Impact of a combination of quantitative indices representing uptake intensity, shape, and asymmetry in DAT SPECT using machine learning: comparison of different volume of interest settings
- 172 Downloads
Abstract
Background
We sought to assess the machine learning-based combined diagnostic accuracy of three types of quantitative indices obtained using dopamine transporter single-photon emission computed tomography (DAT SPECT)—specific binding ratio (SBR), putamen-to-caudate ratio (PCR)/fractal dimension (FD), and asymmetry index (AI)—for parkinsonian syndrome (PS). We also aimed to compare the effect of two different types of volume of interest (VOI) settings from commercially available software packages DaTQUANT (Q) and DaTView (V) on diagnostic accuracy.
Methods
Seventy-one patients with PS and 40 without PS (NPS) were enrolled. Using SPECT images obtained from these patients, three quantitative indices were calculated at two different VOI settings each. SBR-Q, PCR-Q, and AI-Q were derived using the VOI settings from DaTQUANT, whereas SBR-V, FD-V, and AI-V were derived using those from DaTView. We compared the diagnostic value of these six indices for PS. We incorporated a support vector machine (SVM) classifier for assessing the combined accuracy of the three indices (SVM-Q: combination of SBR-Q, PCR-Q, and AI-Q; SVM-V: combination of SBR-V, FD-V, and AI-V). A Mann-Whitney U test and receiver-operating characteristics (ROC) analysis were used for statistical analyses.
Results
ROC analyses demonstrated that the areas under the curve (AUC) for SBR-Q, PCR-Q, AI-Q, SBR-V, FD-V, and AI-V were 0.978, 0.837, 0.802, 0.906, 0.972, and 0.829, respectively. On comparing the corresponding quantitative indices between the two types of VOI settings, SBR-Q performed better than SBR-V (p = 0.006), whereas FD-V performed better than PCR-Q (p = 0.0003). No significant difference was observed between AI-Q and AI-V (p = 0.56). The AUCs for SVM-Q and SVM-V were 0.988 and 0.994, respectively; the two different VOI settings displayed no significant differences in terms of diagnostic accuracy (p = 0.48).
Conclusion
The combination of the three indices obtained using the SVM classifier improved the diagnostic performance for PS; this performance did not differ based on the VOI settings and software used.
Keywords
123I-Ioflupane 123I-FP-CIT DAT SPECT Parkinson’s syndrome Support vector machine Machine learningAbbreviations
- AD
Alzheimer’s disease
- AI
Asymmetry index
- AUC
The areas under the curve
- DAT
Dopamine transporter
- DLB
Dementia with Lewy bodies
- FD
Fractal dimension
- MSA
Multiple system atrophy
- NPS
Non-parkinsonian syndrome
- OSEM
Ordered-subset expectation maximization
- PCR
Putamen-to-caudate ratio
- PS
Parkinsonian syndrome
- PSP
Progressive supranuclear palsy
- RI
Radioisotope
- ROC
Receiver-operating characteristics
- SBR
Specific binding ratio
- SPECT
Single-photon emission computed tomography
- SVM
Support vector machine
- VOI
Volume of interest
Background
Two different types of volume of interest (VOI) settings have been proposed for routine clinical quantification of dopamine transporter single-photon emission computed tomography (DAT SPECT) results using VOI-based analyses. One of the two VOI settings uses a striatal VOI delineated to fit the striatum precisely. Some software packages, such as DaTQUANT (GE Healthcare, Little Chalfont, UK), apply this type of VOI setting [1]. This type of VOI setting needs registration to the normalized SPECT image because the shape and position of the striatum vary among individuals. This fixed-size VOI template can be divided to conform with the striatal structure parts, i.e., the caudate, the anterior putamen, and the posterior putamen, to enable the calculation of the putamen-to-caudate ratio (PCR) that represents the ratio between anterior and posterior striatal radioisotope (RI) uptake [1, 2]. Tossici-Bolt et al. proposed a large pentagonal prism-shaped VOI, which encompasses a wide area around the striatum (Southampton method), as an alternative setting [3]. Some software packages, such as DaTView (AZE, Tokyo, Japan), apply this type of VOI setting. This VOI can be set directly on the original image without normalization and has the merit of reducing the influence of a partial volume effect as it is large in size.
