  • Jean-Michel Arnal
Chapter

Abstract

In pressure support mode, flow curve informs about the patient effort and the patient-ventilator synchrony. Observation of flow and pressure curves at the onset of inspiration assesses the patient-ventilator synchrony to trigger the breath and help to optimize inspiratory trigger setting. Observation of the initial increase in flow informs about the setting of pressure rise time. Flow during insufflation informs about the patient’s effort and the adequacy of cycling setting. Observation of pressure and flow at the end of insufflation helps to optimize the expiratory trigger sensitivity. In volume assist control mode, pressure curve informs about the adequacy of inspiratory flow setting.

Supplementary material

Video 4.1

PS (AVI 13838 kb)

Video 4.2

Ineffective effort in PS (AVI 347930 kb)

Video 4.3

Autotriggering in PS (AVI 235030 kb)

Video 4.4

Double trigger in PS (AVI 255763 kb)

Video 4.5

Expiratory trigger sensitivity (AVI 576042 kb)

Video 4.6

Volume assist control (AVI 317975 kb)

Suggested Readings

  1. Aslanian P, El Atrous S, et al. Effects of flow triggering on breathing effort during partial ventilatory support. Am J Respir Crit Care Med. 1998;157:135–43.CrossRefPubMedGoogle Scholar
  2. Chiumello D, Pelosi P, et al. Effect of different inspiratory rise time and cycling off criteria during pressure support ventilation in patients recovering from acute lung injury. Crit Care Med. 2003;31:2604–10.CrossRefPubMedGoogle Scholar
  3. Dhand R. Ventilator graphics and respiratory mechanics in the patient with obstructive lung disease. Respir Care. 2005;50:246–61.PubMedGoogle Scholar
  4. Du HL, Yamada Y. Expiratory asynchrony. Respir Care Clin N Am. 2005;11:265–80.CrossRefPubMedGoogle Scholar
  5. Fernandez R, Mendez M et al (1999) Effect of ventilator flow rate on respiratory timing in normal humans. Am J Respir Crit Care Med 159:710–9.Google Scholar
  6. Georgopoulos D, Prinianakis G, et al. Bedside waveforms interpretation as a tool to identify patient-ventilator asynchronies. Intensive Care Med. 2006;32:34–47.CrossRefPubMedGoogle Scholar
  7. Grinnan DC, Truwit JD. Clinical review: respiratory mechanics in spontaneous and assisted ventilation. Crit Care. 2005;9:472–84.CrossRefPubMedPubMedCentralGoogle Scholar
  8. Hess DR. Ventilator waveforms and the physiology of pressure support ventilation. Respir Care. 2005;50:166–86.PubMedGoogle Scholar
  9. Hoff FC, Tucci MR, et al. Cycling-off modes during pressure support ventilation: effects on breathing pattern, patient effort, and comfort. J Crit Care. 2014;29:380–5.CrossRefPubMedGoogle Scholar
  10. Kondili E, Prinianakis G, et al. Patient-ventilator interaction. Br J Anaesth. 2003;91:106–19.CrossRefPubMedGoogle Scholar
  11. Mojoli F, Iotti GA, et al. Is the ventilator switching from inspiration to expiration at the right time? Look at waveforms! Intensive Care Med. 2016;42:914–5.CrossRefPubMedGoogle Scholar
  12. Nilsestuen JO, Hargett KD. Using ventilator graphics to identify patient-ventilator asynchrony. Respir Care. 2005;50:202–34.PubMedGoogle Scholar
  13. Parthasarathy S, Jubran A, et al. Cycling of inspiratory and expiratory muscle groups with the ventilator in airflow limitation. Am J Respir Crit Care Med. 1998;158:1471–8.CrossRefPubMedGoogle Scholar
  14. Prinianakis G, Kondili E, et al. Effects of the flow waveform method of triggering and cycling on patient-ventilator interaction during pressure support. Intensive Care Med. 2003;29:1950–9.CrossRefPubMedGoogle Scholar
  15. Prinianakis G, Plataki M, et al. Effects of relaxation of inspiratory muscles on ventilator pressure during pressure support. Intensive Care Med. 2008;34:70–4.CrossRefPubMedGoogle Scholar
  16. Racca F, Squadrone V, et al. Patient-ventilator interaction during the triggering phase. Respir Care Clin N Am. 2005;11:225–45.CrossRefPubMedGoogle Scholar
  17. Tassaux D, Gainnier M, et al. Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med. 2005;172:1283–9.CrossRefPubMedGoogle Scholar
  18. Thille AW, Rodriguez P, et al. Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med. 2006;32:1515–22.CrossRefPubMedGoogle Scholar
  19. Thille AW, Cabello B, et al. Reduction of patient-ventilator asynchrony by reducing tidal volume during pressure-support ventilation. Intensive Care Med. 2008;34:1477–86.CrossRefPubMedGoogle Scholar
  20. Tobin MJ, Jubran A, et al. Patient-ventilator interaction. Am J Respir Crit Care Med. 2001;163:1059–63.CrossRefPubMedGoogle Scholar
  21. Tusman G, Suarez-Sipmann F, et al. Pulmonary blood flow generates cardiogenic oscillations. Respir Physiol Neurobiol. 2009;167:247–54.CrossRefPubMedGoogle Scholar
  22. Yamada Y, Du HL. Analysis of the mechanisms of expiratory asynchrony in pressure support ventilation: a mathematical approach. J Appl Physiol. 2000;88:2143–50.CrossRefPubMedGoogle Scholar

Copyright information

© Springer International Publishing AG 2018

Authors and Affiliations

  • Jean-Michel Arnal
    • 1
    • 2
  1. 1.Service de Réanimation PolyvalenteHopital Sainte MusseToulonFrance
  2. 2.Applied Research and New TechnologyHamilton Medical AGBonaduzSwitzerland

Personalised recommendations