Abstract
Background
Rapid immunochromatographic tests (ICT) for dengue non-structural protein 1 (NS1) have shown good performance for diagnosing acute-phase dengue in serum in laboratory settings, but rarely have been assessed in whole blood and at point of care (POC). This study compare the accuracy and inter- and intra-observer reliability of the NS1 Bioeasy™ ICT in whole blood at POC versus serum in the laboratory, during a DENV-1 epidemic.
Methods
Cross-sectional study involving 144 adults spontaneously demanding care in an emergency department within 4 days of onset of acute febrile illness. Accuracy of NS1 Bioeasy™ ICT was compared in whole blood and serum, both at 15 and 30 min, blinded to the reference RT-PCR or NS1 ELISA. Non-dengue patients were also tested for Zika virus with RT-PCR. Reliability of whole blood and serum readings by the same or different observers was measured by simple kappa (95% CI).
Results
At 15 min, sensitivity (Sn) of NS1 Bioeasy™ ICT in whole blood/POC was 76.7% (95% CI: 68.0–84.1) and specificity (Sp) was 87.0% (95% CI: 66.4–97.2). Sn in serum/laboratory was 82% (95% CI: 74.1–88.6) and Sp 100% (95% CI: 85.8–100). Positive likelihood ratio was 5.9 (95% CI: 2.0–17.0) for whole blood/POC and 19.8 (95% CI: 2.9–135.1) for serum/laboratory. Reliability of matched readings of whole blood/POC and serum/laboratory by the same observer (k = 0.83, 95% CI: 0.74–0.93) or different observers (k = 0.81, 95% CI: 0.72–0.92) was almost perfect, with higher discordant levels in the absence of dengue. Results did not differ statistically at 5%.
Conclusions
NS1 Bioeasy™ ICT in DENV-1 epidemics is a potentially confirmatory test. Invalid results at 15 min should be reread at 30 min. To optimize impact of implementing ICT in the management of false-negatives it should be incorporated into an algorithm according to setting and available specimen.
Trial registration
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Background
Dengue virus (DENV) is an international public health problem, especially serotype 1 (DENV-1), which shows the highest frequency in the world [1]. The continents with the highest incidence of dengue are America and Asia with 84% of the 390 million cases in the world per year [2]. A systematic analysis on the global burden of dengue suggested that, in 2013, disease costs worldwide may have reached U$ 8.9 billion [3].
Clinical suspicion of dengue is based on the signs and symptoms described in the World Health Organization (WHO) guidelines, i.e., nausea/vomiting, rash, aches and pains, leucopenia, any warning sign (abdominal pain or tenderness, persistent vomiting, clinical fluid accumulation, mucosal bleeding, lethargy/restlessness, liver enlargement, and hemoconcentration concurrent with rapid decrease in platelet count). However, this criterion shows low specificity due to the absence of confirmatory laboratory tests [4].
A study conducted in the United States showed that laboratory diagnostic tests assist health decisions in 60–70% of cases [5]. The World Health Organization (WHO) recommends that ideal laboratory diagnostic tests should be affordable, sensitive, specific, user-friendly, robust, rapid (can be stored at room temperature and interpreted within 30 min), equipment-free, and deliverable to those who need them (ASSURED) [6].
Rapid immunochromatographic tests (ICT) for dengue with the detection of non-structural protein 1 (NS1) meet some of the characteristics of an ASSURED test: high specificity, ranging from 82.8% [7] to 100% [8,9,10,11,12], simple presentations, whether in cassette or strip format, and rapid interpretation of results (15 to 30 min) [13]. However, the sensitivity showed substantial discrepancy, ranging from 55.2% to 94.3% [10, 11].
