Abstract
Aim
The aim of this study was to characterise seizure-related adverse events during video-electroencephalography (vEEG) monitoring.
Methods
Our study evaluated adverse events in 230 epilepsy patients during vEEG monitoring while patients were awake and asleep.
Results
A total of 588 seizures were recorded and 231 adverse events were identified including electrode displacement (14.5%), aspiration risk (8.5%), urinary incontinence (7.5%), postictal psychosis (5.8%), tongue biting (5.3%), and patient falls (0.5%). No severe types of adverse events were observed, such as choking or aspiration pneumonia. Of the seizures recorded, 39.1% occurred while patients were sleeping and 38.5% of adverse events occurred during this time, which included electrode displacement (48.2%), aspiration risk (28.0%), tongue biting (60.0%), postictal psychosis (16.7%), patient falls (66.7%), and urinary incontinence (38.6%).
Conclusion
The occurrence of seizure-related adverse events during vEEG monitoring is mild. vEEG is an acceptably safe procedure for epilepsy research and precautions should be put in place to prevent these events.
Article PDF
Similar content being viewed by others
References
Alper K, Devinsky O, Westbrook L, et al. Premorbid psychiatric risk factors for postictal psychosis. J Neuropsychiatry Clin Neurosci 2001; 13: 492–499.
Angus-Leppan H. Seizures and adverse events during routine scalp electroencephalography: a clinical and EEG analysis of 1000 records. Clin Neurophysiol 2007; 118: 22–30.
Appleton RE. Seizure-related injuries in children with newly diagnosed and untreated epilepsy. Epilepsia 2002; 43: 764–767.
Arzy S, Allali G, Brunet D, et al. Antiepileptic drugs modify power of high eeg frequencies and their neural generators. Eur J Neurol 2010; 17: 1308–1312.
Beghi E, Cornaggia C. Morbidity and AEs in patients with epilepsy: results of a european cohort study. Epilepsia 2002; 43: 1076–1083.
Benbadis SR. Tongue biting as a lateralizing sign in partial epilepsy. Seizure 1996; 5: 175–176.
Benbadis SR, Wolgamuth BR, Goren H, Brener S, Fouad-Tarazi F. Value of tongue biting in the diagnosis of seizures. Arch Intern Med 1995; 155: 2346–2349.
Buelow JM, Privitera M, Levisohn P, et al. A description of current practice in epilepsy monitoring units. Epilepsy Behav 2009; 15: 308–313.
Day SM, Wu YW, Strauss DJ, et al. Causes of death in remote symptomatic epilepsy. Neurology 2005; 65: 216–222.
DeToledo JC, Lowe MR, Gonzalez J, et al. Risk of aspiration pneumonia after an epileptic seizure: a retrospective analysis of 1634 adult patients. Epilepsy Behav 2004; 5: 593–595.
Devinsky O, Abramson H, Alper K, et al. Postictal psychosis: a case control series of 20 patients and 150 controls. Epilepsy Res 1995; 20: 247–253.
Dobesberger J, Walser G, Unterberger I, et al. Video-EEG monitoring: safety and adverse events in 507 consecutive patients. Epilepsia 2010; 52: 443–452.
Engel J Jr. A proposed diagnostic scheme for people with epileptic seizures and with epilepsy: report of the ILAE task force on classification and terminology. Epilepsia 2001; 42: 796–803.
Halasz P, Filakovszky J, Vargha A, et al. Effect of sleep deprivation on spike-wave discharges in idiopathic generalised epilepsy: a 4×24 h continuous long-term EEG monitoring study. Epilepsy Res 2002; 51: 123–132.
Haut SR, Swick C, Freeman K, et al. Seizure clustering during epilepsy monitoring. Epilepsia 2002; 43: 711–715.
Hui AC, Kwan P, Leung TW, et al. Diagnostic value and safety of long-term video-EEG monitoring. Hong Kong Med J 2007; 13: 228–230.
Lees A. Retrospective study of seizure-related injuries in older people: a 10-year observation. Epilepsy Behav 2010; 19: 441–444.
Logsdail SJ, Toone BK. Post-ictal psychoses.Aclinical and phenomenological description. Br J Psychiatry 1988; 152: 246–252.
Morgan VR, Mathison JH, Rice JC, et al. A persistent problem. Am J Public Health 1985; 75: 775–777.
Noe KH, Drazkowski JF. Safety of long-term video-electroencephalographic monitoring for evaluation of epilepsy. Mayo Clin Proc 2009; 84: 495–500.
Noe KH, Tapsell LM, Drazkowski JF. Risk of choking and aspiration during inpatient video-EEG monitoring. Epilepsy Res 2010; 93: 84–86.
Nordli DR Jr. Usefulness of video-EEG monitoring. Epilepsia 2006; 47: 26–30.
Rose AB, McCabe PH, Gilliam FG, et al. Occurrence of seizure clusters and status epilepticus during inpatient video-EEG monitoring. Neurology 2003; 60: 975–978.
Sanders PT, Cysyk BJ, Bare MA. Safety in long-term EEG/video monitoring. J Neurosci Nurs 1996; 28: 305–313.
Shafer PO, Buelow J, Ficker DM, et al. Risk of adverse events on epilepsy monitoring units: a survey of epilepsy professionals. Epilepsy Behav 2011; 20: 502–505.
Simon SL, Telfeian A, Duhaime AC. Complications of invasive monitoring used in intractable pediatric epilepsy. Pediatr Neurosurg 2003; 38: 47–52.
Wirrell EC. Epilepsy-related injuries. Epilepsia 2006; 47: 79–86.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Liu, J., Meng, F. & Liu, Z. Seizure-related adverse events during video-electroencephalography monitoring. Epileptic Disord. 14, 51–56 (2012). https://doi.org/10.1684/epd.2012.0490
Received:
Accepted:
Published:
Issue Date: