MV and 18 MV photon treatment beams, we have performed several measurements regarding leakage of the MLC for both energies. The measurements were repeated after adjustments to beam alignment and compared to the previous ones.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2009 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Klüter, S., Sroka-Perez, G., Schubert, K., Debus, J. (2009). Impact of Beam Alignment on Leakage of the Siemens 160 MLCTM . In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/1. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03474-9_139