In addition, the advantages of ICT in cassette format include the possibility of performing the test in whole blood specimens rather than serum and not requiring a specific infrastructure (equipment-free). This allows their deployment in sites with limited resources, for example at point of care (POC) [7, 12,13,14,15]. The tests offer rapid results – allowing the implementation of adequate therapy and care in time to avoid possible complications and even unnecessary costs [16]. Meanwhile, ICT in strip test format requires the use of a centrifuge, reagents, storage in a refrigerator, and test tubes to perform the reaction [7,8,9,10,11,12, 17].
Dengue NS1 ICTs in cassette or strip format have been evaluated in the last decade, mainly for serotypes 1, 2, and 3, with the best performance in the first 5 days of the illness. Still, these studies mostly used serum specimens, whether from biobanks [8, 10, 11] or processed in laboratory settings [8,9,10,11,12]. Lack of the test’s evaluation in whole blood specimens hinders evaluation of ICT performance at POC and the recommendation of its deployment by health systems in epidemic situations [18].
The accuracy and reliability of ICTs for dengue NS1 are essential for guaranteeing quality tests. Still, few published studies on dengue have evaluated agreement between observers with different levels of training [19] or similar levels [14, 20], varying from moderate to almost perfect agreement. No studies were identified that evaluated the reliability of ICTs according to the types of specimens (whole blood versus serum) or settings.
A literature search detected only one study that evaluated duo ICT (NS1/IgM/IgG) and single NS1 using whole blood at POC, with sensitivities of 93.9% (95% CI: 88.8–96.8) and 81.6% (95% CI:74.6–87.1), and specificities of 92% (95% CI: 81.2–96.9) and 98% (95% CI: 89.5–99.7), respectively. Inter-observer reliability proved almost perfect (kappa = 1 95% 95% CI: 0.84–1) [21].
The current study thus aimed to compare the accuracy and reliability of NS1 Bioeasy™ ICT in whole blood at POC and in serum in the laboratory in adults during a DENV-1 epidemic in Rio de Janeiro State, Brazil.
Methods
This was a prospective diagnostic cross-sectional study that evaluated accuracy of the NS1 Bioeasy™ rapid immunochromatographic test and inter- and intra-observer reliability in whole blood specimens at point of care (POC) and serum specimens in the laboratory. The study was conducted according to the Standards for Reporting Studies of Diagnostic Accuracy (STARD) [22].
The study population consisted of patients over 18 years of age with up to 4 days of acute febrile syndrome without an established diagnosis, treated consecutively and by spontaneous demand at an emergency hospital of the Resende Municipal Health Secretariat, Rio de Janeiro State, Brazil, during a dengue epidemic in March 2015.
Data collection
The selected patients answered a questionnaire on clinical signs and symptoms associated with dengue according to WHO 2009 criteria for Dengue diagnosis [4, 23]. All of them signed the free and informed consent form.
Three blood specimens were drawn in vacuum tubes by a nurse and four medical students. Prior asepsis was performed with 70% alcohol, followed by brachial venipuncture with a 25 × 7 mm BD Vacutainer® needle. One tube with EDTA K2 was sent to the hospital laboratory for performing a complete blood count, another with heparin for ICT with whole blood at POC and the third with clotting activator and gel separator for obtaining serum. The latter was transported to the Flavivirus Laboratory, where it remained frozen at −70 °C for subsequent characterization of the specimen and ICT with serum.
The ICTs with whole blood were performed at point of care in the hospital by a nurse, and the ICTs with serum by a biologist and the same nurse (that performed ICT with blood at POC), in the Flavivirus Laboratory of the Oswaldo Cruz Institute, FIOCRUZ, the regional reference laboratory in Rio de Janeiro State. The observers were blinded to the reference test.
Diagnosis
Index test
The index test used was Dengue Eden Test NS1 ICT from the Bioeasy™ company. This is a rectangular cassette with an orifice to add whole blood, serum, or plasma, and which contains an immunochromatographic membrane coated with NS1 antigen on the test line (T) and a window to view the result. When only the control line (C) is visible, the test is considered negative; if lines C and T are visible, the result is positive; and when no line or only the T line is visible, the test result is invalid [13].
