Change in diaphragm and intercostal muscle thickness in mechanically ventilated patients: a prospective observational ultrasonography study
Abstract
Background
Diaphragm atrophy is observed in mechanically ventilated patients. However, the atrophy is not investigated in other respiratory muscles. Therefore, we conducted a two-center prospective observational study to evaluate changes in diaphragm and intercostal muscle thickness in mechanically ventilated patients.
Methods
Consecutive adult patients who were expected to be mechanically ventilated longer than 48 h in the ICU were enrolled. Diaphragm and intercostal muscle thickness were measured on days 1, 3, 5, and 7 with ultrasonography. The primary outcome was the direction of change in muscle thickness, and the secondary outcomes were the relationship of changes in muscle thickness with patient characteristics.
Results
Eighty patients (54 males and 26 females; mean age, 68 ± 14 years) were enrolled. Diaphragm muscle thickness decreased, increased, and remained unchanged in 50 (63%), 15 (19%), and 15 (19%) patients, respectively. Intercostal muscle thickness decreased, increased, and remained unchanged in 48 (60%), 15 (19%), and 17 (21%) patients, respectively. Decreased diaphragm or intercostal muscle thickness was associated with prolonged mechanical ventilation (median difference (MD), 3 days; 95% CI (confidence interval), 1–7 and MD, 3 days; 95% CI, 1–7, respectively) and length of ICU stay (MD, 3 days; 95% CI, 1–7 and MD, 3 days; 95% CI, 1–7, respectively) compared with the unchanged group. After adjusting for sex, age, and APACHE II score, they were still associated with prolonged mechanical ventilation (hazard ratio (HR), 4.19; 95% CI, 2.14–7.93 and HR, 2.87; 95% CI, 1.53–5.21, respectively) and length of ICU stay (HR, 3.44; 95% CI, 1.77–6.45 and HR, 2.58; 95% CI, 1.39–4.63, respectively) compared with the unchanged group.
Conclusions
Decreased diaphragm and intercostal muscle thickness were frequently seen in patients under mechanical ventilation. They were associated with prolonged mechanical ventilation and length of ICU stay.
Trial registration
UMIN000031316. Registered on 15 February 2018
Keywords
Diaphragm Intercostal muscle Atrophy UltrasonographyAbbreviations
- ACV
Assist-control ventilation
- BMI
Body mass index
- CI
Confidence interval
- HFNC
High-flow nasal cannula
- ICU
Intensive care unit
- IQR
Interquartile range
- NPPV
Noninvasive positive pressure ventilation
- VS
Versus
Background
Critically ill patients experience atrophy of various muscles including respiratory muscles, with diaphragm muscle atrophy garnering increasing attention. Patients undergoing prolonged mechanical ventilation experience sustained diaphragm muscle loss, which leads to worse clinical outcomes [1].
Mechanical ventilation can affect not only the diaphragm but also the intercostal muscles, as demonstrated in animal studies [2]. Although diaphragm is the main respiratory muscle, several other respiratory muscles contribute to the respiratory effort as well. Specifically, when respiratory workload increases, other respiratory muscles are more active than the diaphragm [3]. Among these, intercostal muscles are one of the most important respiratory muscles in manipulating the movement of the rib cage [4].
Although diaphragm atrophy was reported to affect clinical outcomes in mechanical ventilation [1, 5], changes in other respiratory muscles during mechanical ventilation are not well investigated. Since diaphragm and intercostal muscles play an important role in critically ill patients, it is reasonable to hypothesize that the thickness of both diaphragm and intercostal muscles could change over the mechanical ventilation. Therefore, we evaluated changes in diaphragm and intercostal muscle thickness and the potential relationship with patient characteristics, medications, ventilator mode, duration of mechanical ventilation, and length of intensive care unit (ICU) stay in patients undergoing mechanical ventilation.
Methods
Study design
This two-center prospective observational study was conducted in the mixed medical/surgical ICUs of Tokushima University Hospital and Tokushima Prefectural Central Hospital between June 2016 and June 2018 (Additional file 1: Table S1). This study was approved by both clinical research ethics committees at Tokushima University Hospital (approval number 2593) and Tokushima Prefectural Central Hospital (approval number 1739). This trial was registered on a clinical trial (UMIN-Clinical Trials Registry: 000031316). At the time of enrollment, written informed consent was obtained from patients or their authorized surrogate decision makers.
