Abstract
New polymeric electrodes responding to tiamulin hydrogen fumarate were constructed by incorporating tiamulin tetraphenylborate and tiamulin phosphotungstate ion-pair complexes into PVC membranes, with the latter giving the best linear and Nernstian response. The electrodes exhibited a linear response over the concentration range 10−2-10−6 mol l−1. The pH did not affect the response of the electrodes within the range 2–8. Direct potentiometry was used to assay tiamulin hydrogen fumarate in pharmaceutical formulations and in the bulk form. The results were compared with those obtained with non-aqueous titration for the antibiotic in the bulk form and with HPLC for commercial formulations.
Similar content being viewed by others
References
D. C. Blood, O. M. Radostits, J. H. Arundel and C. C. Gay, “Veterinary Medicine”, 7th ed., p. 153. Baillière Tindall, London, 1977.
J. F. Prescott and J. D. Baggot, “Veterinary Medicine”, pp. 173–176, Blackwell Scientific Publications, Boston, 1988.
R. J. Bywater, in “Veterinary Applied Pharmacology and Therapeutics”, ed. G. C. Brander, D. M. Pugh, R. J. By water and W. L. Jenkins, p. 218. Baillière Tindall, London, 1977.
J. R. Markus and J. Sherma, J. AOAC Int., 76, 444 (1993).
J. R. Markus and J. Sherma, J. AOAC Int., 76, 447 (1993).
J. R. Markus and J. Sherma, J. AOAC Int., 76, 449 (1993).
J. R. Markusand J. Sherma, J. AOAC Int., 76, 451 (1993).
D. Howard and T. Cowen, Analyst [London], 107, 319 (1982).
G. Gübitz and G. Amann, Mikrochim. Acta, 2, 355 (1979).
R. F. Cosgrove and G. T. Jones, Proc. Anal. Div. Chem. Soc, 320 (1979)
Biochimie GmbH (Kundl, Austria), “Internal Control Procedure 326. 8”, 1992.
V. V. Cosofret and R. P. Buck, Analyst [London], 109, 1321 (1984).
A. A. Bouklouze, A. El-Jammal, G. J. Patriarche and G. D. Christian, J. Pharm. Biotned Anal, 9, 393 (1991).
G. J. Moody and J. D. R. Thomas in “Ion-Selective Electrodes in Analytical Chemistry”, ed. H. Freiser, pp. 287–309, Plenum Press, New York.
A. El-Jammal, A. A. Bouklouze, G. J. Patriarche and G. D. Christian, Talanta, 38, 929 (1991).
A. A. Bouklouze, A. El-Jammal and G. J. Patriarche, Analusis, 19, 27 (1991).
T. Rosatzin, E. Bakker, K. Suzuki and W. Simon, Anal Chim. Acta, 280, 197 (1993).
R. D. Armstrong and G. Horvai, Bectrochim. Acta, 35, 1 (1990).
R. P. Buck and V. V. Cosofret, Pure Appl Chem., 65, 1849 (1993).
Biochimie GmbH (Kundl, Austria), “Internal Control Procedure 1344.4”, 1992.
P. L. Bailey, in “Analysis with Ion-Selective Electrodes”, 2nd ed., Heyden, London, 1977.
Biochimie GmbH (Kundl, Austria), “Internal Control Procedure 1269. 1”, 1989.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Marques, M.R.C., de Neto, G.O. Poly(vinyl chloride) Membrane Electrodes for the Determination of the Antibiotic Tiamulin. ANAL. SCI. 13, 457–461 (1997). https://doi.org/10.2116/analsci.13.457
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.2116/analsci.13.457