The tests were performed according to manufacturer’s instructions and with readings at 15 and 30 min. Tests with invalid results at 15 min were reread at 30 min.
Reference test
All samples were tested using Reverse Transcription Polymerase Chain Reaction (RT-PCR) for dengue and Platelia™ Dengue NS1 Ag-ELISA (Bio-Rad™ Laboratories, France) as reference, performed according to the protocol described by Lanciotti et al. [24] and the manufacturer’s specifications, respectively [25]. Dengue cases were defined by a positive result in at least one of them and non-dengue by negative results in both. Positive specimens according to the reference test were tested by Dengue Serion ELISA classic dengue virus IgM and IgG tests (Virion/Serion, Würzburg, Germany) [26], according to manufacturers’ specifications for characterization of primary and secondary cases, respectively.
Non-dengue patients were also tested for Zika virus with RT-PCR [27] to investigate co-circulation of these viruses.
The reference tests were performed at the Flavivirus Laboratory by two biologists blinded to the index test.
Statistical analysis
The qualitative variables were described by simple frequencies and the quantitative variables by the median with interquartile range (IQR) or minimum and maximum. Shapiro-Wilk test showed rejection of normality for the quantitative variables, indicating the use of a non-parametric test. Verification of association in the dengue and non-dengue groups used the Mann-Whitney non-parametric test and Pearson’s chi-square test for quantitative and qualitative variables, respectively.
To assess the performance of the index test, dengue NS1 ICT, in each setting using whole blood at point of care (POC) and serum in the laboratory in readings at 15′ and 30′, compared to the standard reference, the following measures of accuracy were considered: sensitivity (Sn), specificity (Sp), positive and negative predictive values (PPV, NPV), and positive and negative likelihood ratios (LR+) and (LR-) with the respective 95% confidence intervals (95% CI). Post-test probabilities were calculated for scenarios with prevalence rates of 30%, 50%, and 83.3%, using the Fagan nomogram [28].
The sensitivity and specificity of the index test in different settings were compared for using the χ2 McNemar test [29].
Reliability was assessed by the variability of the ICT results in whole blood/POC and serum/laboratory with readings at 15 min. Inter-observer reliability was analyzed by the readings done by the two professionals (nurse and biologist) and intra-observer reliability by the same professional (nurse), in blinded fashion. The proportions of positive agreement (Ppos) and negative agreement (Pneg) were calculated, as well as simple and prevalence-adjusted bias-adjusted kappa (PABAK) coefficients with respective 95% CI [30]. Kappa values (k) were interpreted according to the classification by Landis & Koch [31]: poor (k < 0.0), slight (0.0 < k < 0.2), fair (0.2 < k < 0.4), moderate (0.4 < k < 0.6), substantial (0.6 < k < 0.8), almost perfect (0.8 < k < 1.00), and perfect agreement (k = 1).
Statistical significance was set at p < 0.05.
The software packages used were R Commander 3.2.1, WinPepi 11.50, and MedCalc 15.8 [32,33,34].
Results
Of 148 patients recruited, four were excluded due to five or more days of fever (Fig. 1). Of the 144 patients included, 52.8% were women. Patients’ median age was 34.5 years, with higher ages in the dengue group (median: 36.1 years, ranging from 18 to 82.4) than in the non-dengue group (median: 27 years, ranging from 18.8 to 62.9), p = 0.019.
Patients presented a median of 1.4 days of fever (95% IQR: 1–2), and the most frequent symptoms were myalgia (91%), headache (89.6%), asthenia (83.3%), and arthralgia (73%). None of the patients evidenced clinical severity signs according to WHO 2009 criteria [4].
Prevalence of the disease was 83.3%. Of the 120 positive cases, 40.3% (n = 58) were primary infections. The dengue virus was characterized by RT-PCR in 105 samples and 15 exclusively by NS1 ELISA. Four samples that were negative for dengue by RT-PCR and NS1 ELISA were positive for Zika virus. One Zika Positive sample tested positive in IgM ELISA.
