Abstract
p-Nitrophenylphosphate (PNPP) is usually employed as the substrate for enzyme-linked immunosorbent assays. p-Nitrophenol (PNP), the product of PNPP, with the catalyst alkaline phosphatase (ALP), will passivate an electrode, which limits applications in electrochemical analysis. A novel anti-passivation ink used in the preparation of a graphene/ionic liquid/chitosan composited (rGO/IL/Chi) electrode is proposed to solve the problem. The anti-passivation electrode was fabricated by directly writing the graphene-ionic liquid-chitosan composite on a single-side conductive gold strip. A glassy carbon electrode, a screen-printed electrode, and a graphene-chitosan composite-modified screen-printed electrode were investigated for comparison. Scanning electron microscopy was used to characterize the surface structure of the four different electrodes and cyclic voltammetry was carried out to compare their performance. The results showed that the rGO/IL/Chi electrode had the best performance according to its low peak potential and large peak current. Amperometric responses of the different electrodes to PNP proved that only the rGO/IL/Chi electrode was capable of anti-passivation. The detection of cardiac troponin I was used as a test example for electrochemical immunoassay. Differential pulse voltammetry was performed to detect cardiac troponin I and obtain a calibration curve. The limit of detection was 0.05 ng/ml.
摘要
目的
酶联免疫分析常用底物对硝基苯磷酸盐分解后产 生对硝基苯酚(PNP),该物质对电极具有钝化 作用,限制了该体系在电化学免疫分析领域的应 用。因此,需要开发一种具有抗钝化作用的电极 浆料。
创新点
石墨烯-子液体-壳聚糖(rGO/IL/Chi)电极浆料 不仅具有抗钝化作用,且拥有良好的电化学性 能,制备过程中无需使用任何有机溶剂,安全环 保。
方法
制备rGO/IL/Chi 复合浆料,涂布于单面导电金箔 上,75 °C,10 min,将其作为工作电极;背面涂布 自制的银浆,作为参比电极和对电极;0.03 g/ml 壳 聚糖溶液作为绝缘浆料。
结论
本研究表明,在比较商品化的玻碳电极、自制丝 网印刷电极、石墨烯修饰的丝网印刷电极和 rGO/IL/Chi 电极对PNP 的循环伏安响应时, rGO/IL/Chi 电极具有最大的峰电流响应和最小的 峰电位。同时,对四种电极进行性能表征时, rGO/IL/Chi 电极具有最小的峰电位差和最大的峰 电流。这表明rGO/IL/Chi 电极具有较好的电化学 性能,且对PNP 有较大的响应。比较四种电极对 PNP 的安培响应之后,发现只有rGO/IL/Chi 电极 具有抗钝化作用。可将该电极用于肌钙蛋白I 的 检测,其检测限为0.05 ng/ml。
Similar content being viewed by others
References
Aneesh K, Berchmans S, 2017. Highly selective sensing of dopamine using carbon nanotube ink doped with anionic surfactant modified disposable paper electrode. J Solid State Electrochem, 21(5):1263–1271. https://doi.org/10.1007/s10008-016-3482-2
Bansod B, Kumar T, Thakur R, et al., 2017. A review on various electrochemical techniques for heavy metal ions detection with different sensing platforms. Biosens Bioelectron, 94:443–455. https://doi.org/10.1016/j.bios.2017.03.031
Budnikov HC, Shirokova VI, 2013. Term “Nano” in electroanalysis:a trendy prefix or a new stage of its development? J Anal Chem, 68(8):663–670. https://doi.org/10.1134/S1061934813080030
Cinti S, Basso M, Moscone D, et al., 2017. A paper–based nanomodified electrochemical biosensor for ethanol detection in beers. Anal Chim Acta, 960:123–130. https://doi.org/10.1016/j.aca.2017.01.010
Dinesh B, Saraswathi R, Kumar AS, 2017. Water based homogenous carbon ink modified electrode as an efficient sensor system for simultaneous detection of ascorbic acid, dopamine and uric acid. Electrochim Acta, 233:92–104. https://doi.org/10.1016/j.electacta.2017.02.139
Ghosale A, Shrivas K, Shankar R, et al., 2017. Low–cost paper electrode fabricated by direct writing with silver nanoparticlebased ink for detection of hydrogen peroxide in wastewater. Anal Chem, 89(1):776–782. https://doi.org/10.1021/acs.analchem.6b03512
González–Sánchez MI, Valero E, Compton RG, 2016. Iodine mediated electrochemical detection of thiols in plant extracts using platinum screen–printed electrodes. Sens Actuators B Chem, 236:1–7. https://doi.org/10.1016/j.snb.2016.05.152
Idris A, Saleh TA, Sanhoob MA, et al., 2017. Electrochemical detection of thiocyanate using phosphate–modified zeolite carbon paste electrodes. J Taiwan Inst Chem Eng, 72:236–243. https://doi.org/10.1016/j.jtice.2017.01.015
Kuila T, Bose S, Khanra P, et al., 2011. Recent advances in graphene–based biosensors. Biosens Bioelectron, 26(12):4637–4648. https://doi.org/10.1016/j.bios.2011.05.039
Li S, Zhang Q, Lu YL, et al., 2017. One step electrochemical deposition and reduction of graphene oxide on screen printed electrodes for impedance detection of glucose. Sens Actuators B Chem, 244:290–298. https://doi.org/10.1016/j.snb.2016.12.142
Li SJ, Zuo Y, Huang WF, 2017. Establishment of a reference interval for high–sensitivity cardiac troponin I in healthy adults from the Sichuan area. Medicine, 96(14):e6252. https://doi.org/10.1097/MD.0000000000006252