Study population
Consecutive adult patients who were expected to be mechanically ventilated longer than 48 h were enrolled in this study. Patients were recruited prospectively within 24 h following ICU admission. Patients who met the following criteria were excluded: age under 18 years, trauma or chest tube at the measurement point, diagnosis of primary neuromuscular disease.
Measurements of diaphragm and intercostal muscle thickness
Imaging was performed with B-mode ultrasound using liner transducers. Diaphragm and intercostal muscle thickness were evaluated with serial ultrasound measurements on days 1, 3, 5, and 7. Recordings were discontinued at extubation, discharge from the ICU, or death of the patient, whichever occurs first.
All measurements were performed with patients in supine anatomical position, reclining in bed at a 30-degree angle. The transducer was perpendicularly placed on the right chest wall at the zones of apposition: between the eighth and tenth intercostal spaces, between the antero-axillary and the midaxillary lines, and 0.5–2 cm below the costophrenic sinus, as previously reported (Additional file 1: Figure S1) [6]. A mark was drawn on the patient’s skin to ensure consistency. Diaphragm and intercostal muscle thickness were measured three times at end expiration, and the mean value was recorded for evaluation. In this area, the diaphragm is observed as a three-layered structure, with the hypoechogenic muscular layer bordered by echogenic layers—the peritoneum and the diaphragmatic pleurae (Additional file 1: Figure S2). Thickening fraction of the diaphragm was calculated as [thickness at inspiration − thickness at expiration]/ [thickness at expiration] × 100 [7]. As intercostal muscles comprise internal and external intercostal muscle fibers that are difficult to discriminate by ultrasonography, intercostal muscles including the internal and external intercostal muscles were measured at the zone of apposition. In order to blind the data analysis from the patient status, the image was stored in ultrasound machine at the bedside and then, the same investigator measured the muscle thickness which was blinded from patients’ name and measurement days. Before commencing the study, intra- and inter-observer reproducibility were 0.92 and 0.96 for the diaphragm and 0.92 and 0.90 for the intercostal muscles, as assessed by two ICU physicians (Additional file 1: Figures S3-S6).
Definitions
The study population was categorized into decreased thickness, increased thickness, and unchanged group, respectively in diaphragm and intercostal muscles. These categories were based on the change in muscle thickness over the study period, based on previous studies [1, 5, 8]. The decreased thickness group was defined as those exhibiting more than 10% decrease in thickness from day one to the lowest value over the measurement period. The increased thickness group was defined as more than 10% increase in thickness without more than 10% decrease, and the remaining patients were categorized into the unchanged group.
Outcomes
The primary outcome was direction and rate of changes in the diaphragm and intercostal muscle thickness. The thickness change was defined as percent variation in muscle thickness compared with the admission day. The secondary outcomes were the relationship of changes in muscle thickness with patient characteristics; ventilator mode; duration of mechanical ventilation; length of ICU and hospital stay; reintubation and tracheostomy; the use of high-flow nasal cannula (HFNC) and noninvasive positive pressure ventilation (NPPV) after extubation; and medications including analgesics, sedatives, catecholamines, steroids, and muscle relaxants.
Statistical analysis
No articles reporting on intercostal muscle atrophy in the ICU were found by the literature search. Therefore, a feasible sample size of 80 patients was planned for enrollment based on two studies on diaphragm atrophy (the average of 54 and 107 patients in these studies) [5, 8]. Continuous data were presented as means ± standard deviation or medians (interquartile range (IQR)), whereas categorical data were expressed as numbers (%). Changes in muscle thickness over time were assessed using a linear mixed model for repeated measures: statistical significance of changes in muscle thickness at each time point was also tested using 95% confidence intervals (CIs), with intervals not including zero considered as statistically significant. Kappa statistics was used to evaluate the relationship between changes in diaphragm and intercostal muscle thickness. Variables were compared using one-way analysis of variance or the Kruskal-Wallis test. Post hoc correction for multiple comparisons was performed with Dunnett’s or Steel’s test to compare decreased or increased thickness group versus unchanged group. Continuous outcomes were evaluated using Cox regression analysis, adjusted for age, sex, and APACHE II score. We presented cumulative incidence of liberation from mechanical ventilation, in which death was treated as a competing risk. Data were compared using Gray’s test with Bonferroni correction for two pairwise comparisons (significant at p < 0.025 vs. unchanged group). Data analyses were conducted using JMP 13.1.0 and SAS 9.4 (SAS Institute, Cary, NC). All statistical tests were two-tailed and a p value < 0.05 was regarded as statistically significant.