For all the five invalid cases on readings performed at 15 min in whole blood at POC, readings repeated at 30 min tested positive. For two serum samples at the laboratory showing invalid readings at 15 min, one turned positive at 30 min and the other remained invalid (Fig. 1).
ICT performance was similar in the two settings independently of reading times, both in the point estimate and 95% CI. Sensitivity of ICT with the 15-min reading varied from 76.7% in whole blood at point of care to 82.2% in serum in the laboratory. For the 30-min reading, sensitivity was 78.3% using whole blood at point of care and 84.9% in serum in the laboratory (Table 1). Specificity at 15-was 87% for whole blood at point of care and 100% for serum at the laboratory. For 30-min readings, specificity was 87.5% for whole blood at point of care and 95.8% for serum in the laboratory.
Positive predictive value was high (Table 1) in the prevalence observed in the study (83.3%). In scenarios with prevalence rates of 30% (72.4% CI: 49.3–87.7) or 50% (86%, 95% CI: 69.3–94.3), the PPV remains high. The negative predictive value would increase to 78.9% (95% CI: 72.3–84.3) with a decrease in prevalence in a scenario with 50% prevalence and to 89.7% (95% CI: 85.9–92.6) in a scenario with 30% prevalence.
The reliability between whole blood/POC and serum/laboratory was almost perfect, when evaluated both by the same observer (k = 0.83 CI: 0.74–0.93) and by different observers (k = 0.81 CI: 0.72–0.92), and remained almost perfect even when adjusted by prevalence bias. However, considering the lower limit of Kappa 95% Confidence Interval, reliability interpretation decreased from almost perfect to substantial both for inter-observer (0.73) and for intra-observer 0.75) (Table 2).
Discussion
This study assessed the Bioeasy™ rapid immunochromatographic test for DENV-1 NS1 antigen in a high-prevalence setting with adults in Rio de Janeiro State, Brazil, and found similar performance when testing was compared in whole blood/POC and serum/laboratory.
This was the first study to date comparing the performance of NS1 Bioeasy™ ICT in two scenarios, one in a laboratory setting with professionals trained to perform diagnostic tests and the other in a real-world setting (point of care). The evaluation at POC was conducted in the spontaneous treatment flow in an emergency care unit during an epidemic period, with whole blood samples drawn by different professionals. The accuracy of NS1 Bioeasy™ ICT in whole blood/POC showed sensitivity of 76.7% and specificity of 87% at the 15-minuute reading, similar to that found by Gan et al. [21] for Bioline™ ICT SD with whole blood/POC in a predominantly DENV-2 sample, i.e., Sn = 81.6%, 95% CI: 74.6–87.1 and Sp = 98%, 95% CI: 89.5–99.7.
DENV-1 ICT in serum in the laboratory showed a slight increase in accuracy, i.e., Sn = 82.2% and Sp = 100%. The literature reports variability in sensitivity according to serotype, with better performance in serotypes 1, 2, and 3, varying from 69% [7] to 81.8% [8] in DENV-1, from 62% [7] to 81.8% in DENV-2, and from 61.3% to 81.2% [8] in DENV-3, as compared to variation from 44.5% [23] to 84.8% [8] in DENV-4, so that the tests need to be improved in order to perform independently of serotype.
NS1 ICT manufacturers recommend readings at 15 or 20 min [13, 35]. However, in the current study five cases of whole blood specimens at point of care that were invalid at 15 min turned positive at 30 min, and one of the two invalid results at 15 min with serum in the laboratory turned positive at 30 min. Dussart et al. [8] and Osorio et al. [14] used ICT with serum in the laboratory and also reported improved accuracy at the 30-min reading.