These two VOI settings can be used to evaluate three different types of quantitative indices that have complementary roles: the specific binding ratio (SBR), PCR/fractal dimension (FD), and asymmetry index (AI). The SBR index represents the amount of RI accumulation in the striatum, the PCR/FD index represents the distribution or the shape of the RI uptake, and the AI index represents the lateral difference between the right and left side RI uptake, respectively. Since the VOI from Tossici-Bolt et al. cannot be separated into anterior and posterior parts, the PCR index cannot be calculated using this VOI analysis. Therefore, another quantitative index that can replace the PCR index is required. Iwabuchi et al. proposed the FD index as a new quantitative index for assessing DAT SPECT results [4]. It can be calculated with the VOI from Tossici-Bolt et al. The FD index is used to evaluate the spatial heterogeneity or distribution of various types of medical images [5]. We consider that the PCR and FD indices have similar roles in assessing DAT SPECT as both indices represent the shape or the distribution of striatal RI accumulation.
In visual interpretations, a comprehensive evaluation of the three characteristic findings is important, i.e., the decline of striatal RI accumulation, the shape of RI accumulation, and the lateral difference in RI accumulation [6]. This is because the normal striatum shows symmetrical comma-shaped regions with high accumulation in both the caudate and putamen, while asymmetric dot-shaped patterns are observed in patients with Parkinsonian syndrome (PS). Therefore, we hypothesized that, even with quantitative assessments, a comprehensive evaluation with three characteristic indices, namely the SBR, PCR/FD, and AI, would be similarly important. Although several previous studies have demonstrated the utility of these quantitative indices individually [7, 8, 9, 10, 11] or assessed the combined accuracy of SBR and FD [4], to our knowledge, no study has reported the combined usefulness of the indices using the three features (SBR, PCR/FD, and AI).
Further, the VOI-setting style is thought to affect the diagnostic accuracy in quantitative assessment [12]. Although Shimizu et al. have compared the usefulness of the two types of VOI settings in distinguishing dementia with Lewy bodies (DLB) from Alzheimer’s disease (AD) [12]; they examined only the SBR index and did not examine other indices. Tossici-Bolt et al. and Yin et al. have demonstrated the influence of VOI settings on the SBR index itself [13, 14]. However, their studies did not assess the diagnostic accuracy for PS. To our knowledge, no study has assessed the effect of VOI settings on diagnostic accuracy for PS so far.
The purpose of this study was to assess the combined diagnostic accuracy of three types of DAT SPECT quantitative indices—SBR, PCR/FD, and AI—for PS using a support vector machine (SVM) classifier, and to compare the effect of two different types of VOI settings from commercially available software packages, i.e., DaTQUANT (Q) and DaTView (V).
Materials and methods
Patients
In this retrospective study, 183 consecutive patients who underwent DAT SPECT from February 2014 to August 2015 in our hospital were selected. Of these patients, 72 were excluded from the study because of the insufficiency of the image quality (n = 9) or clinical diagnosis (n = 63). The remaining 111 patients (median age, 68.5 years; range, 30–89 years; men/women, 62/49), including 71 patients with PS and 40 without PS (NPS), were included. Of the 71 patients with PS, 51 patients were diagnosed with PD based on the clinical diagnostic criteria of the UK Parkinson’s disease society brain bank, and the other 20 patients were diagnosed with atypical parkinsonian syndrome including clinical DLB (n = 11), progressive supranuclear palsy (PSP; n = 7), and multiple system atrophy (MSA; n = 2) on the basis of established diagnostic criteria [15, 16, 17].
The institutional review board of the hospital granted permission for this retrospective review of imaging and clinical data and waived the requirement for obtaining informed consent from the patients.
