Abstract
X-ray absorption spectral (XAS) analysis is performed with a combination of a 9 V dry electric battery X-ray generator and a portable Si PIN X-ray detector. The calcium K edge (4.0 keV) in paper is measured with a grazing incidence geometry, which suppressed the artifact due to the Kα X-ray fluorescence peak at 3.5 keV. The 9 V dry battery X-ray emitter is useful for portable XAS measurements.