Skip to main content

Advertisement

Log in

Electrolyte Detection by Ion Beam Analysis, in Continuous Glucose Sensors and in Microliters of Blood using a Homogeneous Thin Solid Film of Blood, HemaDrop™

  • Published:
MRS Advances Aims and scope Submit manuscript

Abstract

Percolation of blood and of interstitial fluids into implantable continuous glucose sensors (CGS) for diabetics presently limits sensor lifetime between 3 and 7 days. Na+ mobile ions in body fluids damage Si-based CGS sensors electronics. The direct detection of Na percolation is investigated by Ion Beam Analysis (IBA) and Proton Induced X-ray Emission (PIXE) in previously used CGS. Based on these results, a new technology called HemaDropTM is then tested to prepare small volume (5-10 µL) of blood for IBA. A species’s detectability by IBA scales with the square of the ratio of element’s atomic number Z to that of the substrate. Because Na has a low atomic number (Z=11), Si signals from sensor substrates can prevent Na detection in Si by 2 mega electron volt (MeV) IBA.

Using 4.7 MeV 23Na (α, α)23Na nuclear resonance (NR) can increase the 23Na scattering cross section and thus its detectability in Si. The NR energy, width, and resonance factor, is calibrated via two well-known alpha (α) particle signals with narrow energy spreads: a 5.486 ± 0.007 MeV 241Am α-source (ΔΕ = 0.12%) and the 3.038 ± 0.003 MeV 16O(α, α)16O NR (ΔΕ = 0.1%). Next, the NR cross section is calibrated via 100 nm NaF thin films on Si(100) by scanning the beam energy. The23Na (α, α) NR energy is found to be 4.696 ± 0.180 MeV, and the NR/RBS cross section 141 ± 7%. This is statistically significant but small compared to the 4.265 MeV 12C NR (1700%) and 3.038 MeV 16O NR (210%), and insufficient to detect small amounts of 23Na in Si. Next, a new method of sample preparation HemaDropTM, is tested for detection of elements in blood, such Fe, Ca, Na, Cl, S, K, C, N, and O, as an alternative to track fluid percolation and Na diffusion in damaged sensors. Detecting more abundant, heavier elements in blood and interstitial fluids can better track fluid percolation and Na+ ions in sensors. Both Na detection and accuracy of measured blood composition by IBA is greatly improved by using HemaDropTM sample preparation to create Homogeneous Thin Solid Films (HTSFs) of blood from 5-10 µL on most substrates. HTSF can be used in vacuo such as 10-8 –10-6 Torr).

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Medtronics. Continuous Glucose Monitoring. [cited 2016, Mar. 17]. www.medtronicdiabetes.com/treatment-and-products/continuous-glucose-monitoring.

  2. Raider, S.I., Gregor, L.V., & Flitsch, R., 1979. Transfer of Mobile Ions from Aqueous Solutions to the Silicon Dioxide Surface. J. of The Electrochemical Society 120(3), 425–431.

    Article  Google Scholar 

  3. Jang J, Lee K 1982. Determination of Positive Mobile Ion Density in SiO2 by Hydrogenated Amorphous Si Field Effect Technique. J. of the Electrochem. Soc.129(12), 2270–2772.

    Article  Google Scholar 

  4. Beuthner, P., Dziesiaty, J., Piprek, P., 1985. Implantation Effect on Transport Properties of Mobile Ions in Silicon Dioxide. Phys. Status solidi (a), 91, K113–K117.

    Article  CAS  Google Scholar 

  5. Imaoka, K., Higashi, M., Shiraiwa, H., Inoue, F., Kajita, T., Sugawa, S., 2009. Mobile-Ion-Induced Charge Loss Failure in Silicon–Oxide–Nitride–Oxide–Silicon Two-Bit Storage Flash Memory. Jap. J. of Appl. Physics 48, 066510 doi:10.1143/JJAP.48.066510.

    Article  Google Scholar 

  6. Carreyrou, J. “Hot Startup Theranos Has Struggled With Its Blood-Test Technology.” WSJ. Wall Street Journal, 16 Oct. 2015. Web. 14 Mar. 2016.

