Innovative non-invasive method for absolute intracranial pressure measurement without calibration

  • A. Ragauskas
  • G. Daubaris
  • A. Dziugys
  • V. Azelis
  • V. Gedrimas
Part of the Acta Neurochirurgica Supplementum book series (NEUROCHIRURGICA, volume 95)

Summary

A new absolute ICP (aICP) measurement method was designed which does not need calibration. In this study we compared a new method with invasive aICP method in ICU on the patients with closed severe traumatic brain injury.

A new method is based on two-depth TCD technique for aICP and external absolute pressure aPe comparison using the eye artery (EA) as natural “balance”. The intracranial segment of EA is compressed by aICP and the extracranial segment is compressed by aPe applied to the tissues surrounding the eye. The blood flow parameters in both EA segments are approximately the same when aPe=aICP. Two-depth TCD device is used as an indicator of balance aPe=aICP when the pulsatility index of blood flow velocity waveform in intracranial and extracranial segments are the same.

Fifty seven simultaneous invasive and non-invasive aICP measurements were performed in aICP range from 3.0 to 37.0 mmHg. Bland Altman plot of the differences between simultaneous invasive and non-invasive aICP measurements shows the negligible mean difference (mean=0.94 mmHg) with a standard deviation of 6.18 mmHg.

This validation study shows that it is possible to measure aICP non-invasively without calibration of the system with 95% confidence interval of 12 mmHg.

Keywords

Non-invasive absolute intracranial pressure calibration transcranial Doppler traumatic brain injury eye artery 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Czosnyka M, Pickard JD (2004) Monitoring and interpretation of intracranial pressure. J Neurol Neurosurg Psychiatry 75: 813–821PubMedCrossRefGoogle Scholar
  2. 2.
    Hassler W, Steinmetz H, Gawlowski J (1988) Transcranial Doppler ultrasonography in raised intracranial pressure and in intracranial circulatory arrest. J Neurosurg 68: 745–751PubMedGoogle Scholar
  3. 3.
    Kushiwaki H et al (1997) Human dural thickness measured by ultrasonographic method: reflection of intracranial pressure. J Ultrasound Med 16:11: 725–730Google Scholar
  4. 4.
    Motschmann M, Muller C, Kuchenbecker J et al (2001) Ophthalmodynamometry: a reliable method for measuring intracranial pressure. Strabismus 9: 13–16PubMedCrossRefGoogle Scholar
  5. 5.
    Querfurth HW, Lagreze WD, Hedges TR, Heggerick PA (2002) Flow velocity and pulsatility of the ocular circulation in chronic intracraqnial hypertension. Acta Neurol Scand 105(6): 431–440PubMedCrossRefGoogle Scholar
  6. 6.
    Ragauskas A, Daubaris G, Dziugys A (1999) Method and apparatus for determining the pressure inside the brain. US patent 5,951,477Google Scholar
  7. 7.
    Raksin PB, Alperin N, Sivaramakrishnan A et al (2003) Noninvasive intracranial compliance and pressure based on dynamic magnetic resonance imaging of blood flow and cerebrospinal fluid flow: review of principles, implementation, and other noninvasive approaches. Neurosurg Focus 14(4): article 4Google Scholar
  8. 8.
    Reid A, Marchbanks RJ, Bateman DE et al (1989) Mean intracranial pressure monitoring by a non-invasive audiological technique: a pilot study. J Neurol Neurosurg Psychiatry 52: 610–612PubMedCrossRefGoogle Scholar
  9. 9.
    Russegger L, Ennemoser O (1990) Atraumatic measurement of intracranial pressure. Wien Klin Wochenschr 102: 543–547 (in German)PubMedGoogle Scholar
  10. 10.
    Schmidt B, Klingelhofer J, Schwarce JJ et al (1997) Noninvasive prediction of intracranial pressure curves using transcranial Doppler ultrasonography and blood pressure curves. Stroke 28: 2465–2472PubMedGoogle Scholar
  11. 11.
    Schoser BG, Riemenschneider N, Hansen HC (1999) The impact of raised intracranial pressure on cerebral venous hemodynamics: a prospective venous transcranial Doppler ultrasonography. J Neurosurg 91: 744–749PubMedGoogle Scholar
  12. 12.
    Shakhnovich AR, Shakhnovich VA, Galushina AA (1999) Noninvasive assessment of the elastance and reserve capacity of the craniovertebral contents via flow velocity measurements in the straights sinus by TCD during body tilting test. J Neuroimaging 9: 141–149PubMedGoogle Scholar
  13. 13.
    Taylor WR, Chen JW, Meltzer H et al (2003) Quantitative pupillometry, a new technology: normative data and preliminary observations in patients with acute head injury. Technical Note. J Neurosurg 98: 205–213PubMedCrossRefGoogle Scholar
  14. 14.
    Ueno T, Shuer LM, Yost WT et al (1998) Development of a non-invasive technique for the measurement of intracranial pressure. Biol Sci Space 12: 270–271PubMedGoogle Scholar
  15. 15.
    US Patents (1980–2004) 4204547, 4984567, 5117835, 5388583, 5617873, 5919144, 6086533, 6117089, 6129682, 6231509, 6390989, 6413227, 6475147, 6589189, 6702743, 6740048, 6746410, 6773407Google Scholar

Copyright information

© Springer-Verlag 2005

Authors and Affiliations

  • A. Ragauskas
    • 1
  • G. Daubaris
    • 1
  • A. Dziugys
    • 1
  • V. Azelis
    • 2
  • V. Gedrimas
    • 2
  1. 1.Kaunas University of TechnologyKaunasLithuania
  2. 2.Kaunas University of MedicineKaunasLithuania

Personalised recommendations