Abstract—An immunochromatographic test system for the combined analysis of two cardiomarkers, such as fatty acid binding protein (FABP) and troponin I (TnI), in serum and whole blood is developed. The main stages of the design of a multiparametric system are investigated, including the influence of the antibody position and gold nanoparticle size on the results. The detection limits of the method are found to be 10 ng/mL for FABP and 1 ng/mL for TnI, and the coefficient of variation did not exceed 10%. The developed test system allows us to confirm the diagnosis of acute myocardial infraction (N = 9). The analysis time is 15 min.
Similar content being viewed by others
REFERENCES
Rosen, S., in Lateral Flow Immunoassay, Wong, R.C. and Tse, H.Y., Eds., New York: Humana, 2009, p. 35.
Posthuma-Trumpie, G., Korf, J., and van Amerongen, A., Anal. Bioanal. Chem., 2009, vol. 393, p. 569.
Huang, X., Aguilar, Z.P., Xu, H., Lai, W., and Xiong, Y., Biosens. Bioelectron., 2016, vol. 75, p. 166.
Song, S., Liu, N., Zhao, Z., Ediage, E.N., Wu, S., Sun, C., De Saeger, S., and Wu, A., Anal. Chem., 2014, vol. 86, p. 4995. https://doi.org/10.1021/ac500540z
Wang, L., Cai, J., Wang, Y., Fang, Q., Wang, S., Cheng, Q., Du, D., Lin, Y., and Liu, F., Microchim. Acta, 2014, vol. 181, p. 1565.
Taranova, N.A., Berlina, A.N., Zherdev, A.V., and Dzantiev, B.B., Biosens. Bioelectron., 2015, vol. 63, p. 255.
Xing, C., Liu, L., Song, S., Feng, M., Kuang, H., and Xu, C., Biosens. Bioelectron., 2015, vol. 66, p. 445. https://doi.org/10.1016/j.bios.2014.12.004
Wang, C., Hou, F., and Ma, Y., Biosens. Bioelectron., 2015, vol. 68, p. 156. https://doi.org/10.1016/j.bios.2014.12.051
Chinnasamy, T., Segerink, L.I., Nystrand, M., Gantelius, J., and Svahn, H.A., Analyst, 2014, vol. 139, p. 2348.
Babuin, L. and Jaffe, A.S., Can. Med. Assoc. J., 2005, vol. 173, p. 1191.
Lescuyer, P., Allard, L., Hochstrasser, D.F., and Sanchez, J.C., Mol. Diagn., 2005, vol. 9, p. 1.
Lippi, G., Mattiuzzi, C., and Cervellin, G., Clin. Biochem., 2013, vol. 46, p. 26.
Abbasi, W.A., Saleem, M., Rasheed, S., and Kiyani, A.M., J. Ayub Med. Coll. Abbottabad, 2017, vol. 29, p. 107.
Liang, Y., Chan, C., Cheung, K., Cautherley, G., and Renneberg, R., J. Immunoassay Immunochem., 2011, vol. 32, p. 342.
Li, C.J., Li, J.Q., Liang, X.F., Li, X.X., Cui, J.G., Yang, Z.J., Guo, Q., Cao, K.J., and Huang, J., Acta Pharmacol. Sin., 2010, vol. 31, p. 307.
Zyryanova, A.V., Yarokhno, N.N., and Nikolaev, K.Yu., Patol. Krovoobrashch,Kardiokhir., 2010, vol. 4, p. 46.
Ramparany, L., Ramirez, J., Nizou, J., Le Saux, D., Richard, V., and Talarmin, A, Clin. Vaccine Immunol., 2011, vol. 18, p. 414.
Zhu, J., Zou, N., Zhu, D., Wang, J., Jin, Q., Zhao, J., and Mao, H., Clin. Chem., 2011, vol. 57, p. 1732.
Diagnostic Automation/Cortez Diagnostics. http://www. rapidtest.com/pdf/3_in_1_Troponin_I_CK-MB_ Myoglobin_Serum_WB_166779-1(8-21-2012).pdf.
Yakovleva, E.A., Andreeva, I.P., Grigorenko, V.G., and Osipov, A.P., Moscow Univ. Chem. Bull. (Engl. Transl.), 2011, vol. 66, p. 356. https://doi.org/10.3103/S0027131411060101
Yakovleva, E.A., Andreeva, I.P., Grigorenko, V.G., Egorov, A.M., and Osipov, A.P., Moscow Univ. Chem. Bull. (Engl. Transl.), 2012, vol. 67, p. 249. https://doi.org/10.3103/S0027131412060107
Grigorenko, V., Andreeva, I., Borchers, T., Spener, F., and Egorov, A., Anal. Chem., 2001, vol. 73, p. 1134.
Andreeva, I.P., Grigorenko, V.G., Egorov, A.M., and Osipov, A.P., Anal. Lett., 2016, vol. 49, p. 579.
Liu, X., Atwater, M., Wang, J., and Huo, Q., Colloids Surf., B, 2007, vol. 58, p. 3.
Funding
This work was supported by the Ministry of Education and Science as part of the state task “Organization of research” (agreement no. 16.6548.2017/VU).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
The authors declare that they have no conflict of interest.
Additional information
Translated by O. Maslova
About this article
Cite this article
Andreeva, I.P., Yakovleva, E.A., Grigorenko, V.G. et al. The Development of a Multiparametric Immunochromatographic Test System for the Analysis of Cardiomarkers. Moscow Univ. Chem. Bull. 74, 315–321 (2019). https://doi.org/10.3103/S002713141906004X
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.3103/S002713141906004X