Optimizing Homogeneous Thin Solid Films (HTSFs) from μL-Blood Droplets via Hyper-Hydrophilic Coatings (HemaDrop™) for Accurate Compositional Analysis via IBA, XRF, and XPS

  • Articles
  • Published:
MRS Advances Aims and scope Submit manuscript

Abstract

Liquid phase analysis dominates the field of blood diagnostics and requires drawing blood volumes of several ml for each test. To achieve acceptable accuracy, each single liquid blood test requires ∼7 mL per blood sample, and repeated blood tests are often needed. Frequent testing ca result in Hospital Acquired Anemia for infants, chronically ill, and critically ill patients. Blood testing methods that can be utilized with small amounts of blood are a critical need to save lives. Theranos claimed to have developed novel methods requiring only a few nL of blood. However, Theranos’ techniques led to errors that exceeded beyond the medically acceptable threshold of 10%. This work investigates solid state blood analysis using low volumes of several µL. The most common blood tests used as first line for diagnostics and monitoring patients’ status, always include blood electrolytes, iron, and in some cases, heavy metals.

The present work investigates the formation of rapidly solidified Homogeneous Thin Solid Films (HTSFs) formed from blood drops, in order to make them suitable for solid state analysis in vacuo and in air. The solidification of ∼5 micro-liter (µL)-sized blood droplets into HTSFs is studied with two goals: achieve reproducible HTSFs optimized for producing accurate analysis, and successfully measure the potential accuracy of measurements made on HTSFs for blood electrolytes Na, K, Mg, Ca, and Cl and heavy metals such as Fe.

The blood volumes selected for this work are in the µL range, one thousandth volumes drawn for current liquid phase analysis. Balanced Saline Solution (BSS) is used as an initial liquid for testing solidification uniformity and a potential calibration material. Next, canine and human blood are studied on two types of HemaDrop™ coatings for solidification: super-hydrophilic and hyper-hydrophilic. HTSF formation from BSS and blood drops are compared on both coated and uncoated surfaces.

Three solid state analytical methods are investigated in parallel to probe composition at different depths and test each for reproducibility and accuracy: Ion Beam Analysis (IBA), X-ray Fluorescence (XRF), and X-ray Photoelectron Spectroscopy (XPS). The results show that using solid films of blood yields composition, which can be reproducibly measured by IBA, XPS and XRF to varying degrees. XPS’s depth of analysis, limited to ∼5 nm, probes a small fraction of the HTSF, but provides insights into the range of thickness for homogeneous compositions in HTSFs. Statistical and error analysis help establish whether measurements taken in sets of three typically used in lab fall below the medically accepted error threshold (<10%) for each technique and element detected. Measurements are repeated and taken at various locations and on different HTSFs to establish reproducibility. XRF is of particular interest, because it is fast, accurate, portable and can be conducted in air, making it ideal for areas with limited resources.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. A M Harris, C L J Atterbury, B Chaffe, C Elliott, T Hawkins, S J Hennem, C Howell, J.Jones, S Murray, H V New, et al. Guideline on the Administration of Blood Components.

  2. M P Arsie, L Marchioro, A Lapolla, G F Giacchetto, M R Bordin, P Rizzotti, and D Fedele. Evaluation of diagnostic reliability of DCA 2000 for rapid and simple monitoring of HbA1c. Acta diabetologica, 37(1):1–7, 2000.

    Article  CAS  Google Scholar 

  3. Ming Zhi, Eric L Ding, Jesse Theisen-Toupal, Julia Whelan, and Ramy Arnaout. The landscape of inappropriate laboratory testing: a 15-year meta-analysis. PloS one, 8(11): e78962, 2013.

    Article  Google Scholar 

  4. J Buettner, S Puser, et al., editors. Senses, sensors, and systems: a journey through the history of laboratory diagnosis. F Hoffmann-La Roche & Co., Basel, Switzerland, 2004.

    Google Scholar 

  5. J G Van der Bom and S C Cannegieter. Hospital-acquired anemia: the contribution of diagnostic blood loss. Journal of Thrombosis and Haemostasis, 13(6):1157–1159, 2015.

    Article  Google Scholar 

  6. Adam C Salisbury, Kimberly J Reid, Karen P Alexander, Frederick A Masoudi, Sue-Min Lai, Paul S Chan, Richard G Bach, Tracy Y Wang, John A Spertus, and Mikhail Kosiborod. Diagnostic blood loss from phlebotomy and hospital-acquired anemia during acute myocardial infarction. Archives of internal medicine, 171(18):1646–1653, 2011.

    Article  Google Scholar 

  7. Rajiv N Thakkar, Daniel Kim, Amy M Knight, Stefan Riedel, Dhananjay Vaidya, and Scott M Wright. Impact of an educational intervention on the frequency of daily blood test orders for hospitalized patients. American journal of clinical pathology, 143(3):393–397, 2015.