The high number of false-negatives suggests that this ICT should not be used to screen for cases, as recommended even by the manufacturer [13] and Andries et al. [15]. The low number of false-positives and the high likelihood of post-testing in this study suggest that the test could be used for diagnostic confirmation. Robustness of this result is strengthened by the use of composite reference test, comprising PCR Dengue [24] and ELISA Dengue NS1 Platelia. The latter has shown very high specificity (100%) in DENV-1 in a previous study [36].
Inter and intra-observer reliability was almost perfect in this study independently of the type of specimen examined (whole blood or serum), the setting (POC or laboratory), and the level of training of the professionals that performed the test. This result corroborates Gan et al. [21].
Evidence for Zika and Dengue cross reactivity has, so far, been identified for IgM ELISA dengue but not for ELISA NS1 dengue tests [27, 37]. In our study, one of four Zika positive samples showed cross-reactivity with IgM Dengue ELISA, but none cross-react with NS1 ELISA Dengue. Increasingly acute febrile syndromes will demand investigation for several flaviviruses.
A possible limitation to the study was the high prevalence of dengue, but the post-test likelihoods were provided for different prevalence scenarios. The evaluation of a single serotype, DENV-1, may have been a limitation, but it resulted from a consecutive patient sample in an epidemic period. The comparison of whole blood/POC and serum/laboratory precludes other comparisons such as serum/POC and whole blood/laboratory, but it represents the possible routine use in emergency care departments without the necessary infrastructure for processing serum specimens during epidemics.
Considering NS1 Bioeasy™ ICT an ASSURED test [6], these recommendations may be particularly useful at sites without laboratory infrastructure or in hyperendemic scenarios, since they would offer faster turnaround of test results, thus favoring diagnosis in emergencies or remote sites [16].
In health services with some laboratory infrastructure, we recommend the use of NS1 ICT with serum in the laboratory, since it is approximately three times more positive in individuals with dengue than without (LR + = 19.8) when compared to whole blood specimens at point of care (LR + = 5.9). However, in services without laboratory infrastructure or during DEN-V 1 epidemics, we recommend screening according to WHO criteria [4] based on clinical signs followed by NS1 Bioeasy™ ICT in whole blood/POC and reading at 15 min and rereading of invalid results at 30 min.
In addition, investment to improve the performance of these tests in whole blood would allow better diagnostic confirmation, potentially optimizing the flow and efficiency of care and treatment. Training strategies for different levels of health professionals involved in performing and reading the NS1 ICT in whole blood at point of care can guarantee the reliability and adequate implementation in the clinical management of suspected cases.
Conclusions
Bioeasy™ NS1 antigen rapid immunochromatographic test showed a good accuracy during the epidemic of DENV-1, particularly for confirmatory diagnosis. It may be useful at the point of care, with no more than a drop of whole blood, since it is rapid and simple to do by minimal trained health personnel. In this context, Bioeasy™ is relevant for the early diagnosis of dengue, particularly at sites without laboratorial infrastructure and in epidemic or hyperendemic scenarios.
Change history
01 November 2017
An error occurred during the publication of the original article [1].
Abbreviations
- ASSURED:
-
Affordable, sensitive, specific, user-friendly, robust, rapid, equipment-free, and deliverable to those who need them
- C:
-
Control line
- CI:
-
Confidence interval
- DALYs:
-
Disability-adjusted life years
- DENV:
-
Dengue virus
- DENV-1:
-
Dengue virus serotype 1
- DENV-2:
-
Dengue virus serotype 2
- DENV-3:
-
Dengue virus serotype 3
- DENV-4:
-
Dengue virus serotype 4
- EDTA:
-
Ethylenediaminetetraacetic acid
- ELISA:
-
Enzyme-linked immunosorbent assay
- ICR:
-
Immunochromatographic rapid test
- IgG:
-
Immunoglobulin G
- IgM:
-
Immunoglobulin M
- Kappa:
-
Simple Kappa
- LR−:
-
Negative likelihood ratio
- LR+:
-
Positive likelihood ratio
- NPV:
-
Negative predictive value
- NS1:
-
Nonstructural protein-1
- PABAK:
-
Prevalence and bias adjusted kappa
- POC:
-
Point of care
- PPV:
-
Positive predictive value
- RT-PCR:
-
Reverse-transcription polymerase chain reaction
- S:
-
Sensitivity
- Sp:
-
Specificity
- STARD:
-
Standards for Reporting of Diagnostic Accuracy Study
- T:
-
Test line
- WHO:
-
World Health Organization
References
Guzman MG, Halstead SB, Artsob H, Buchy P, Farrar J, Gubler DJ, et al. Dengue: a continuing global threat. Nat Rev Microbiol. 2010;8(12 Suppl):S7–16.
