Abstract

Energetic beam electrons stimulate the atoms of the specimen to emit “characteristic” X-ray photons with sharply defined energies that are specific to each atom species. The critical condition for generating characteristic X-rays is that the energy of the beam electron must exceed the electron binding energy, the critical ionization energy Ec, for the particular atom species and the K-, L-, M-, and/or N- atomic shell(s). For efficient excitation, the incident beam energy should be at least twice the critical excitation energy, E0 > 2 Ec. Characteristic X-rays can be used to identify and quantify the elements present within the interaction volume. Simultaneously, beam electrons generate bremsstrahlung, or braking radiation, which creates a continuous X-ray spectrum, the “X-ray continuum,” whose energies fill the range from the practical measurement threshold of 50 eV to the incident beam energy, E0. This continuous X-ray spectrum forms a spectral background beneath the characteristic X-rays which impacts accurate measurement of the characteristic X-rays and determines a finite concentration limit of detection. X-rays are generated throughout a large fraction of the electron interaction volume. The spatial resolution, lateral and in-depth, of electron-excited X-ray microanalysis can be roughly estimated with a modified Kanaya–Okayama range equation or much more completely described with Monte Carlo electron trajectory simulation. Because of their generation over a range of depth, X-rays must propagate through the specimen to reach the surface and are subject to photoelectric absorption which reduces the intensity at all photon energies, but particularly at low energies.

References

  1. Crawford J, Cohen D, Doherty G, Atanacio A (2011) Calculated K, L and M-shell X-ray line intensities for light ion impact on selected targets from Z = 6 to 100. ANSTO/E-774, Australian Nuclear Science and Technology Organisation (September, 2011)Google Scholar
  2. Drouin D, Couture A, Joly D, Tastet X, Aimez V, Gauvin R (2007) CASINO V2.42 – a fast and easy-to-use modeling tool for scanning electron microscopy and microanalysis users. Scanning 29:92–101CrossRefGoogle Scholar
  3. Fiori CE, Swyt CR, Myklebust RL (1992) Desktop Spectrum Analyzer (DTSA), a comprehensive software engine for electron-excited X-ray spectrometry. National Institute of Standards and Technology (NIST) – National Institutes of Health (NIH). United States Patent 5,299,138 (March 29, 1994) (Standard Reference Data Program, NIST, Gaithersburg, MD 20899)Google Scholar
  4. Heinrich KFJ, Fiori CE, Myklebust RL (1979) Relative transition probabilities for the X-ray lines from the K level. J Appl Phys 50:5589CrossRefGoogle Scholar
  5. Joy D (2006) Monte Carlo simulation. Available at ► http://www.lehigh.edu/~maw3/link/mssoft/mcsim.html
  6. Kramers H (1923) On the theory of X-ray absorption and the continuous X-ray spectrum. Phil Mag 46:836CrossRefGoogle Scholar
  7. Krause M, Oliver J (1979) Natural widths of atomic K and L levels, Kα X-ray lines and several KLL Auger lines. J Phys Chem Ref Data Monogr 8:329CrossRefGoogle Scholar
  8. Lifshin E, Ciccarelli MF, Bolon RB (1980) In: Beaman DR, Ogilvie RE, Wittry DB (eds) Proceedings 8th international conference on X-ray optics and microanalysis, vol 141. Pendell, MidlandGoogle Scholar
  9. Pouchou JL, Pichoir F (1991) Quantitative analysis of homogeneous or stratified microvolumes applying the model ‘PAP’. In: Heinrich KFJ, Newbury DE (eds) Electron probe quantitation, vol 31. Plenum Press, New YorkGoogle Scholar
  10. Powell C (1976) Use of Monte Carlo Calculations. In: Heinrich KFJ, Newbury DE, Yakowitz H (eds) Electron probe microanalysis and scanning electron microscopy. National Bureau of standard Special Publication 460. U.S. Government Printing Office, Washington, DC, 97Google Scholar
  11. Ritchie NWM (2015) NIST DTSA-II software, including tutorials, is available free at: ► www.cstl.nist.gov/div837/837.02/epq/dtsa2/index.html

Copyright information

© Springer Science+Business Media LLC 2018

Open Access This chapter is licensed under the terms of the Creative Commons Attribution-NonCommercial 2.5 International License (http://creativecommons.org/licenses/by-nc/2.5/), which permits any noncommercial use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license and indicate if changes were made.

The images or other third party material in this chapter are included in the chapter's Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the chapter's Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

Authors and Affiliations

  • Joseph I. Goldstein
    • 1
  • Dale E. Newbury
    • 2
  • Joseph R. Michael
    • 3
  • Nicholas W. M. Ritchie
    • 2
  • John Henry J. Scott
    • 2
  • David C. Joy
    • 4
  1. 1.University of MassachusettsAmherstUSA
  2. 2.National Institute of Standards and TechnologyGaithersburgUSA
  3. 3.Sandia National LaboratoriesAlbuquerqueUSA
  4. 4.University of TennesseeKnoxvilleUSA

Personalised recommendations