    Article  Google Scholar 

  8. Richard John Hamilton and Peter Alexis Sewell. Introduction to high performance liquid chromatography, 1982.

  9. J J Kirkland and R M McCormick. Liquid phase separation methods: HPLC, FFF, electrophoresis. Chromatographia, 24(1):58–76, 1987.

    Article  CAS  Google Scholar 

  10. Carl L Shackelford and Kenneth Rainin. Modular liquid chromatography column apparatus, 1987.

  11. Gerhard Maldener. Requirements and tests for HPLC apparatus and methods in pharmaceutical quality control. Chromatographia, 28(1–2):85–88, 1989.

    Article  Google Scholar 

  12. Rolf Andersson and Bruno Hedlund. HPLC analysis of organic acids in lactic acid fermented vegetables. Zeitschrift für Lebensmittel-Untersuchung und Forschung, 176(6):440–443, 1983.

    Article  CAS  Google Scholar 

  13. J A Hamann, K Johnson, and Delbert T Jeter. HPLC determination of clenbuterol in pharmaceutical gel formulations. Journal of chromatographic science, 23(1):34–36, 1985.

    Article  CAS  Google Scholar 

  14. Masato Homma, Anura L Jayewardene, John Gambertoglio, and Francesca Aweeka. High-performance liquid chromatographic determination of ribavirin in whole blood to assess disposition in erythrocytes. Antimicrobial agents and chemotherapy, 43(11):2716–2719, 1999.

    Article  CAS  Google Scholar 

  15. Brian A Kidd, Gabriel Hoffman, Noah Zimmerman, Li Li, Joseph W Morgan, Patricia K Glowe, Gregory J Botwin, Samir Parekh, Nikolina Babic, Matthew W Doust, et al. Evaluation of direct-to-consumer low-volume lab tests in healthy adults. The Journal of clinical investigation, 126(5):1734–1744, 2016.

    Article  Google Scholar 

  16. Chunyang Y Zheng, Guanghui Ma, and Zhiguo Su. Native PAGE eliminates the problem of PEG-SDS interaction in SDS-PAGE and provides an alternative to HPLC in characterization of protein PEGylation. Electrophoresis, 28(16):2801–2807, 2007.

    Article  CAS  Google Scholar 

  17. Elizabeth Holmes. Systems and methods for multi-analysis, 2013.

  18. Eleftherios P Diamandis. Theranos phenomenon: promises and fallacies. Clinical Chemistry and Laboratory Medicine (CCLM), 53(7):989–993, 2015.

    Article  CAS  Google Scholar 

  19. Blood Pressure Lowering Treatment Trialists’ Collaboration et al. Effects of different blood-pressure-lowering regimens on major cardiovascular events: results of prospectively-designed overviews of randomized trials. The Lancet, 362(9395):1527–1535, 2003.

    Article  Google Scholar 

  20. David C Klonoff, Joan Lee Parkes, Boris P Kovatchev, David Kerr, Wendy C Bevier, Ronald L Brazg, Mark Christiansen, Timothy S Bailey, James H Nichols, and Michael A Kohn. Investigation of the accuracy of 18 marketed blood glucose monitors. Diabetes Care, 41(8):1681–1688, 2018.

    Article  CAS  Google Scholar 

  21. Yash Pershad, Ashley A Mascareno, Makoyi R Watson, Alex L Brimhall, Nicole Herbots, Clarizza F Watson, Abijith Krishnan, Nithin Kannan, Mark W Mangus, Robert J Culbertson, et al. Electrolyte Detection by Ion Beam Analysis, in Continuous Glucose Sensors and in Microliters of Blood using a Homogeneous Thin Solid Film of Blood, HemaDrop™. MRS Advances, 1(29):2133–2139, 2016.

    Article  CAS  Google Scholar 

  22. Yash Pershad, Nicole Herbots, Grady Day, Ryan Van Haren, Shawn Whaley, Alvaro Martinez, Sabrina Suhartono, Robert Culbertson, Mark Mangus, and Barry Wilkens. Determining Canine Blood and Human Blood Composition by Congealing Microliter Drops into Homogeneous Thin Solid Films (HTSFs) via HemaDrop™. MRS Advances, 2(45):2451–2456, 2017.

    Article  CAS  Google Scholar 

  23. Nico Grüner, Oumaima Stambouli, and R Stefan Ross. Dried blood spots-preparing and processing for use in immunoassays and in molecular techniques. JoVE (Journal of Visualized Experiments), (97): e52619, 2015.

  24. Sharmila M Mukhopadhyay. Sample preparation for microscopic and spectroscopic characterization of solid surfaces and films. CHEMICAL ANALYSIS-NEW YORK-INTERSCIENCE THEN JOHN WILEY-, pages 377–412, 2003.