Constenla D, Armien B, Arredondo J, Carabali M, Carrasquilla G, Castro R, et al. Costing dengue fever cases and outbreaks: recommendations from a costing dengue working group in the Americas. Value Health Reg Issues. 2015;8C:80–91.
Shepard DS, Undurradga EA, Halasa YA, Stanaway JD. The global economic burden of dengue: a systematic analysis. Lancet Infect Dis. 2016;16(8):935–41.
World Health Organization. Dengue – Guidelines for diagnosis, treatment, prevention and control. Geneva: WHO; 2009. p. 11. http://www.who.int/tdr/publications/documents/dengue-diagnosis.pdf. Accessed 09 Apr 2016.
Goodman C, Faulkner E, Gould C, Karnes E, Smith A, Aguiar C, et al. The Value of Diagnostics Innovation, Adoption and Diffusion into Health Care. The Lewin Group, Inc. AdvaMed. 2005. https://dx.advamed.org/sites/dx.advamed.org/files/resource/Lewin%20Value%20of%20Diagnos tics%20Report.pdf. Accessed 15 Dec 2016.
Peeling RW, Mabey D. Point-of-care tests for diagnosing infections in the developing world. Clin Microbiol Infect. 2010;16(8):1062–9.
Fry SR, Meyer M, Semple MG, Simmons CP, Sekaran SD, Huang JX, et al. The diagnostic sensitivity of dengue rapid test assays is significantly enhanced by using a combined antigen and antibody testing approach. PLoS Negl Trop Dis. 2011;5(6):e1199.
Dussart P, Petit L, Labeau B, Bremand L, Leduc A, Moua D, et al. Evaluation of two new commercial tests for the diagnosis of acute dengue virus infection using NS1 antigen detection in human serum. PLoS Negl Trop Dis. 2008;2(8):e280.
Pok KY, Lai YL, Sng J, Ng LC. Evaluation of nonstructural 1 antigen assays for the diagnosis and surveillance of dengue in Singapore. Vector-Borne and Zoonotic Dis. 2010;10(10):1009–16.
Tricou V, Vu HTT, Quynh NVN, Nguyen CW, Tran CT, Farrar J, et al. Comparison of two dengue NS1 rapid tests for sensitivity, specificity and relationship to viraemia and antibody responses. BMC Infect Dis. 2010;10:142.
Lima MdRQ, Nogueira RMR, Schatzmayr HG, Santos FBd. Comparison of three commercially available dengue NS1 antigen capture assays for acute diagnosis of dengue in Brazil. PLoS Negl Trop Dis. 2010;4(7):e738.
Ferraz FO, Bomfim MRQ, Totola AH, Ávila TV, Cisalpino D, Pessanha JEM, et al. Evaluation of laboratory tests for dengue diagnosis in clinical specimens from consecutive patients with suspected dengue in Belo Horizonte, Brazil. J Clin Virol. 2013;58:41–6.
Dengue Eden Test Bioeasy®. Korea: Bioeasy Diagnóstica Ltda. 2015. http://www.masterdiagnostica.com.br/public/files/produtos/1394214631139421463132631 16009.pdf. Accessed 6 May 2017.