SPECT acquisition and reconstruction
All SPECT images were acquired 3 h after the injection of 123I-Ioflupane (185 MBq) using the Discovery NM 630 or Discovery NM/CT 670 (GE Healthcare, Milwaukee, WI) with a FAN beam collimator. The imaging parameters were as follows: matrix size, 128 × 128; pixel size, 4.4 mm; slice thickness, 4.4 mm; and energy window, 159 keV ± 10%. The projection data acquired for 30 min were reconstructed on a Xeleris workstation (GE Healthcare). The ordered-subset expectation maximization (OSEM) method (iterations, 3; subset, 10) and a Butterworth filter setting with a critical frequency of 0.5 and power of 10.0 were applied for DaTView software. The OSEM (iterations, 2; subset, 10) and a Butterworth filter setting with a critical frequency of 0.6 and power of 10.0 were applied for DaTQUANT. Neither attenuation correction nor scatter correction was used. These settings were determined based on the manufacturer’s recommended settings.
Quantitative indices calculated from VOI-based analyses
Two types of VOI settings using DaTQUANT and DaTView. a DaTQUANT uses separated VOIs delineated to fit to the striatum. The reference VOIs are set to the bilateral occipital lobes. b DaTView uses large pentagonal prism-shaped VOIs encompassing a wide area around the striatum. The reference VOI is set to the whole brain. Abbreviation: VOI, volume of interest
The three types of indices derived with the VOI setting of DaTQUANT were named SBR-Q, PCR-Q, and AI-Q, while the indices derived with the VOI setting of DaTView were named SBR-V, FD-V, and AI-V. Five indices (SBR-Q, PCR-Q, AI-Q, SBR-V, and AI-V), except for FD-V, can be computed automatically with DaTQUANT and DaTView software. SBR-Q is defined as the mean counts of the striatal VOI (background-subtracted) divided by the mean counts of the occipital lobe VOI, which simply represents the count ratio [1]. On the other hand, SBR-V is defined as the count concentration of the striatal VOI divided by the count concentration of the whole brain, which represents the concentration ratio [3]. Both the AI-Q and AI-V are defined as the difference between the SBR of both sides divided by the mean of the SBR of both sides. PCR-Q is defined as the mean counts of the putamen VOI divided by the mean counts of the caudate VOI [1]. The FD value can be calculated using the VOI setting from Tossici-Bolt et al. After setting 5-stage threshold values to delineate the striatum, the number of pixels obtained at each threshold was measured [4, 18, 19, 20, 21, 22]. The FD is calculated on the basis of the relationship between the threshold values and the number of pixels. For details, refer to a previous report [4].
Support vector machine classifier
We compared the diagnostic accuracy between SVM-Q and SVM-V to determine the combined diagnostic accuracy of the two types of VOI settings.
Statistical analysis
Sex and age distribution were compared using the Fisher’s exact test or the t test. The Mann-Whitney U test was used to compare eight indices (SBR-Q, PCR-Q, AI-Q, SVM-Q, SBR-V, FD-V, AI-V, and SVM-V) between the PS and NPS groups. The receiver-operating characteristics (ROC) analysis of each index was performed to evaluate the AUC. We used the DeLong method to examine the difference between the two AUCs [26]. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy of each index were calculated using the optimal cutoff values determined based on the ROC curves. Differences with p values < 0.05 were considered statistically significant.
All statistical analyses were performed using the statistical package R (version 3.2.2; available as a free download from http://www.r-project.org) and SPSS software (version 14.0J; SPSS Inc., Chicago, IL).