Results
Patient characteristics
Patient characteristics
Characteristics | Overall (n = 80) |
---|---|
Age, mean ±SD, year | 68 ± 14 |
Male/female | 54/26 |
Body mass index, mean ± SD, kg/m2 | 24 ± 4 |
APACHE II score | 24 (19–30) |
SOFA, mean in the first 3 days | 9 (5–12) |
Sepsis (sepsis-3 criteria), n (%) | 31 (39) |
ICU admission reasons, n (%) | |
Respiratory failure | 22 (28) |
Post-cardiac surgery | 15 (19) |
Heart failure | 8 (10) |
Sepsis, nonrespiratory | 7 (9) |
Stroke | 7 (9) |
Cardiac arrest | 5 (6) |
Traumas | 2 (3) |
Others | 14 (18) |
Main outcomes
Time course of the diaphragm and intercostal muscle thickness. Time course for the measurement of the diaphragm and intercostal muscle thickness over the first 7 days of mechanical ventilation. The horizontal line represents the time from admission to the intensive care unit (ICU), and the vertical line represents the change in diaphragm and intercostal muscle thickness. Solid lines represent the changes in diaphragm muscle thickness, and dotted lines represent the changes in intercostal muscle thickness. Data are expressed as means and 95% confidence intervals
Throughout the study period, the intercostal muscle thickness decreased, increased, and remained unchanged in 48 (60%), 15 (19%), and 17 (21%) patients, respectively. In the decreased thickness group, it decreased by 15.4% (95% CI, 11.5–19.2%), 20.9% (95% CI, 16.5–25.4%), and 19.8% (95% CI, 14.6–25.0%) on days 3, 5, and 7 (p < 0.01). In the increased thickness group, it increased by 25.7% (95% CI, 15.8–35.6%), 28.4% (95% CI, 16.5–40.3%), and 28.6% (95% CI, 10.4–46.8%) on days 3, 5, and 7 (p < 0.01). Finally, in the unchanged group, it changed by 0.5% (95% CI, − 2.2–1.2%), 2.9% (95% CI, − 0.5–6.4%), and − 1.3% (95% CI, − 5.2–2.7%) on days 3, 5, and 7 (p = 0.26).
The relationship between the diaphragm and the intercostal muscle
The relationship of changes in muscle thickness between the diaphragm and the intercostal muscles. Among groups stratified according to changes in diaphragm thickness, the percentages of patients with decreased, increased, or unchanged intercostal muscle thickness are shown by a bar graph. Numbers indicate the number of patients in each group. Changes in diaphragm and intercostal muscles thickness were associated with a kappa value 0.28 (95% confidence interval, 0.14–0.41, p < 0.001), suggesting a poor association
The relationship of changes in muscle thickness with ventilator mode
The relationship of changes in diaphragm and intercostal muscle thickness with patient characteristics, medications, and ventilator mode
Variables | Change in diaphragm thickness | Change in intercostal muscle thickness | ||||||
---|---|---|---|---|---|---|---|---|
Decreased thickness (n = 50) | Unchanged (n = 15) | Increased thickness (n = 15) | p value | Decreased thickness (n = 48) | Unchanged (n = 17) | Increased thickness (n = 15) | p value | |
Age, mean ± SD, year | 69 ± 2 | 69 ± 3 | 61 ± 3 | 0.10 | 70 ± 2 | 64 ± 3 | 66 ± 3 | 0.32 |
Male/female | 34/16 | 10/5 | 10/5 | 0.99 | 32/16 | 13/4 | 9/6 | 0.60 |
Body mass index, mean ± SD, kg/m2 | 23 ± 1 | 24 ± 1 | 24 ± 1 | 0.72 | 23 ± 1a | 26 ± 1 | 24 ± 1 | < 0.01 |
APACHE II score | 23 (18–29) | 22 (20–31) | 26 (22–32) | 0.62 | 25 (17–30) | 22 (17–27) | 26 (22–32) | 0.35 |
SOFA, mean in the first 3 days | 9 (6–12) | 8 (4–9) | 10 (5–14) | 0.33 | 9 (5–12) | 7 (5–9) | 9 (6–14) | 0.22 |
Sepsis (sepsis-3 criteria), n (%) | 22 (44) | 5 (33) | 4 (27) | 0.43 | 18 (38) | 6 (35) | 7 (47) | 0.77 |
