Advertisement

FPGA-Based Multi Cycle Parallel Architecture for Real-Time Processing in Ultrasound Applications

  • Valentino MeacciEmail author
  • Enrico Boni
  • Alessandro Dallai
  • Alessandro Ramalli
  • Monica Scaringella
  • Francesco Guidi
  • Dario Russo
  • Stefano Ricci
Conference paper
Part of the Lecture Notes in Electrical Engineering book series (LNEE, volume 573)

Abstract

A typical echograph transmits ultrasound signals and receives backscattered echoes through several tens of small transducers that compose the array of the probe. The system should process the acquired echoes in a very short time, typically few tens of ms, so that the medical doctor does not perceive any delay between the probe movements and the image update. This implies, especially when high-frame rate is required, the real-time processing of several GB of data per second: a huge constraint that is even emphasized in the new advanced ultrasound modalities investigated nowadays. In this paper, we present a Field Programmable Gate Array architecture which uses a combination of parallel and serial strategies capable to achieve a throughput rate of 3.5 GB/s in ultrasound data processing. The proposed architecture is implemented in the research ultrasound system ULA-OP 256. The performance of this approach is demonstrated by the real-time implementation of Plane Wave Vector Doppler, a novel investigation method that shows the blood velocity fields over a 2D section of an artery. Tests in a carotid artery of a volunteer are presented.

Notes

Acknowledgements

Authors would like to express their deep gratitude to Professor Piero Tortoli, for his guidance in this research work.

References

  1. 1.
    Davis, R.I., Grolleau, E.: Editorial note on Special issue on scheduling and timing analysis for advanced real-time systems. Real-Time Syst. 51(2), 125–127 (2015)CrossRefGoogle Scholar
  2. 2.
    Evans, D.H., McDicken, W.N.: Doppler Ultrasound: Physics, Instrumentation and Signal Processing, 2nd edn. Wiley, Chichester (2000)Google Scholar
  3. 3.
    Costa, A., De Gloria, A., Faraboschi, P., Olivieri, M.: A parallel architecture for the Color Doppler flow technique in ultrasound imaging. Microprocess. Microprogr. 38(1–5), 545–551 (1993)CrossRefGoogle Scholar
  4. 4.
    Birk, M., Guth, A., Zapf, M., Balzer, M., Ruiter, N., Hübner, M., Becker, J.: Acceleration of image reconstruction in 3D ultrasound computer tomography: an evaluation of CPU, GPU and FPGA computing. In: Proceedings of Conference on Design & Architectures for Signal & Image Processing (DASIP 2011), pp. 1–8Google Scholar
  5. 5.
    Boni, E., Yu, A., Freear, F., Jensen, J., Tortoli. P.: Ultrasound open platforms for next-generation imaging technique development. IEEE Trans. Ultrason. Ferroelect. Freq. Contr. 65(7), 1078–1092 (2018)CrossRefGoogle Scholar
  6. 6.
    Boni, E., Bassi, L., Dallai, A., Meacci, V., Ramalli, A., Scaringella, M., Guidi, F., Ricci, S., Tortoli, P.: Architecture of an ultrasound system for continuous real-time high frame rate imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control.  https://doi.org/10.1109/tuffc.2017.2727980CrossRefGoogle Scholar
  7. 7.
    Steinberg, B.D.: Digital beamforming in ultrasound. IEEE Trans. Ultrason. Ferroelect. Freq. Contr. 39, 716–721 (1992)MathSciNetCrossRefGoogle Scholar
  8. 8.
    Meacci, V., Bassi, L., Ricci, S., Boni, E., Tortoli, P.: High-performance FPGA architecture for multi-line beamforming in ultrasound applications. In: Proceedings of Euromicro Conference on Digital System Design (DSD 2016), pp. 584–590, Cyprus, August 2016Google Scholar
  9. 9.
    Jensen, J.A., Nikolov, S.I., Yu, A.C.H., Garcia, D.: Ultrasound vector flow imaging-Part I: sequential systems. IEEE Trans. Ultrason. Ferroelectr. Freq. Contr. 63(11), 1704–1721 (2016)Google Scholar
  10. 10.
    Jensen, J.A., Nikolov, S.I., Yu A.C.H., Garcia, D.: Ultrasound vector flow imaging-Part II: parallel systems. IEEE Trans. Ultrason. Ferroelectr. Freq.Contr. 63(11), 722–1732 (2016)CrossRefGoogle Scholar
  11. 11.
    Ricci, S., Bassi, L., Meacci, V., Ramalli, A., Boni, E., Tortoli, P.: Multi-line measurements of blood velocity vectors in real-time. In: Proceedings IEEE Ultrasonics Symposium, pp. 1–4, Sep. 2016Google Scholar
  12. 12.
    Ricci, S., Ramalli, A., Bassi, L., Boni, E., Tortoli, P.: Real-time blood velocity vector measurement over a 2D region. IEEE Trans. Ultrason. Ferroelect. Freq. Contr. 65(2), 201–209 (2018).  https://doi.org/10.1109/tuffc.2017.2781715CrossRefGoogle Scholar
  13. 13.
    Costa, A., De Franciscis, A., De Gloria, A., Faraboschi P., Olivieri, M.: Spectral estimation for 2-D Doppler ultrasound imaging in Electronics Letters, vol. 28, no. 23, pp. 2177–2179, 5 Nov. 1992CrossRefGoogle Scholar

Copyright information

© Springer Nature Switzerland AG 2019

Authors and Affiliations

  • Valentino Meacci
    • 1
    Email author
  • Enrico Boni
    • 1
  • Alessandro Dallai
    • 1
  • Alessandro Ramalli
    • 2
  • Monica Scaringella
    • 1
  • Francesco Guidi
    • 1
  • Dario Russo
    • 1
  • Stefano Ricci
    • 1
  1. 1.Department of Information EngineeringUniversity of FlorenceFlorenceItaly
  2. 2.Laboratory of Cardiovascular Imaging and Dynamics, Department of Cardiovascular SciencesKU LeuvenLouvainBelgium

Personalised recommendations