Results
Demographic features
Patient characteristics
ALL | PS | NPS | p value | |
---|---|---|---|---|
Number of cases | 111 | 71 | 40 | – |
Age (year, mean ± SD) | 69 ± 11.6 | 70 ± 10.2 | 66 ± 13.5 | 0.12✱ |
Men/women (N) | 62/49 | 39/32 | 23/17 | 0.84✝ |
Comparison among individual quantitative indices
Box-and-whisker plots of the SBR-Q (a), PCR-Q (b), AI-Q (c), SBR-V (d), FD-V (e), and AI-V (f). There were significant differences in all six quantitative indices between NPS and PS, as assessed using the Mann-Whitney U test. Abbreviations: AI, asymmetry index; FD, fractal dimension; NPS, non-parkinsonian syndrome; PCR, putamen-to-caudate ratio; PS, parkinsonian syndrome; SBR, specific binding ratio
ROC analysis of the three types of indices. a The AUCs for SBR-Q and SBR-V were 0.978 and 0.906, respectively. b The AUCs for PCR-Q and FD-V were 0.837 and 0.972, respectively. c The AUCs for AI-Q and AI-V were 0.802 and 0.829, respectively. Abbreviations: AI, asymmetry index; AUC, area under the curve; FD, fractal dimension; PCR, putamen-to-caudate ratio; ROC, receiver-operating characteristics; SBR, specific binding ratio
Sensitivity, specificity, PPV, NPV, and accuracy of the six indices
Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) | |
---|---|---|---|---|---|
SBR-Q | 94.4 (67/71) | 92.5 (37/40) | 95.7 (67/70) | 90.2 (37/41) | 93.7 (104/111) |
SBR-V | 78.9 (56/71) | 85.0 (34/40) | 90.3 (56/62) | 69.4 (34/49) | 81.1 (90/111) |
PCR-Q | 80.3 (57/71) | 87.5 (35/40) | 91.9 (57/62) | 71.4 (35/49) | 82.9 (92/111) |
FD-V | 90.1 (64/71) | 97.5 (39/40) | 98.5 (64/65) | 84.8 (39/46) | 92.8 (103/111) |
AI-Q | 71.8 (51/71) | 85.0 (34/40) | 89.5 (51/57) | 63.0 (34/54) | 76.6 (85/111) |
AI-V | 73.2 (52/71) | 85.0 (34/40) | 89.7 (52/58) | 64.2 (34/53) | 77.5 (86/111) |
Combined diagnostic accuracy of the three types of indices
Scatter plots of the PS (red triangles) and NPS (blue circles) cases. The blue planes are the boundary surfaces determined by the SVM classifier. The apostrophe represents the means after standardization. a The SVM-Q computed with SBR-Q, PCR-Q, and AI-Q. b The SVM-V computed with SBR-V, FD-V, and AI-V. Abbreviations: AI, asymmetry index; FD, fractal dimension; NPS, non-parkinsonian syndrome; PCR, putamen-to-caudate ratio; PS, parkinsonian syndrome; SBR, specific binding ratio; SVM, support vector machine
Box-and-whisker plots and ROC analysis of SVM-Q and SVM-V. There were significant differences between NPS and PS in both the SVM-Q (a) and SVM-V (b), as assessed using a Mann-Whitney U test. c The AUCs for the SVM-Q and SVM-V were 0.988 and 0.994, respectively. Abbreviations: AUC, area under the curve; NPS, non-parkinsonian syndrome; PS, parkinsonian syndrome; ROC, receiver-operating characteristics; SVM, support vector machine
Sensitivity, specificity, PPV, NPV, and accuracy of the SVM-Q and SVM-V
Sensitivity (%) | Specificity (%) | PPV (%) | NPV (%) | Accuracy (%) | |
---|---|---|---|---|---|
SVM-Q | 94.7 (18/19) | 100 (9/9) | 100 (18/18) | 90.0 (9/10) | 96.4 (27/28) |
SVM-V | 94.7 (18/19) | 100 (9/9) | 100 (18/18) | 90.0 (9/10) | 96.4 (27/28) |
Discussion
The combined diagnostic accuracy of the three types of indices (AUC for SVM-Q = 0.988, AUC for SVM-V = 0.994) was numerically higher than the individual accuracies of the six indices (AUC = 0.802–0.978). This result suggests that the combination of the three features is useful, even for quantitative evaluation, similar to the visual interpretation described in the SNM practice guideline for DAT imaging [6]. Although several studies have demonstrated the utility of each index individually [7, 8, 9, 10, 11], this study is the first to quantitatively evaluate the combined usefulness of these indices. We believe that our proposed method involving a combination of the three characteristic indices can easily be used in routine clinical practice because these indices can be obtained simultaneously using simple VOI settings. This method can also reduce inter-operator variability as operators will only have to set the striatal VOI once; therefore, improvements in diagnostic accuracy may be achieved.
Our results demonstrate that no significant difference was observed when comparing the combined accuracies between two different types of VOI settings (p = 0.48). However, the SBR-Q delivered better accuracy than the SBR-V (p < 0.01). The method of Tossici-Bolt et al. can be influenced by the dilatation of the ventricles or the cerebral sulcus because this method uses a large VOI for the striatum [27, 28]. This might be why the accuracy of SBR-V was lower than that of SBR-Q in our study. In addition, FD-V showed better accuracy than PCR-Q (p < 0.001). PCR-Q simply represents the ratio between the anterior and posterior striatal accumulation, while FD-V can represent a change in the image through different stages using a calculation based on the threshold settings of the five stages [4]. We speculate that FD-V might be able to capture a change of distribution in more detail than PCR-Q. These results suggest that the combined use of the indices might compensate for individual differences between the corresponding indices of the two VOI settings.