Surgical admissions, n (%) | 19 (38) | 3 (20) | 6 (40) | 0.40 | 18 (38) | 3 (18) | 7 (47) | 0.19 |
Medications, n (%) | ||||||||
Catecholamine* | 35 (70) | 11 (73) | 10 (67) | 0.92 | 35 (73) | 10 (59) | 11 (73) | 0.53 |
Neuromuscular blocking agents† | 7 (14) | 1 (7) | 0 (0) | 0.25 | 6 (13) | 1 (6) | 1 (7) | 0.66 |
Steroids‡ | 18 (36) | 5 (33) | 2 (13) | 0.25 | 18 (38) | 3 (18) | 4 (27) | 0.29 |
Aminoglycoside | 1 (2) | 1 (7) | 0 (0) | 0.47 | 1 (2) | 0 (0) | 1 (7) | 0.46 |
Opioid | 44 (88) | 13 (87) | 11 (73) | 0.37 | 42 (88) | 16 (94) | 10 (67) | 0.07 |
Midazolam | 18 (36) | 7 (46) | 7 (47) | 0.64 | 20 (42) | 7 (41) | 5 (33) | 0.84 |
Dexmedetomidine | 24 (48) | 6 (40) | 4 (27) | 0.33 | 23 (48) | 4 (24) | 7 (47) | 0.20 |
Propofol | 8 (16) | 3 (20) | 6 (40) | 0.14 | 7 (15) | 7 (41) | 3 (20) | 0.07 |
Ventilatory settings during first 3 days | ||||||||
Controlled (ACV)/partial assist (PSV) | 47/3 | 14/1 | 14/1 | 0.96 | 46/2 | 15/2 | 14/1 | 0.76 |
Set inspiratory pressure above PEEP | 12 (10–12) | 10 (10–12) | 12 (10–14) | 0.28 | 12 (10–13)a | 10 (10–12) | 12 (10–14)a | 0.02 |
PEEP, cmH2O | 8 (6–10) | 8 (6–10) | 6 (6–8) | 0.31 | 8 (6–10) | 8 (7–11) | 8 (6–8) | 0.11 |
Tidal volume/PBW, mL/kg | 8.2 (7.3–9.6) | 8.6 (7.6–10.2) | 7.8 (7.2–9.7) | 0.52 | 8.2 (7.2–9.7) | 8.0 (7.8–9.7) | 8.1 (7.2–9.7) | 0.90 |
Patient outcomes
Outcomes
Outcomes | Change in diaphragm thickness | Change in intercostal muscle thickness | ||||||
---|---|---|---|---|---|---|---|---|
Decreased thickness (n = 50) | Unchanged (n = 15) | Increased thickness (n = 15) | p value | Decreased thickness (n = 48) | Unchanged (n = 17) | Increased thickness (n = 15) | p value | |
Duration of mechanical ventilation, day | 7 (5–15)a | 4 (3–5) | 7 (4–12)a | < 0.01 | 8 (5–17)a | 4 (3–6) | 6 (5–8) | < 0.01 |
Length of ICU stay, day | 10 (6–16)a | 5 (5–7) | 8 (5–15) | < 0.01 | 10 (6–18)a | 5 (4–9) | 7 (5–10) | < 0.01 |
Length of hospital stay, day | 34 (22–54) | 32 (13–118) | 27 (11–65) | 0.71 | 34 (22–72) | 29 (13–46) | 34 (16–65) | 0.37 |
Reintubation, n (%) | 9 (18) | 1 (7) | 1 (7) | 0.36 | 10 (21) | 1 (6) | 0 (0) | 0.07 |
Tracheostomy, n (%) | 12 (24) | 1 (7) | 2 (13) | 0.27 | 12 (25) | 1 (6) | 2 (13) | 0.19 |
The use of HFNC, n (%) | 29 (58) | 8 (53) | 7 (47) | 0.73 | 27 (56) | 9 (53) | 8 (53) | 0.96 |
The use of NPPV, n (%) | 3 (6) | 0 (0) | 0 (0) | 0.39 | 2 (4) | 1 (6) | 0 (0) | 0.66 |
Mortality in the ICU, n (%) | 10 (20) | 0 (0) | 1 (7) | 0.10 | 8 (17) | 0 (0) | 3 (20) | 0.17 |
Mortality in the hospital, n (%) | 18 (36) | 1 (7) | 5 (33) | 0.09 | 17 (35) | 1 (6) | 6 (40) | 0.048 |
Outcomes by multivariate analysis
Outcomes | Change in diaphragm thickness | Change in intercostal muscle thickness | ||
---|---|---|---|---|
Decreased thickness vs. unchanged | Increased thickness vs. unchanged | Decreased thickness vs. unchanged | Increased thickness vs. unchanged | |
Duration of mechanical ventilation, day | 4.19 (2.14–7.93)* | 2.38 (1.08–5.29)† | 2.87 (1.53–5.21)* | 1.71 (0.79–3.81) |
Length of ICU stay, day | 3.44 (1.77–6.45)* | 1.99 (0.92–4.39) | 2.58 (1.39–4.63)* | 1.43 (0.66–3.16) |
Length of hospital stay, day | 1.34 (0.66–2.60) | 0.89 (0.36–2.28) | 2.04 (1.06–3.81)† | 2.21 (0.94–5.61) |