  7. Munro, D. “Walgreens Puts The Brakes On Theranos Roll Out.” Forbes. Forbes Magazine, 24 Oct. 2015. Web. 14 Mar. 2016.

  8. Plebani, M. “Evaluating And Using Innovative Technologies: A Lesson From Theranos?.” Clinic. Chem. & Lab. Med. 53.7 (2015): 961–962. Acad. Search Comp. Web. 14 Mar. 2016.

    Google Scholar 

  9. Several US Patents Pending, Inventors: Herbots, N., Watson C.F, Culbertson E.J., Watson G. P.K., Thilmany, P. R., Krishnan A., Martins, Tempe, I.P.O, Pershad. Y., Gupta. N., Brimhall A., Davis E., Mascareno A., Watson M., Assignee: SiO2 NanoTech LLC.

  10. Herbots N, Xing Q, Hart M, Bradley J, Sell DA, Culbertson RJ, Wilkens B. “IBMM OF OH ADSORBATES AND INTERPHASES ON SI-BASED MATERIALS” Nuclear Instruments and Methods B (NIMB) (2011) 272:330–333

    Article  Google Scholar 

  11. Herbots N, Shaw JM, Hurst QB, Grams MP, Culbertson RJ, Smith DJ, Atluri V, Zimmerman P, Queeney KT. “The Formation of Ordered, Ultrathin Sio2/Si(100) Interfaces Grown On (1×1) Si(100)” Materials Science and Engineering B (2001) 87(3):303–316

    Article  Google Scholar 

  12. Xing Q, Herbots N, Hart M, Bradley J, Wilkens BJ, Sell DA, Sell CH, Kwong HM, Culbertson RJ, Whaley S. “Ion Beam Analysis of Silicon-Based Surfaces and Correlation with Surface Energy Measurements” (2011) Applic. of Accelerators in Research and Industry AIP Conf. Proc. 1336, 201–207 (2011); doi: 10.1063/1.3586089 [accessed Apr 9, 2016].

    CAS  Google Scholar 

  13. Leavitt, J.A., McIntyre, Jr., L.C., Stoss, P., Oder, J.G., Ashbaugh, M.D., Dezfouly-Arjomandy, B., Yang, Z.-M., Lin, Z., 1989. Cross sections for 170.5° backscattering of 4He from carbon for 4He energies between 1.6 and 5.0 MeV. NIMB 40/41, 776–779.

    Article  Google Scholar 

  14. Chu, W., Mayer, J. W., & Nicolet, M., 1978. Backscattering spectrometry. NY: Acad. Press.

    Book  Google Scholar 

  15. Frost, M.C., Meyerhoff, M.E.. In Vivo Chemical Sensors: Tackling Biocompatibility. Analytical Chemistry 78(21), 370–7377; 2006 [cited 2015 Feb 16].

    Article  Google Scholar 

  16. Akers, J.; Setter, S.M.: ‘Pharmacy Times (2007). Ret. 2/4/2008, Web. pharmacytimes.com/issues/articles/2007-05 4633.asp’ Medical Devics & Surgery Technology 2006, pg 335.

  17. Mangus MW, Neric N, Woolson AJ, Herbots N, Culbertson RJ, Wilkens BJ, Causey AW, Brimhall AL, Watson CF, Sinha SA, Acharya AJ. IBA of Materials Used in Hermetic Single-Device Human Implants integrating Bio-sensors with Medical Electronics. 2014 Int. Conf. for Appl. Of Accel. In Res. And Industry, San Antonio, TX, May 2014.

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Pershad, Y., Mascareno, A.A., Watson, M.R. et al. Electrolyte Detection by Ion Beam Analysis, in Continuous Glucose Sensors and in Microliters of Blood using a Homogeneous Thin Solid Film of Blood, HemaDrop™. MRS Advances 1, 2133–2139 (2016). https://doi.org/10.1557/adv.2016.469

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1557/adv.2016.469

Navigation