To our knowledge, this is the first study to use the SVM classifier to combine the three types of indices (SBR, PCR/FD, and AI). Although several studies have evaluated the utility of SVM analysis in assessing DAT SPECT [29, 30, 31, 32], no report has applied the three indices as features. Oliveira et al. used seven indices as features, including the SBR and PCR, as well as the length and volume of striatum uptake [29]. However, they did not include the AI index as a feature, and the seven indices used in their study cannot be obtained simultaneously using one VOI setting. Iwabuchi et al. previously assessed the combined accuracy of SBR and FD using an unsophisticated method of simply dividing SBR by FD, and did not use an SVM classifier to combine these quantitative indices, or include AI as a characteristic quantitative index [4].
This study had some limitations. First, the number of patients enrolled was relatively small. Additional studies with larger numbers of participants are needed to confirm our observations. Second, the diagnoses of PD and NPS were based on clinical observations, not pathology. Although this may have influenced our results, it is practically difficult to perform DAT SPECT and pathological examination at the same time. When autopsy is used as a reference standard, diagnostic discrepancies between DAT SPECT and autopsy findings can be noted, as the diagnosis can change with disease progression [33]. Third, we did not assess the effect of reconstruction parameters and scanner characterization. As indicated by Tossici-Bolt et al. [13], these factors would have differing effects on different VOI settings, and this matter requires further investigation.
Conclusion
A combination of the three types of indices obtained using the SVM classifier improved the diagnostic performance for PS, and no significant difference was observed between the combined diagnostic performance of indices calculated using the VOI settings from two different software packages.
Notes
Acknowledgements
The authors thank the staff of the Division of Nuclear Medicine at the Department of Radiology for their valuable support.
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Availability of data and materials
All data generated or analysed during this study are included in this published article.
Authors’ contributions
YI, TN, MK, and MJ were responsible for the design of the study; YI and YM prepared the SPECT data; YI and YY performed the statistical analysis; TO, DI, and HT supported the selection of clinical data; MH supported the analysis of image processing; and all authors read and critiqued the manuscript and approved the final version of the manuscript.
Ethics approval and consent to participate
This study was approved by the Ethics Committee of Keio University School of Medicine (registration number: 20150441). This article does not contain any studies with animals performed by any of the authors.
Consent for publication
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The institutional review board of the hospital granted permission for this retrospective review of imaging and clinical data, and waived the requirement for obtaining informed consent from the patients.
Competing interests
TN and MJ received research grants from Nihon Medi-Physics Co., Ltd. and GE Healthcare Corp. MK received a research grant from Nihon Medi-Physics Co., Ltd. The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Varrone A, Dickson JC, Tossici-Bolt L, Sera T, Asenbaum S, Booij J, et al. European multicentre database of healthy controls for [123I]FP-CIT SPECT (ENC-DAT): age-related effects, gender differences and evaluation of different methods of analysis. Eur J Nucl Med Mol Imaging. 2013;40:213–27.Google Scholar
