Alternative Bearing Surfaces: Alumina Ceramic Bearings for Total Hip Arthroplasty

  • W. N. Capello
  • J. A. D’Anthonio
  • J. R. Feinberg
  • M. T. Manley
Part of the Ceramics in Orthopaedics book series (CIO)

Abstract

Osteolysis resulting from polyethylene wear debris is one of the most common causes of implant failure in young, active individuais who undergo total hip arthroplasty. Reducing wear may help extend implant life in younger, more active patients. Contemporary alumina ceramic/alumina ceramic bearing articulations are harder, are scratch resistant, and are more hydrophilic than other bearing couples. This results in reduced wear and reduction of particle load to the surrounding tissue. Therefore, bearings made of alumina ceramic may be a preferable bearing choice for younger, more active patients.

To investigate this hypothesis, a total of 495 patients (514 hips), average age of 53 years, were enrolled in a prospective, randomized, multicenter study comparing an alumina-on-alumina ceramic bearing to a cobalt-chrome-on-polyethylene bearing control. At an average of four years post-implantation, no difference in clinical outcome was observed between groups. There were no ceramic head or liner fractures in this group, nor were there any revisions due to the ceramic liner.

Another investigational group was added to the study one year after enrollment in the original study was closed. The same inclusion/exclusion criteria were used. A total of 194 consecutive patients (1209 hips) received an alumina liner that included a thin metal backing designed to allow bearing replacement and ease operative assembly. At an average follow-up of 30 months, no liner or head chips or fractures were observed in this group.

Keywords

Bearing Surface Ceramic Head Ceramic Liner Acetabular Shell Acetabular Side 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Dumbleton J. H., Manley M. T., Edidin A. A., A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplasty 2002; 17:649–61.PubMedCrossRefGoogle Scholar
  2. 2.
    Boutin P., Christel P., Dorlot J.-M., et al: The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res 1988; 22:1203–1232.PubMedCrossRefGoogle Scholar
  3. 3.
    Mahoney O. M., Dimon J. H.: Unsatisfactory results with a ceramic total hip prosthesis. J Bone Joint Surg 1990; 72-A:663–671.Google Scholar
  4. 4.
    Knahr K., Bohler M., Frank P., Plenk H., Salzer M. Survival analysis of an uncemented ceramic acetabular component in total hip replacement. Arch Orthop Trauma Surg 1987; 106:523–531.CrossRefGoogle Scholar
  5. 5.
    Skinner H. B.: Ceramic bearing surfaces. Clin Orthop 1999; 369:83–91.PubMedCrossRefGoogle Scholar
  6. 6.
    Hoffinger S. A., Keggi K. J., Zatorski L. E. Primary ceramic hip replacement: A prospective study of 119 hips. Orthopedics 1991; 14:523–531.PubMedGoogle Scholar
  7. 7.
    Bierbaum B. E., Nairus J., Kuesis D., Morison J. C., Ward D: Ceramic-on-ceramic bearings in total hip arthroplasty. Clin Orthop 2002; 405:158–163.PubMedGoogle Scholar
  8. 8.
    Heros R. J., Willman G. Ceramics in total hip arthroplasty: History, mechanical properties, clinical results, and current manufacturing state of the art. Sem Arhroplasty 1998; 9:114–122.Google Scholar
  9. 9.
    Willmann G. Ceramic femoral head retrieval data. Clin Orthop 2000; 379:22–28.PubMedGoogle Scholar
  10. 10.
    Willmann G., von Chamier W. The improvements of the material properties of BIOLOX offer benefits for THR. In Puhl (ed): Bioceramics in Orthopedics: New Applications. Stuttgart, Enke 19–25, 1998.Google Scholar
  11. 11.
    Dorlot J. M. Long-term effects of alumina components in total hip prostheses. Clin Orthop 1992; 282:47–52.PubMedGoogle Scholar
  12. 12.
    Dorlot J. M., Christel P., Neunier A. Wear analysis of retrieved alumina heads and sockets of hip prostheses. J Biomed Mater Res 1989; 23(Supple A3):299–310.PubMedCrossRefGoogle Scholar
  13. 13.
    Cooper J. R., Dowson D., Fisher K., Jobbins B. Ceramic bearing surfaces in total artificial joints: resistance to third body wear damage from bone cement particles. J Med Eng Technol 1991; 15:63–67.PubMedGoogle Scholar
  14. 14.
    Harris W. H. Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty: An end result study using a new method of result evaluation. J Bone Joint Surg 1969; 51-A:737–755.Google Scholar
  15. 15.
    Engh C. A., Massin P., Suthers K. E. Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop 1990; 257:107–128.PubMedGoogle Scholar
  16. 16.
    Taylor S. K., Serekian P., Manley M. Wear performance of a contemporary alumina: alumina bearing couple under hip joint simulation. Trans. 44th Ann Mtg, Orthopaedic Research Society, New Orleans, 1998.Google Scholar
  17. 17.
    Clarke I. C. Role of ceramic implants: Design and clinical success with total hip prosthetic ceramic-on-ceramic bearings. Clin Orthop 1992; 282:19–30.PubMedGoogle Scholar
  18. 18.
    Callaway G. H., Flynn W., Ranawat C. Case report: Fracture of the femoral head after ceramic-on-polyethylene total hip arthroplasty. J Arthroplasty 1995; 10:855–859.PubMedCrossRefGoogle Scholar
  19. 19.
    D’Antonio J. A., Capello W. N., Manley M. T. Remodeling of bone around hydroxyapatite-coated femoral stems. J Bone Joint Surg 1996; 78A:1226–1234.Google Scholar

Copyright information

© Steinkopff Verlag Darmstadt 2005

Authors and Affiliations

  • W. N. Capello
  • J. A. D’Anthonio
  • J. R. Feinberg
  • M. T. Manley

There are no affiliations available

Personalised recommendations