Abstract
This paper presents the experimental results on the intrafractional motion effect on dose distributions in a target in the context of spot-scanning beam proton therapy. The evaluation is performed by a qualitative analysis of the structure and shape of dose distributions and a quantitative analysis of the average dose and dose homogeneity within the region of interest for various motion parameters and acceleration cycles. The results show that intrafractional motion leads to a significant distortion of the dose field shape and structure. Distortions appear as hot and cold spots (overdosing and underdosing areas, respectively) dose field spread along the motion trajectory. These effects are reflected in a 16% reduction in the mean dose in the region of interest and dose homogeneity from 96.7 to 75.5% as the motion amplitude increases from 0 to 10 mm. In this case, the level of distortions weakly depends on time parameters of motion and the accelerator cycle. The observed effects lead to an efficiency degradation of proton therapy in the treatment of moving tumors; therefore, the use of such motion compensation methods as rescanning, gating, tracking, and their combinations is necessary for clinical practice.
Similar content being viewed by others
Change history
25 April 2023
An Erratum to this paper has been published: https://doi.org/10.3103/S1068335623020112
REFERENCES
Chernyaev, A.P., Klenov, G.I., Bushmanov, A.Yu., Pryanichnikov, A.A., Belikhin, M.A., and Lykova, E.N., Proton accelerators for radiation therapy, Med. Radiol. Radiat. Saf., 2019, vol. 64, no. 2, pp. 11–22. https://doi.org/10.12737/article_5ca5a0173e4963.18268254
Kubiak, T., Particle therapy of moving targets – the strategies for tumor motion monitoring and moving targets irradiation, Br. J. Radiol., 2016, vol. 89, no. 1066, p. 20150275. https://doi.org/10.1259/bjr.20150275
Trnková, P. et al., Clinical implementations of 4D pencil beam scanned particle therapy: Report on the 4D treatment planning workshop 2016 and 2017, Phys. Med., 2018, vol. 54, pp. 121–130. https://doi.org/10.1016/j.ejmp.2018.10.002
Bert, C. et al., Scanned carbon beam irradiation of moving films: comparison of measured and calculated response, Radiat. Oncol., 2012, vol. 7, p. 55. https://doi.org/10.1186/1748-717X-7-55
Bert, C. et al., Quantification of interplay effects of scanned particle beams and moving targets, Phys. Med. Biol., 2008, vol. 53, no. 9, p. 2253. https://doi.org/10.1088/0031-9155/53/9/003
Balakin, V.E., Belikhin, M.A., Pryanichnikov, A.A., Shemyakov, A.E., and Strelnikova, N.S., Clinical application of new immobilization system in seated position for proton therapy, The 2nd International Symposium on Physics, KnE Energy/Engineering and Technologies for Biomedicine, 2018, pp. 45–51. https://doi.org/10.18502/ken.v3i2.1790
Pryanichnikov, A.A., Sokunov, V.V., and Shemyakov, A.E., Some results of the clinical use of the Proton Therapy Complex “Prometheus,” Phys. Part. Nucl. Lett., 2018, vol. 15, no. 7, pp. 981–985. https://doi.org/10.1134/S1547477118070592
Balakin, V.E., Bazhan, A.I., Alexandrov, V.A., Chashurin, V.I., Lunev, P.A., Pryanichnikov, A.A., Shemyakov, A.E., Belikhin, M.A., and Shestopalov, A.I., Updated status of Protom synchrotrons for radiation therapy, Proc. RuPAC'21, 2021, pp. 120–123. https://doi.org/10.18429/JACOW-RUPAC2021-FRB05
Siksin, V.V., Ryabov, V.A., and Shemyakov, A.E., Registration of Bragg peaks in scanning mode of a target with a low-intensity proton beam from the PROMETEUS medical accelerator, Bull. Lebedev Phys. Inst., 2021, vol. 48, no. 12, pp. 16–21.
Siksin, V.V. and Shemyakov, A.E., Measurements of the Bragg peak depth by a PPC-40 ionization chamber and a digital imaging detector, Bull. Lebedev Phys. Inst., 2020, vol. 47, no. 9, 285–289. https://doi.org/10.3103/S1068335620090055
Wang, N. et al., Evaluation and comparison of New 4DCT based strategies for proton treatment planning for lung tumors, Radiat. Oncol., 2013, vol. 8, p. 73. https://doi.org/10.1186/1748-717X-8-73
Funding
This study was supported within the Federal Scientific and Technical program no. 075-15-2021-1347 (October 5, 2021) by the Ministry of Science and Higher Education of the Russian Federation “Development of new technologies for diagnostics and beam therapy of socially significant diseases by proton and ion beams using binary nuclear methods,” as well as technically supported by the Protom Ltd.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
The authors declare that they have no conflicts of interest.
Additional information
Translated by A. Kazantsev
About this article
Cite this article
Belikhin, M.A., Grigoryeva, M.S., Zavestovskaya, I.N. et al. EXPERIMENTAL STUDY OF THE TARGET MOTION EFFECT ON THE DOSE DISTRIBUTION IN SPOT-SCANNING BEAM PROTON THERAPY. Bull. Lebedev Phys. Inst. 49, 132–136 (2022). https://doi.org/10.3103/S1068335622050025
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.3103/S1068335622050025