- 2.Koch W, Radau PE, Hamann C, Tatsch K. Clinical testing of an optimized software solution for an automated, observer-independent evaluation of dopamine transporter SPECT studies. J Nucl Med. 2005;46:1109–18.PubMedGoogle Scholar
- 3.Tossici-Bolt L, Hoffmann SM, Kemp PM, Mehta RL, Fleming JS. Quantification of [123I]FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio. Eur J Nucl Med Mol Imaging. 2006;33:1491–9.CrossRefGoogle Scholar
- 4.Iwabuchi Y, Nakahara T, Kameyama M, Yamada Y, Hashimoto M, Ogata Y, et al. Quantitative evaluation of the tracer distribution in dopamine transporter SPECT for objective interpretation. Ann Nucl Med. 2018;32:363–71.CrossRefGoogle Scholar
- 5.Michallek F, Dewey M. Fractal analysis in radiological and nuclear medicine perfusion imaging: a systematic review. Eur Radiol. 2014;24:60–9.CrossRefGoogle Scholar
- 6.Djang DS, Janssen MJ, Bohnen N, Booij J, Henderson TA, Herholz K, et al. SNM practice guideline for dopamine transporter imaging with 123I-ioflupane SPECT 1.0. J Nucl Med. 2012;53:154–63.CrossRefGoogle Scholar
- 7.El Fakhri G, Habert MO, Maksud P, Kas A, Malek Z, Kijewski MF, et al. Quantitative simultaneous (99m)Tc-ECD/123I-FP-CIT SPECT in Parkinson’s disease and multiple system atrophy. Eur J Nucl Med Mol Imaging. 2006;33:87–92.CrossRefGoogle Scholar
- 8.Varrone A, Marek KL, Jennings D, Innis RB, Seibyl JP. [(123)I]beta-CIT SPECT imaging demonstrates reduced density of striatal dopamine transporters in Parkinson’s disease and multiple system atrophy. Mov Disord. 2001;16:1023–32.CrossRefGoogle Scholar
- 9.Niimi Y, Ito S, Murate K, Hirota S, Hikichi C, Ishikawa T, et al. Usefulness of combining (123)I-FP-CIT-SPECT striatal asymmetry index and cardiac (123)I-metaiodobenzylguanidine scintigraphy examinations for diagnosis of parkinsonisms. J Neurol Sci. 2017;377:174–8.CrossRefGoogle Scholar
- 10.Zijlmans J, Evans A, Fontes F, Katzenschlager R, Gacinovic S, Lees AJ, et al. [123I] FP-CIT SPECT study in vascular parkinsonism and Parkinson’s disease. Mov Disord. 2007;22:1278–85.CrossRefGoogle Scholar
- 11.Pirker W, Asenbaum S, Bencsits G, Prayer D, Gerschlager W, Deecke L, et al. [123I]beta-CIT SPECT in multiple system atrophy, progressive supranuclear palsy, and corticobasal degeneration. Mov Disord. 2000;15:1158–67.CrossRefGoogle Scholar
- 12.Shimizu S, Namioka N, Hirose D, Kanetaka H, Hirao K, Hatanaka H, et al. Comparison of diagnostic utility of semi-quantitative analysis for DAT-SPECT for distinguishing DLB from AD. J Neurol Sci. 2017;377:50–4.CrossRefGoogle Scholar
- 13.Tossici-Bolt L, Dickson JC, Sera T, Booij J, Asenbaun-Nan S, Bagnara MC, et al. [(123)I]FP-CIT ENC-DAT normal database: the impact of the reconstruction and quantification methods. EJNMMI Phys. 2017;4:8.CrossRefGoogle Scholar
- 14.Yin TK, Lee BF, Yang YK, Chiu NT. Differences of various region-of-interest methods for measuring dopamine transporter availability using 99mTc-TRODAT-1 SPECT. Sci World J. 2014;2014:837439.Google Scholar
- 15.McKeith IG, Boeve BF, Dickson DW, Halliday G, Taylor JP, Weintraub D, et al. Diagnosis and management of dementia with Lewy bodies: fourth consensus report of the DLB consortium. Neurology. 2017;89:88–100.CrossRefGoogle Scholar
- 16.Gilman S, Wenning GK, Low PA, Brooks DJ, Mathias CJ, Trojanowski JQ, et al. Second consensus statement on the diagnosis of multiple system atrophy. Neurology. 2008;71:670–6.CrossRefGoogle Scholar
- 17.Hoglinger GU, Respondek G, Stamelou M, Kurz C, Josephs KA, Lang AE, et al. Clinical diagnosis of progressive supranuclear palsy: the movement disorder society criteria. Mov Disord. 2017;32:853–64.CrossRefGoogle Scholar