Cumulative incidence of liberation from mechanical ventilation by diaphragm or intercostal muscle thickness changes during the first week. Data were compared using Gray’s test with Bonferroni correction for two pairwise comparisons (significant at p < 0.025 vs. unchanged group). Death was treated as a competing risk. The horizontal line represents the time from admission to the intensive care unit (ICU), and the vertical line represents the cumulative incidence of liberation from mechanical ventilation. a Diaphragm: compared with unchanged, those with decreased or increased diaphragm thickness had a lower cumulative incidence of liberation from mechanical ventilation (p < 0.01 in both group). b Intercostal muscle: compared with unchanged, those with decreased diaphragm thickness had a lower cumulative incidence of liberation from mechanical ventilation (p = 0.018). VS, versus. *Significant at < 0.025 vs. unchanged group by Bonferroni correction
Discussion
This prospective observational ultrasonography study investigated the time course of change in diaphragm and intercostal muscle thickness daily for up to 7 days in mechanically ventilated patients. The diaphragm muscle thickness can decrease, increase, or remain unchanged during mechanical ventilation. A similar change was observed in intercostal muscles, which are also essential respiratory muscles. Our study found that atrophy of diaphragm or intercostal muscles was associated with prolonged mechanical ventilation and length of ICU stay.
While the diaphragm is the main respiratory muscle, accounting for 60–80% of the inspiratory work [9], other respiratory muscles also play an important role in moving the rib cage, especially in cases where the diaphragm function deteriorates or work associated with breathing increases [3]. Several studies detected rib cage muscle activity at high inspiratory efforts by electromyography [10, 11]. In one study, the activity of rib cage muscles exceeded that of the diaphragm during increased inspiratory workload [9]. Rib cage muscles comprise intercostal, sternocleidomastoid, scalene, pectoralis, serratus, oblique, transverse abdominis, and rectus abdominis muscles. Among these, the intercostal muscles are the most important respiratory muscles that significantly contribute to the movement of the rib cage [9]. In patients with chronic obstructive pulmonary disease, intercostal muscle atrophy occurs in advanced stages [12], and the decreased mass of intercostal muscles is associated with exacerbation [13]. Our findings highlight the important role of intercostal muscles in critically ill patients and suggest a possible association of their atrophy with difficult weaning from mechanical ventilation and prolonged ICU stay.
Several studies reported on rib cage muscle atrophy under mechanical ventilation. In one study, mechanical ventilation led to intercostal muscle atrophy as well as diaphragm muscle in animals [2]. Capdevila et al. suggested that 48 h of mechanical ventilation led to a 29% decrease in intercostal muscle mass in rabbits [2]. Another study by Levine et al. investigating atrophy of the pectoralis major muscle and the diaphragm in mechanically ventilated patients [14] found that the muscle fibers of the pectoralis major muscle did not decrease, although the cross-sectional area of the diaphragm muscle fibers was decreased. However, that study compared a limited number of 22 patients with different pathologies. In contrast, Nakanishi et al. reported that biceps brachii, which can also function to expand rib cage, showed significant progressive atrophy in critically ill patients [15]. Thus, rib cage muscles can undergo atrophy in critically ill patients.
