Background
Adequate sagittal plane motion of the first metatarsalphalangeal joint (1st MTPJ) is important during normal gait and goniometric measurement is commonly used as a diagnostic and outcome assessment tool for 1st MTPJ pathology. We aimed to determine the intra and inter-rater reliability of a universal plastic goniometer and a smartphone application when measuring dorsiflexion at the 1st MTPJ.
Method
A double-blind repeated measures design was used (three raters, one observer) to measure 1st MTPJ dorsiflexion in both feet of 32 healthy volunteers (64 data sets) using a pre-defined measurement protocol. Reliability of the smartphone app (Dr. Goniometer v1.8 installed on an iphone 4s) and a standard universal goniometer was assessed using Interclass correlation coefficients (ICCs) with 95% confidence intervals.
Results
Both instruments demonstrated excellent intra-rater reliability (ICC> 0.75), with moderate to good inter-rater reliability:
-
Universal Goniometer
-
Inter-rater reliability ICC 0.693 (95% CI 0.580 – 0.788)
-
Smartphone app
-
Inter-rater reliability ICC 0.708 (95% CI 0.597 - 0.799)
Conclusions
Moderate to high intra and inter-rater reliability of passive 1st MTPJ motion can be achieved with traditional and smartphone-based goniometric measurement. Smartphone applications may provide a slightly higher reliability, but devices should not be used inter-changeably as significant variations in measurement between devices may occur.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
This article is published under an open access license. Please check the 'Copyright Information' section either on this page or in the PDF for details of this license and what re-use is permitted. If your intended use exceeds what is permitted by the license or if you are unable to locate the licence and re-use information, please contact the Rights and Permissions team.
About this article
Cite this article
Hales, G., Keating, R., Bear, N. et al. Reliability of a smartphone goniometer app compared with traditional goniometer for measuring passive motion at the first metatarsophalangeal joint. J Foot Ankle Res 8 (Suppl 2), P12 (2015). https://doi.org/10.1186/1757-1146-8-S2-P12
Published: