Simulation of Positron Emission Mammography Imaging with Pixelated CdTe.

Conference paper
Part of the Lecture Notes in Computer Science book series (LNCS, volume 9699)

Abstract

The Voxel Imaging PET (VIP) Pathfinder project presents a new approach for the design of nuclear medicine imaging diagnostic devices by using highly segmented pixel CdTe sensors. State-of-the-art PET devices are made from scintillator crystals which have an energy resolution of about 10 %. Because of their limited energy resolution, the scatter fraction is relatively large. Their limited spatial resolution will introduce a parallax error and deficiency in the depth of interaction (DOI) determination (> 5 mm FWHM) and therefore a wrong line of response (LOR). In this study, we present a design for a breast dedicated PET based on CdTe detectors which have an energy resolution of about 1 % and can be easily segmented into small voxels for optimal spatial resolution to provide a solution for these setbacks. This simulation study will assess the advantages of this design for breast imaging purposes.

Notes

Acknowledgements

This work has received funding from the European Research Council under the European Union’s Seventh Framework Programme (ERC Grant Agreement 250207, “VIP”). We also acknowledge the support from the Spanish MINECO under the Severo Ochoa excellence program (grant SO-2012-0234).

References

  1. 1.
    Hruska, C.B., O’Connor, M.K.: Nuclear imaging of the breast: translating achievements in instrumentation into clinical use. Med Phys. 40(5), 050901 (2013)CrossRefGoogle Scholar
  2. 2.
    Luo, W., et al.: Performance evaluation of a PEM scanner using the NEMA NU 4–2008 small animal PET standards. IEEE Trans. Nucl. Sci. 57, 94–103 (2010)CrossRefGoogle Scholar
  3. 3.
    MacDonald, L., et al.: Clinical imaging characteristics of the positron emission mammography camera: PEM Flex Solo II. J. Nucl. Med. 50, 1666–1675 (2009)CrossRefGoogle Scholar
  4. 4.
    Lerche, C.W., et al.: Dependency of energy-, position- and depth of interaction resolution on scintillation crystal coating and geometry. IEEE Trans. Nucl. Sci. 55, 1344–1351 (2008)CrossRefGoogle Scholar
  5. 5.
    Mikhaylova, E., et al.: Simulation of the expected performance of a seamless scanner for brain PET based on highly pixelated CdTe Detectors. IEEE Trans. Med. Imaging 33, 332–339 (2014)CrossRefGoogle Scholar
  6. 6.
    Ariño, G., et al.: Characterization of CdTe detector for use in PET. In: IEEE Conference Record on NSS & MIC, pp. 4598–4603 (2011)Google Scholar
  7. 7.
    De Lorenzo, G., et al.: Pixelated CdTe detectors to overcome intrinsic limitations of crystal based positron emission mammographs. J. Instrum. 8, C01030 (2013)Google Scholar
  8. 8.
    Calderón, Y.: Design, Development, and Modeling of a Compton Camera Tomographer Based on Room Temperature Solid State Pixel Detector. Ph.D. Thesis, Universitat Autònoma de Barcelona (UAB) (2014)Google Scholar
  9. 9.
    Arce, P., et al.: Gamos: a framework to do Geant4 simulations in different physics fields with an user-friendly interface. Nucl. Instr. Meth. Phys. Res. A 735, 304–313 (2014)CrossRefGoogle Scholar
  10. 10.
    Ariño, G., et al.: Energy and coincidence time resolution measurements of CdTe detectors for PET. J. Instrum. 8, C02015 (2013)CrossRefGoogle Scholar
  11. 11.
    Kolstein, M., et al.: Evaluation of list-mode ordered subset expectation maximization image reconstruction for pixelated solid-state compton gamma camera with large number of channels. J. Instrum. 9, C04034 (2014)CrossRefGoogle Scholar
  12. 12.
    NEMA standards publication NU 4–2008: Performance measurements for small animal positron emission tomographs. National Electrical Manufacturers Association, Rosslyn, VA (2008)Google Scholar
  13. 13.
    Furuta, M., et al.: Basic evaluation of a C-shaped breast PET scanner. In: IEEE NSS & MIC Conference Record, pp. 2548–2552. IEEE (2009)Google Scholar
  14. 14.
    Ferreira, C.S., et al.: Scatter and attenuation corrections for a PEM detector using list-mode OSEM. In: IEEE Conference Record on NSS & MIC, pp. 1–4 (2013)Google Scholar

Copyright information

© Springer International Publishing Switzerland 2016

Authors and Affiliations

  1. 1.Institut de Física D’Altes Energies (IFAE), The Barcelona Institute of Science and TechnologyBarcelonaSpain

Personalised recommendations