- 18.Nagao M, Murase K, Ichiki T, Sakai S, Yasuhara Y, Ikezoe J. Quantitative analysis of technegas SPECT: evaluation of regional severity of emphysema. J Nucl Med. 2000;41:590–5.PubMedGoogle Scholar
- 19.Nagao M, Murase K, Kikuchi T, Ikeda M, Nebu A, Fukuhara R, et al. Fractal analysis of cerebral blood flow distribution in Alzheimer’s disease. J Nucl Med. 2001;42:1446–50.PubMedGoogle Scholar
- 20.Nagao M, Murase K, Yasuhara Y, Ikezoe J. Quantitative analysis of pulmonary emphysema: three-dimensional fractal analysis of single-photon emission computed tomography images obtained with a carbon particle radioaerosol. AJR Am J Roentgenol. 1998;171:1657–63.CrossRefGoogle Scholar
- 21.Nagao M, Sugawara Y, Ikeda M, Fukuhara R, Hokoishi K, Murase K, et al. Heterogeneity of cerebral blood flow in frontotemporal lobar degeneration and Alzheimer’s disease. Eur J Nucl Med Mol Imaging. 2004;31:162–8.CrossRefGoogle Scholar
- 22.Nagao M, Sugawara Y, Ikeda M, Fukuhara R, Ishikawa T, Murase K, et al. Heterogeneity of posterior limbic perfusion in very early Alzheimer’s disease. Neurosci Res. 2006;55:285–91.CrossRefGoogle Scholar
- 23.Vapnik V, Lerner A. Pattern recognition using generalized portrait method. Autom Remote Control. 1963;24:774–80.Google Scholar
- 24.Vapnik VN. The nature of statistical learning theory. Berlin: Springer-Verlag; 1995.Google Scholar
- 25.Kohavi R. A study of cross-validation and bootstrap for accuracy estimation and model selection. Proceedings of the 14th international joint conference on artificial intelligence - volume 2. Montreal, Quebec: Morgan Kaufmann Publishers Inc.; 1995. p. 1137–43.Google Scholar
- 26.DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics. 1988;44:837–45.CrossRefGoogle Scholar
- 27.Nonokuma M, Kuwabara Y, Hida K, Tani T, Takano K, Yoshimitsu K. Optimal ROI setting on the anatomically normalized I-123 FP-CIT images using high-resolution SPECT. Ann Nucl Med. 2016;30:637–44.CrossRefGoogle Scholar
- 28.Mizumura S, Nishikawa K, Murata A, Yoshimura K, Ishii N, Kokubo T, et al. Improvement in the measurement error of the specific binding ratio in dopamine transporter SPECT imaging due to exclusion of the cerebrospinal fluid fraction using the threshold of voxel RI count. Ann Nucl Med. 2018;32:288–96.CrossRefGoogle Scholar
- 29.Oliveira FPM, Faria DB, Costa DC, Castelo-Branco M, Tavares J. Extraction, selection and comparison of features for an effective automated computer-aided diagnosis of Parkinson’s disease based on [(123)I]FP-CIT SPECT images. Eur J Nucl Med Mol Imaging. 2017;45:1052–62.CrossRefGoogle Scholar
- 30.Illan IA, Gorrz JM, Ramirez J, Segovia F, Jimenez-Hoyuela JM, Ortega Lozano SJ. Automatic assistance to Parkinson’s disease diagnosis in DaTSCAN SPECT imaging. Med Phys. 2012;39:5971–80.CrossRefGoogle Scholar
- 31.Huertas-Fernandez I, Garcia-Gomez FJ, Garcia-Solis D, Benitez-Rivero S, Marin-Oyaga VA, Jesus S, et al. Machine learning models for the differential diagnosis of vascular parkinsonism and Parkinson’s disease using [(123)I]FP-CIT SPECT. Eur J Nucl Med Mol Imaging. 2015;42:112–9.CrossRefGoogle Scholar
- 32.Taylor JC, Fenner JW. Comparison of machine learning and semi-quantification algorithms for (I123)FP-CIT classification: the beginning of the end for semi-quantification? EJNMMI Physics. 2017;4:29.CrossRefGoogle Scholar
- 33.Grandal Leiros B, Perez Mendez LI, Zelaya Huerta MV, Moreno Eguinoa L, Garcia-Bragado F, Tunon Alvarez T, et al. Prevalence and concordance between the clinical and the post-mortem diagnosis of dementia in a psychogeriatric clinic. Neurologia. 2018;33:13–7.CrossRefGoogle Scholar
Copyright information
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.