In contrast to previous studies reporting the incidence of diaphragm atrophy at 41–44% [1, 5], 63% of the patients in the current study had decreased diaphragm thickness, which we propose might be due to two potential explanations. First, majority of the patients in the current study were controlled by ACV, which was previously shown to be associated with excessive respiratory support and neglected spontaneous breathing [16, 17]. Therefore, an average thickening fraction was 8% in the current study, which should be approximately 25% in individuals with normal breathings. Second, the mean patient age was higher in the current study compared with previous studies (68 vs. 59–60 years), which is associated with prominent muscle atrophy and decreased functional status [18]. As previously reported, we also found that decreased diaphragm thickness was associated with the duration of mechanical ventilation and the length of ICU stay, and diaphragm thickness decreased quickly in the early phase after mechanical ventilation [1, 8].
The increased diaphragm thickness was observed in 19% of the patients and remained unchanged in another 19%, which were previously reported as 12–24% and 35–44%, respectively [1, 5]. Increased diaphragm thickness was associated with prolonged mechanical ventilation. Increased diaphragm thickness has been garnering increased interest based on a study by Goligher et al. who recently reported that increased diaphragm thickness predicted prolonged mechanical ventilation (OR, 1.38; 95% CI, 1.00–1.90). The rate of increased diaphragm thickness in the current study is comparable with those reported in previous studies [1, 5]. However, increased diaphragm thickness was not significantly associated with excessive thickening fraction as previously reported [5]. The potential reason is the limited number of high thickening fraction, defined as more than 25% thickening (2.5% in our study population vs. 12.3% in Goligher’s study) [1]. Despite the relatively lower thickening fraction in the current study, increased thickness was observed in 15 patients, suggesting increased diaphragm thickness is likely multifactorial. The clinical significance of increased diaphragm thickness is unclear in this research because of the relatively lower number of increased thickness in the study cohort (15 with increased thickness vs. 50 patients with atrophy). Furthermore, most studies on diaphragm focuses on atrophy and not increased thickness [8, 17, 19, 20], reflected on the paucity of data on its increased thickness [1, 5]. Increased diaphragm thickness needs further investigation.
The changes in the diaphragm and intercostal muscle thickness showed a poor association (κ = 0.27, p < 0.01). Although the association was poor, the atrophy of both the diaphragm and intercostal muscles was related to prolonged mechanical ventilation. This result shows both of the diaphragm and intercostal muscles are important for critically ill patients, but different factors may influence the atrophy. The extent of disuse may be different because diaphragm and rib cage muscles contribute to respiration differently [9]. In spontaneous breathing patients, mechanical ventilation can support some part of diaphragm function and may disuse intercostal muscles. On the other hand, intercostal muscles may exceed the diaphragm function in increased respiratory workload [3]. Therefore, the relationship was limited to a poor association between the diaphragm and the intercostal muscles.
Ventilator settings caused contradictory results in diaphragm and intercostal muscle thickness. We did not find a significant difference in the inspiratory pressure between the patients with changed and preserved diaphragm thickness. The limited range of set inspiratory pressure (IQR, 12–14 cmH2O) may explain our results, since a previous study compared patients with a set inspiratory pressure of ≥ 12 cmH2O to those with a set inspiratory pressure of 5–12 cmH2O [20]. Conversely, high inspiratory pressure was associated with both the decreased and increased thickness of intercostal muscles. Changes in thickness were easier to monitor for intercostal muscles than the diaphragm because of the bigger muscle size (2.1 mm for diaphragm vs. 4.2 mm for intercostal muscles). Therefore, the impact of pressure was likely detected in the intercostal muscles. As excessive inspiratory support causes atrophy of the disused muscle, excessive pressure might cause injury to respiratory muscles and structural changes with gradual increased thickness. Since set inspiratory pressure does not reflect the actual pressure loaded to respiratory muscles, transpulmonary pressure should be measured to elucidate its precise influence. Similarly, the significance of the observed association between low body mass index (BMI) and intercostal muscle atrophy is unclear. It is possible that this result reflects the association of low BMI with malnutrition and increased respiratory workload. Further elucidation of this aspect is limited due to the small study sample size that was not powered to detect the significant differences.