Osorio L, Ramirez M, Bonelo A, Villar LA, Parra B. Comparison of the diagnostic accuracy of commercial NS1-based diagnostic tests for early dengue infection. Virol J. 2010;7:361.
Andries AC, Duong V, Ngan C, Ong S, Huy R, Sroin KK, et al. Field evaluation and impact on clinical management of a rapid diagnostic kit that detects dengue NS1, IgM and IgG. PLoS Negl Trop Dis. 2012;6(12):e1993.
Pai NP, Vadnais C, Denkinger C, Engel N, Pai M. Point-of-care testing for infectious diseases: diversity, complexity, and barriers in low- and middle-income countries. PLoS Med. 2012;9(9):e1001306.
Ty Hang V, Minh Nguyet N, The Trung D, Tricou V, Yoksan S, Dung NM, et al. Diagnostic accuracy of NS1 ELISA and lateral flow rapid tests for dengue sensitivity, specificity and relationship to viraemia and antibody responses. PLoS Negl Trop Dis. 2009;3(1):e360.
Palamountain KM, Baker J, Cowan LP, Essajee S, Mazzola LT, Metzler M, et al. Perspectives on introduction and implementation of new point-of-care diagnostic tests. J Infect Dis. 2012;205(Suppl 2):S181–90.
Martínez-Veja RA, Díaz-Quijano FA, Coronel-Ruiz C, Gómez SY, Villar-Centeno LA. Evaluación de la utilidad de la prueba rápida de casete por imunocromatografía para el diagnóstico de dengue en una región endémica colombiana. Biomedica. 2009;29:616–24.
Hunsperger EA, Yoksan S, Buchy P, Nguyem VC, Sekaran SD, Enria DA, et al. Evaluation of commercially available diagnostic tests for the detection of dengue virus NS1 antigen and anti-dengue virus IgM antibody. PLoS Negl Trop Dis. 2014;8(10):e3171.
Gan VC, Tan L-K, Lye DC, Pok K-Y, Mok S-Q, Chua RC-R, et al. Diagnosing dengue at the point-of-care: utility of a rapid combined diagnostic kit in Singapore. PLoS ONE. 2014;9(3):e90037.
Korevaar DA, Wang J, van Enst WA, Leeflang MM, Hooft L, Smidt N, et al. Reporting diagnostic accuracy studies: some improvements after 10 years of STARD. Radiology. 2015;274:781–9.
Buonora SN, Passos SRL, do Carmo CN, Quintela FM, de Oliveira DNR, dos Santos FB, et al. Accuracy of clinical criteria and an immunochromatographic strip test for dengue diagnosis in a DENV-4 epidemic. BMC Infect Dis. 2016;16:37.
Lanciotti RS, Calisher CH, Gubler DJ, Chang GJ, Vorndam AV. Rapid detection and typing of dengue viruses from clinical samples by using reverse transcriptase-polymerase chain reaction. J Clin Microbiol. 1992;30(3):545–51.
Platelia™ Dengue NS1 AG®. France: Bio-Rad Laboratories, Inc.. 2008. http://www.biorad.com/webroot/web/pdf/inserts/CDG/en/72830_881149_EN.pdf. Accessed 6 May 2017.
Lanciotti RS, Kosoy OL, Laven JJ, Velez JO, Lambert AJ, Johnson AJ, et al. Genetic and serologic properties of Zika virus associated with an epidemic, Yap State, Micronesia, 2007. Emerg Infect Dis. 2008;14(8):1232–9.
Serion Elisa Classic Dengue Virus IgG/IgM Virion\Serion®. Germany: Institut Virion\Serion GmbH. 2012. http://www.virionserion. de/download/flyer/Flyer_ELISA_classic_Dengue_Virus_EN.pdf. Accessed 6 May 2017.
