A Novel Interoperable Safety System for Improved Coordination and Communication in Cardiac Surgery

  • David ArneyEmail author
  • Geoffrey Rance
  • Srey Rithy
  • Julian M. Goldman
  • Marco A. Zenati
Conference paper
Part of the Lecture Notes in Computer Science book series (LNCS, volume 11041)

Abstract

During cardiac surgery there is an unmet need for safe transfer of responsibility for patient oxygenation back and forth from the anesthesia to the perfusion teams. Prior to cardiopulmonary bypass (CPB), lung ventilation is performed by the anesthesia machine ventilator and is the responsibility of the anesthesia team. During CPB, lung ventilation is halted and oxygenation is performed by the CPB oxygenator and perfusion team This recurrent transfer throughout the procedure introduces the rare but serious possibility of a “never event”, resulting in the patient’s lungs not being ventilated upon stopping the CPB and potentially leading to catastrophic hypoxemia. Monitors and alarms on the anesthesia and bypass machines would not be useful when the other device is operating so they are routinely put into a standby mode until needed. Consequently, in the event that the handoff is missed, there are no alarms to catch the situation. To solve this unmet need, we propose a novel interoperable, context-aware system capable of detecting and acting if this rare situation occurs. Our system is built on the open-source OpenICE framework, allowing it to seamlessly work with a variety of ventilator and bypass machines.

Keywords

Cardiopulmonary bypass Never event Interoperability Alarms Surgical safety systems 

References

  1. 1.
    Arney, D., Goldman, J.M., Whitehead, S.F., Lee, I.: Synchronizing an x-ray and anesthesia machine ventilator: a medical device interoperability case study. In: BIODEVICES 2009, pp. 52–60 (2009)Google Scholar
  2. 2.
    Arney, D., Goldman, J.M., Whitehead, S.F., Lee, I.: Improving patient safety with X-ray and anesthesia machine ventilator synchronization: a medical device interoperability case study. In: Fred, A., Filipe, J., Gamboa, H. (eds.) BIOSTEC 2009. CCIS, vol. 52, pp. 96–109. Springer, Heidelberg (2010).  https://doi.org/10.1007/978-3-642-11721-3_7CrossRefGoogle Scholar
  3. 3.
    Arney, D., Plourde, J., Goldman, J.: OpenICE medical device interoperability platform overview and requirement analysis. Biomed. Eng./Biomed. Tech. 63, 39–47 (2017)CrossRefGoogle Scholar
  4. 4.
    ASTM F2761–09(2013): Medical Devices and Medical Systems - Essential safety requirements for equipment comprising the patient-centric integrated clinical environment (ICE) - Part 1: General requirements and conceptual model. http://www.astm.org/Standards/F2761.htm
  5. 5.
    Conboy, H., et al.: Cognitive support during high-consequence episodes of care in cardiovascular surgery. In: 2017 IEEE Conference on Cognitive and Computational Aspects of Situation Management, pp. 1–3, March 2017Google Scholar
  6. 6.
    Pajic, M., Mangharam, R., Sokolsky, O., Arney, D., Goldman, J.M., Lee, I.: Model-driven safety analysis of closed-loop medical systems. IEEE Trans. Ind. Inform. 10(1), 3–16 (2014).  https://doi.org/10.1109/TII.2012.2226594CrossRefGoogle Scholar

Copyright information

© Springer Nature Switzerland AG 2018

Authors and Affiliations

  • David Arney
    • 1
    • 2
    Email author
  • Geoffrey Rance
    • 3
  • Srey Rithy
    • 3
  • Julian M. Goldman
    • 1
    • 2
  • Marco A. Zenati
    • 2
  1. 1.Massachusetts General HospitalBostonUSA
  2. 2.Harvard Medical SchoolBostonUSA
  3. 3.VA Boston Healthcare SystemBostonUSA

Personalised recommendations