Changes in respiratory muscle thickness occur frequently in critically ill patients. As reported for the diaphragm, intercostal muscles exhibit significant changes in these patients, and the underlying causes are multifactorial. The severity of the clinical status, medications, and ventilatory settings should be further investigated as potential factors. Compared with immobilized limb muscles, respiratory muscles are permanently activated by the ventilator [15]. Therefore, the duration and settings of mechanical ventilation may be the key factors to preserve respiratory muscles. Whether prolonged mechanical ventilation led to respiratory muscle atrophy or respiratory muscle atrophy induced prolonged mechanical ventilation remains to be elucidated. Weaning patients from mechanical ventilation may be essential to prevent or minimize respiratory muscle changes. It is suggested that maintaining and improving respiratory muscles thickness is essential in order to do fast-track extubation in the ICU. Further studies are needed to determine approaches to maintain respiratory muscles. Taken together, the current findings emphasize that it is essential to monitor changes in diaphragm and intercostal muscles.
Limitations
The current study has several limitations. First, although the patients were enrolled from two ICUs, this cohort study was limited to one country. Second, ultrasound evaluation of intercostal muscles is not established well, while reproducibility of measurement was good. Third, investigators performing measurements were not blinded to patients’ condition, although image analyses were blinded. Fourth, we classified decreased thickness first, and then, we classified increased thickness. Therefore, this classification is not fair to increased thickness. Finally, changes in intercostal muscle thickness may depend on the specific intercostal muscle that was measured.
Conclusion
This two-center prospective observational study evaluating changes in diaphragm and intercostal muscle thickness in mechanically ventilated patients measured with ultrasound revealed that decreased diaphragm and intercostal muscle thickness were frequently seen in patients under mechanical ventilation. The decreased diaphragm and intercostal muscle thickness were associated with prolonged mechanical ventilation and length of ICU stay.
Notes
Acknowledgements
The authors thank Yoshihiro Okayama of the Clinical Trial Center for Development Therapeutics of the Tokushima University Hospital for his statistical support.
Authors’ contributions
NN was involved in study design, analysis, and interpretation of the data, and drafting of the manuscript. OJ took part in study design, analysis, and drafting of the manuscript. UY was involved in acquisition of the data. NE and IT did analyze and interpreted the data. NM took part in study concept, interpretation of the data, and critical revision of the manuscript. All authors read and approved the final manuscript.
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Ethics approval and consent to participate
Ethics approval was obtained from the clinical research ethics committee at Tokushima University Hospital (approval number 2593) and Tokushima Prefectural Central Hospital (approval number 1739). Informed consent to participate in the study was also obtained from patients or from an authorized surrogate.
Consent for publication
Not applicable
Competing interests
The authors declare that they have no competing interests
Supplementary material
References
- 1.Goligher EC, Dres M, Fan E, Rubenfeld GD, Scales DC, Herridge MS, Vorona S, Sklar MC, Rittayamai N, Lanys A, et al. Mechanical ventilation-induced diaphragm atrophy strongly impacts clinical outcomes. Am J Respir Crit Care Med. 2018;197(2):204–13.CrossRefGoogle Scholar
- 2.Capdevila X, Lopez S, Bernard N, Rabischong E, Ramonatxo M, Martinazzo G, Prefaut C. Effects of controlled mechanical ventilation on respiratory muscle contractile properties in rabbits. Intensive Care Med. 2003;29(1):103–10.CrossRefGoogle Scholar