Fagan TJ. Nomogram for Bayes theorem [letter]. N Engl J Med. 1975;293(5):257.
Trajman A, Luiz RR. McNemar χ2 test revisited: comparing sensitivity and specificity of diagnostic examinations. Scand J Clin Lab Invest. 2007;68(1):77–80.
Byrt T, Bishop J, Carlin JB. Bias, prevalence and kappa. J Clin Epidemiol. 1993;46(5):423–9.
Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159–74.
R Core Team. R: A language and environment for statistical computing. Vienna: R Foundation for Statistical Computing; 2015. https://www.R-project.org/.
Abramson JH. WINPEPI updated: computer programs for epidemiologists, and their teaching potential. Epidemiol Perspect Innov. 2011;8:1.
MedCalc Statistical Software version 12.7.8. Ostend: MedCalc Software bvba; 2014. https://www.medcalc.org.
IVB Dengue Ag NS1 OrangeLife®. Brazil: Orangelife comércio e indústria Ltda. 2016. http://www.dmed.com.br/manuais/denguens1orangelife.pdf. Accessed 07 Jan 2016.
Pal S, Dauner AL, Mitra I, Forshey BM, Garcia P, Morrison AC, Halsey ES, Kochel TJ, Wu SJ. Evaluation of dengue NS1 antigen rapid tests and Elisa kits using clinical samples. PLoS One. 2014;9(11):e113411.
Matheus S, Boukhari R, Labeau B, Ernault V, Bremand L, Kazanji M, Rousset D. Specificity of Dengue NS1 antigen in differential diagnosis of dengue and Zika virus infection. Emerg Infect Dis. 2016;22(9):1691–2.
Acknowledgements
The authors wish to thank for field work, Rogério Ferrari, Ana Victória Palagi Viganó, Indiara Firmino Brito and Letícia Borges dos Santos.
Funding
This study was supported by the National Council for Scientific and Technological Development (CNPq) [grant numbers 401,366/2013–8] - Brazilian Network for Technological and Health Assessment (REBRATS), Excellence Program in Clinical Research of the Oswaldo Cruz Foundation (PROEP/PCLIN) [grant numbers N° 402,068/2012–2] and the Foundation for Research Support in the State of Rio de Janeiro (FAPERJ) [grant numbers E-25/110.188/2014]. Fellowship grant from CNPq [grant numbers 311,414/2013–3] and UNESA - Productivity Research to [SRLP] and scholarship from FAPERJ [grant numbers 221,354 E_01/2016] to [VEM].
Availability of data and materials
The datasets analyzed during the current study is available from the corresponding author upon request.
Author information
Authors and Affiliations
Contributions
VEM conceived of the study, participated in its design, data collection, recording, carried out rapid immunochromatographic test and coordination and helped to draft the manuscript. SRLP conceived of the study, participated in its design and coordination and helped to draft the manuscript. YHMH and MABS participated in the study design, data analysis and interpretation and helped to draft the manuscript. GMB, LVBF and ACFSRM participated in its design, data collection and recording. CDSR and ASS carried out the immunoassays, viral characterization and writing the manuscript. RVCO participated in the design of the study and performed the statistical analysis. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The study was approved by the Institutional Review Board of the Evandro Chagas Clinical Research Institute (CAAE 0066.0.009.000-11). All patients signed the free and informed consent form.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Additional information
The original article has been updated due to a technical error causing certain portions of text to appear in Portuguese. More information on what has been amended can be found in the Correction notice for this article.
A correction to this article is available online at https://doi.org/10.1186/s12879-017-2804-z.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Mata, V.E., Passos, S.R.L., Hökerberg, Y.H.M. et al. Accuracy and reliability of an NS1 rapid immunochromatographic test for DENV-1 diagnosis at point of care and in the laboratory. BMC Infect Dis 17, 594 (2017). https://doi.org/10.1186/s12879-017-2679-z
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12879-017-2679-z