Ma XY, Chao MY, Wang ZX, et al., 2012. Electrochemical detection of dopamine in the presence of epinephrine, uric acid and ascorbic acid using a graphene–modified electrode. Anal Methods, 4(6):1687–1692. https://doi.org/10.1039/c2ay25040c
Molazemhosseini A, Magagnin L, Vena P, et al., 2017. Singleuse nonenzymatic glucose biosensor based on CuO nanoparticles ink printed on thin film gold electrode by micro–plotter technology. J Electroanal Chem, 789:50–57. https://doi.org/10.1016/j.jelechem.2017.01.041
Nesakumar N, Sethuraman S, Krishnan UM, et al., 2016. Electrochemical acetylcholinesterase biosensor based on ZnO nanocuboids modified platinum electrode for the detection of carbosulfan in rice. Biosens Bioelectron, 77:1070–1077. https://doi.org/10.1016/j.bios.2015.11.010
Ping JF, Wu J, Ying YB, 2010. Development of an ionic liquid modified screen–printed graphite electrode and its sensing in determination of dopamine. Electrochem Commun, 12(12):1738–1741. https://doi.org/10.1016/j.elecom.2010.10.010
Ping JF, Wang YX, Ying YB, et al., 2012a. Application of electrochemically reduced graphene oxide on screen–printed ion–selective electrode. Anal Chem, 84(7):3473–3479. https://doi.org/10.1021/ac203480z
Ping JF, Wu J, Wang YX, et al., 2012b. Simultaneous determination of ascorbic acid, dopamine and uric acid using high–performance screen–printed graphene electrode. Biosens Bioelectron, 34(1):70–76. https://doi.org/10.1016/j.bios.2012.01.016
Preechaworapun A, Dai Z, Xiang Y, et al., 2008. Investigation of the enzyme hydrolysis products of the substrates of alkaline phosphatase in electrochemical immunosensing. Talanta, 76(2):424–431. https://doi.org/10.1016/j.talanta.2008.03.025
Saita T, Yamamoto Y, Hosoya K, et al., 2017. An ultraspecific and sensitive sandwich ELISA for imatinib using two anti–imatinib antibodies. Anal Chim Acta, 969:72–78. https://doi.org/10.1016/j.aca.2017.03.034
Sajid M, Nazal MK, Mansha M, et al., 2016. Chemically modified electrodes for electrochemical detection of dopamine in the presence of uric acid and ascorbic acid:a review. TrAC Trend Anal Chem, 76:15–29. https://doi.org/10.1016/j.trac.2015.09.006
Secor EB, Ahn BY, Gao TZ, et al., 2015. Rapid and versatile photonic annealing of graphene inks for flexible printed electronics. Adv Mater, 27(42):6683–6688. https://doi.org/10.1002/adma.201502866
Tang WZ, Wu J, 2014. Amperometric determination of organophosphorus pesticide by silver electrode using an acetylcholinesterase inhibition method. Anal Methods, 6(3):924–929. https://doi.org/10.1039/C3AY41932K
Tang WZ, Zhou JZ, Yang QQ, et al., 2014. Determination of methyl parathion by solid–phase extraction on an ionic liquid–carbon nanotube composite electrode. Anal Methods, 6(15):5886–5890. https://doi.org/10.1039/C4AY01062K
Wang YX, Ping JF, Ye ZZ, et al., 2013. Impedimetric immunosensor based on gold nanoparticles modified graphene paper for label–free detection of Escherichia coli O157:H7. Biosens Bioelectron, 49:492–498. https://doi.org/10.1016/j.bios.2013.05.061
Yang SX, Chen YC, Nicolini L, et al., 2015. “Cut–and–Paste” manufacture of multiparametric epidermal sensor systems. Adv Mater, 27(41):6423–6430. https://doi.org/10.1002/adma.201502386
Yu XW, Sheng KX, Shi GQ, 2014. A three–dimensional interpenetrating electrode of reduced graphene oxide for selective detection of dopamine. Analyst, 139(18):4525–4531. https://doi.org/10.1039/C4AN00604F
Zheng QQ, Yu YH, Fan K, et al., 2016. A nano–silver enzyme electrode for organophosphorus pesticide detection. Anal Bioanal Chem, 408(21):5819–5827. https://doi.org/10.1007/s00216-016-9694-6
Author information
Authors and Affiliations
Corresponding author
Additional information
Project supported by the National Natural Science Foundation of China (No. 31571918)
Electronic supplementary materials: The online version of this article (https://doi.org/10.1631/jzus.B1700510) contains supplementary materials, which are available to authorized users
Electronic supplementary material
Rights and permissions
About this article
Cite this article
Zheng, Qq., Lu, Yc., Ye, Zz. et al. An anti-passivation ink for the preparation of electrodes for use in electrochemical immunoassays. J. Zhejiang Univ. Sci. B 19, 726–734 (2018). https://doi.org/10.1631/jzus.B1700510
Received:
Revised:
Published:
Issue Date:
DOI: https://doi.org/10.1631/jzus.B1700510