- 3.Hershenson MB, Kikuchi Y, Tzelepis GE, McCool FD. Preferential fatigue of the rib cage muscles during inspiratory resistive loaded ventilation. J Appl Physiol. 1989;66(2):750–4.CrossRefGoogle Scholar
- 4.Ratnovsky A, Elad D, Halpern P. Mechanics of respiratory muscles. Respir Physiol Neurobiol. 2008;163(1-3):82–9.CrossRefGoogle Scholar
- 5.Goligher EC, Fan E, Herridge MS, Murray A, Vorona S, Brace D, Rittayamai N, Lanys A, Tomlinson G, Singh JM, et al. Evolution of diaphragm thickness during mechanical ventilation. Impact of inspiratory effort. Am J Respir Crit Care Med. 2015;192(9):1080–8.CrossRefGoogle Scholar
- 6.Bunnell A, Ney J, Gellhorn A, Hough CL. Quantitative neuromuscular ultrasound in intensive care unit-acquired weakness: a systematic review. Muscle Nerve. 2015;52(5):701–8.CrossRefGoogle Scholar
- 7.Umbrello M, Formenti P, Longhi D, Galimberti A, Piva I, Pezzi A, Mistraletti G, Marini JJ, Iapichino G. Diaphragm ultrasound as indicator of respiratory effort in critically ill patients undergoing assisted mechanical ventilation: a pilot clinical study. Crit Care. 2015;19:161.CrossRefGoogle Scholar
- 8.Schepens T, Verbrugghe W, Dams K, Corthouts B, Parizel PM, Jorens PG. The course of diaphragm atrophy in ventilated patients assessed with ultrasound: a longitudinal cohort study. Crit Care. 2015;19:422.CrossRefGoogle Scholar
- 9.Ratnovsky A, Elad D. Anatomical model of the human trunk for analysis of respiratory muscles mechanics. Respir Physiol Neurobiol. 2005;148(3):245–62.CrossRefGoogle Scholar
- 10.Costa D, Vitti M, de Oliveira TD, Costa RP. Participation of the sternocleidomastoid muscle on deep inspiration in man. An electromyographic study. Electromyogr Clin Neurophysiol. 1994;34(5):315–20.PubMedGoogle Scholar
- 11.Breslin EH, Garoutte BC, Kohlman-Carrieri V, Celli BR. Correlations between dyspnea, diaphragm and sternomastoid recruitment during inspiratory resistance breathing in normal subjects. Chest. 1990;98(2):298–302.CrossRefGoogle Scholar
- 12.Ju S, Lee SJ, Park MJ, Cho YJ, Jeong YY, Jeon KN, Bae K, Lee JD, Kim HC. Clinical importance of cross-sectional area of intercostal muscles in patients with chronic obstructive pulmonary disease. Clin Respir J. 2018;12(3):939–47.CrossRefGoogle Scholar
- 13.Guerri R, Gayete A, Balcells E, Ramirez-Sarmiento A, Vollmer I, Garcia-Aymerich J, Gea J, Orozco-Levi M. Mass of intercostal muscles associates with risk of multiple exacerbations in COPD. Respir Med. 2010;104(3):378–88.CrossRefGoogle Scholar
- 14.Levine S, Nguyen T, Taylor N, Friscia ME, Budak MT, Rothenberg P, Zhu J, Sachdeva R, Sonnad S, Kaiser LR, et al. Rapid disuse atrophy of diaphragm fibers in mechanically ventilated humans. N Engl J Med. 2008;358(13):1327–35.CrossRefGoogle Scholar
- 15.Nakanishi N, Oto J, Tsutsumi R, Iuchi M, Onodera M, Nishimura M. Upper and lower limb muscle atrophy in critically ill patients: an observational ultrasonography study. Intensive Care Med. 2018;44(2):263–4.CrossRefGoogle Scholar
- 16.Zambon M, Greco M, Bocchino S, Cabrini L, Beccaria PF, Zangrillo A. Assessment of diaphragmatic dysfunction in the critically ill patient with ultrasound: a systematic review. Intensive Care Med. 2017;43(1):29–38.CrossRefGoogle Scholar
- 17.Glau CL, Conlon TW, Himebauch AS, Yehya N, Weiss SL, Berg RA, Nishisaki A. Progressive diaphragm atrophy in pediatric acute respiratory failure. Pediatr Crit Care Med. 2018;19(5):406–11.CrossRefGoogle Scholar
- 18.Sacanella E, Perez-Castejon JM, Nicolas JM, Masanes F, Navarro M, Castro P, Lopez-Soto A. Functional status and quality of life 12 months after discharge from a medical ICU in healthy elderly patients: a prospective observational study. Crit Care. 2011;15(2):R105.CrossRefGoogle Scholar
- 19.Grosu HB, Lee YI, Lee J, Eden E, Eikermann M, Rose KM. Diaphragm muscle thinning in patients who are mechanically ventilated. Chest. 2012;142(6):1455–60.CrossRefGoogle Scholar
- 20.Zambon M, Beccaria P, Matsuno J, Gemma M, Frati E, Colombo S, Cabrini L, Landoni G, Zangrillo A. Mechanical ventilation and diaphragmatic atrophy in critically ill patients: an ultrasound study. Crit Care Med. 2016;44(7):1347–52.CrossRefGoogle Scholar
Copyright information
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.