Abstract
Two prototype position sensitive parallel plate avalanche counters (PPACs) have been fabricated at the Rare Isotope Science Project (RISP). Both detectors have active areas of 10 × 10 cm2 and utilize a stripped electrode structure, together with a delay line readout, to extract position information. Testing was conducted with an 241Am α source, and 2-dimensional position information was obtained. A position resolution of 0.55 mm full width at half maximum (FWHM), together with a position linearity of ±55 μm over a 67-mm detection range, were measured. A time resolution of 450 ps (FWHM) was also measured.
Similar content being viewed by others
References
J. E. Draper, Nucl. Instrum. Meth. 30, 148 (1964).
K. Tshoo, H. Chae, J. Park, J. Y. Moon, Y. K. Kwon, G. A. Souliotis, T. Hashimoto, C. Akers, G. P. A. Berg, S. Choi, S. C. Jeong, S. Kato, Y. K. Kim, S. Kubono, K. B. Lee and C. B. Moon, Nucl. Instrum. Meth. B 376, 188 (2016).
A. Zaghloul, Y. J. Kim, D. G. Kim, H. C. Jo and M. J. Kim, J. Korean Phys. Soc. 67, 1422 (2015).
H. Stelzer, Nucl. Instrum. Meth. 133, 409 (1976).
M. Nakhostin, Nucl. Instrum. Meth. A 615, 53 (2010).
H. Kumagai, A. Ozawa, N. Fukuda, K. Summerer and I. Tanihata, Nucl. Instrum. Meth. A 470, 562 (2001).
D. Harrach and H. J. Specht, Nucl. Instrum. Meth. 164, 477 (1979).
H. Kumagai, T. Ohnishi, N. Fukuda, H. Takeda, D. Kameda, N. Inabe, K. Yoshida and T. Kubo, Nucl. Instrum. Meth. B 317, 717 (2013).
J. F. Ziegler, M. Ziegler and J. Biersack, Nucl. Instrum. Meth. B 268, 1818 (2010).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Akers, C., Lee, K.B., Kim, Y.J. et al. Construction and performance test of prototype parallel plate avalanche counters at RAON. Journal of the Korean Physical Society 70, 682–686 (2017). https://doi.org/10.3938/jkps.70.682
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.3938/jkps.70.682