Abstract

A nation-wide survey of patient dose data was carried out, using data recorded in 2013–2015. Data from 32,000 women were collected. The average dose for oblique views, for DR systems, was 1.65 mGy for all women, and 1.35 mGy for 50–60 mm breasts. There was a wide range of doses for different systems, with the highest more than twice the dose of the lowest (2.03 mGy and 0.91 mGy respectively for the Hologic Dimensions and Philips MicroDose L30 systems, averaged over all breasts). Image quality, as indicated by the threshold gold thickness for 0.25 mm details, was better (0.21 µm) for the Hologic systems; for all the others it was practically the same (0.28 µm), although their doses to the average breast varied over a wide range.

Keywords

Mammography Patient dose DR systems 

Notes

Acknowledgments

The data collection was conducted under the auspices of the UK National Breast Screening Quality Assurance Coordinating Group for Physics. The authors are grateful to the NHSBSP radiographers who recorded data in the breast screening centres, and the physicists involved in the NHSBSP who provided technical data relating to the systems.

References

  1. 1.
    Dance, D.R., Skinner, C.L., Young, K.C., Beckett, J.R., Kotre, C.J.: Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys. Med. Biol. 45, 3225–3240 (2000)CrossRefGoogle Scholar
  2. 2.
    CDCOM version 1.6 software. EUREF. www.euref.org
  3. 3.
    CDMAM analysis UK version 1.4 software, NCCPM, Guildford, UK. www.nccpm.org
  4. 4.
    Kulama, E., Burch, A., Castellano, I., Lawinski, C.P., Marshall, N., Young, K.C.: Commissioning and Routine Testing of Full Field Digital Mammography Systems (NHSBSP Equipment report 0604). Sheffield: NHS Cancer Screening Programmes (2009)Google Scholar
  5. 5.
    Oduko, J.M., Young, K.C., Burch, A.: A survey of patient doses from digital mammography systems in the UK in 2007 to 2009. In: Martí, J., Oliver, A., Freixenet, J., Martí, R. (eds.) IWDM 2010. LNCS, vol. 6136, pp. 365–370. Springer, Heidelberg (2010)CrossRefGoogle Scholar
  6. 6.
    Young, K.C., Oduko, J.M.: Radiation doses received in the United Kingdom breast screening programme in 2010 to 2012. Br. J. Radiol. 89, 01508312 (2016)CrossRefGoogle Scholar
  7. 7.
    Strudley, C.J., Young, K.C.: Technical Evaluation of Philips MicroDose SI digital mammography system (NHSBSP Equipment Report 1310). London: Public Health England, NHS Cancer Screening Programmes (2016)Google Scholar

Copyright information

© Springer International Publishing Switzerland 2016

Authors and Affiliations

  1. 1.National Coordinating Centre for the Physics of MammographyGuildfordUK
  2. 2.Department of PhysicsUniversity of SurreyGuildfordUK

Personalised recommendations