    Google Scholar 

  25. John P Greer, Bertil Glader DAA, Alan F List, Robert T Means, Frixos Paraskevas, and George M Rodgers. Wintrobe’s Clinical Hematology, (2014). Lippincott Williams & Wilkins, Philadelphia, 14 edition, 2014.

    Google Scholar 

  26. Rong Fan, Ophir Vermesh, Alok Srivastava, Brian KH Yen, Lidong Qin, Habib Ahmad, Gabriel A Kwong, Chao-Chao Liu, Juliane Gould, Leroy Hood, et al. Integrated barcode chips for rapid, multiplexed analysis of proteins in microliter quantities of blood. Nature biotechnology, 26(12):1373, 2008.

    Article  CAS  Google Scholar 

  27. Xing, Q., Hart, M. A., Culbertson, R., Bradley, J. D., Herbots, N., Wilkens, B. J., Watson, C. F. (2011). Particle-Induced X-ray Emission (PIXE) of silicate coatings on high impact resistance polycarbonates. In AIP Conference Proceedings (Vol. 1336, pp. 303–309) https://doi.org/10.1063/L3586109 Qian Xing. Modeling Mechanisms of Water Affinity and Condensation on Si-based Surfaces via Experiments and Applications, Ph D thesis, Arizona State University (2011).

  28. Nicole Herbots, N. C. Suresh, S.R. Narayan., J. M. Day., H. L.Thinakaran, S. Ram, and Y. Pershad, 2019. Patent Pending.

  29. Nicole Herbots, Justin M Shaw, Q B Hurst, M P Grams, R J Culbertson, David J Smith, V Atluri, P Zimmerman, and K T Queeney. The formation of ordered, ultrathin SiO2/Si (1 0 0) interfaces grown on (1*1) Si (1 0 0). Materials Science and Engineering: B, 87(3):303–316, 2001.

    Article  Google Scholar 

  30. Khavar Dar, Tim Williams, Richard Aitken, Kent L Woods, and Susan Fletcher. Arterial versus capillary sampling for analysing blood gas pressures. Bmj, 310(6971):24–25, 1995.

    Article  CAS  Google Scholar 

  31. Antti AM Luukkonen, Tiina M Lehto, Pirjo SM Hedberg, and Tommy E Vaskivuo. Evaluation of a handheld blood gas analyzer for rapid determination of blood gases, electrolytes and metabolites in intensive care setting. Clinical Chemistry and Laboratory Medicine (CCLM), 54(4):585–594, 2016.

    Article  CAS  Google Scholar 

  32. Hiroto Hatakeyama, Sherry Y Wu, Lingegowda S Mangala, Gabriel Lopez-Berestein, and Anil K Sood. Assessment of in vivo siRNA delivery in cancer mouse models, 2016.

  33. M Mayer. SIMNRA user’s guide, report IPP 9/113Max-Planck-Institut fur Plasmaphysik, Garching, Germany, 1(9):9, 1997.

    Google Scholar 

  34. M Mayer. SIMNRA, a simulation program for the analysis of NRA, RBS and ERDA. In AIP Conference Proceedings, volume 475, pages 541–544, 1999.

    Article  CAS  Google Scholar 

  35. W Eckstein and M Mayer. Rutherford backscattering from layered structures beyond the single scattering model. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 153(1–4):337–344, 1999.

    Article  CAS  Google Scholar 

  36. M Mayer. Ion beam analysis of rough thin films. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 194(2):177–186, 2002

    Article  CAS  Google Scholar 

  37. M Mayer, Kai Arstila, Kai Nordlund, Erik Edelmann, and Juhani Keinonen. Multiple scattering of MeV ions: Comparison between the analytical theory and Monte-Carlo and molecular dynamics simulations. Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 249(1–2):823–827, 2006.

    Article  CAS  Google Scholar 

  38. Luigia Sabbatini and Pier G Zambonin. XPS and SIMS surface chemical analysis of some important classes of polymeric biomaterials. Journal of electron spectroscopy and related phenomena, 81(3):285–301, 1996.

    Article  CAS  Google Scholar 

  39. Ronald T Boro and Sam J Cipolla. Proton-excited X-ray analysis on whole blood using a small accelerator. Nuclear Instruments and Methods, 131(2):343–352.

  40. Wei-Ku Chu, James W. Mayer, and Marc-A Nicolet. Backscattering Spectrometry, Academic Press Inc., San Diego, California, 1997

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Herbots, N., Suresh, N.C., Khanna, S. et al. Optimizing Homogeneous Thin Solid Films (HTSFs) from μL-Blood Droplets via Hyper-Hydrophilic Coatings (HemaDrop™) for Accurate Compositional Analysis via IBA, XRF, and XPS. MRS Advances 4, 2489–2513 (2019). https://doi.org/10.1557/adv.2019.398

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1557/adv.2019.398

Navigation