Abstract
Accurate noninvasive quantification of volume blood flow in the uterine arteries (UtAs) would have clinical and research benefits. We evaluated the correlation and agreement between uterine artery volume blood flow (UtABF) as calculated (cUtABF)from color/pulsed-wave Doppler acquisitions and that measured (mUtABF) by bilateral perivascular transit-time flow probes in 6 pregnant sheep at 2 gestational ages. Out of 22 Doppler acquisitions, 19 were successful. The overall correlation between cUtABF and mUtABF was 0.55 (n = 19, P = .01). Calculated UtABF and mUtABF were significantly correlated in late gestation (n = 11, r = 0.71, P = .01) but not at mid-gestation (n = 8, r= .02, P = .96). By Bland-Altman analysis, the mean cUtABF/mUtABF was 1.15 with 95% limit of agreement (−0.26 to 2.56), similar to results previously achieved using power/pulsed-wave Doppler. Despite the acceptable correlation, the limits of agreement between Doppler and transit-time flow probe measurements remain wide. This makes Doppler ultrasonography less than a desirable method to quantify UtABF in studies where accurate quantification is required.
Similar content being viewed by others
References
Gill RW. Pulsed Doppler with B-mode imaging for quantitative blood flow measurement. Ultrasound Med Biol. 1979; 5(3):223–235.
Konje JC, Abrams K, Bell SC, de Chazal RC, Taylor DJ. The application of color power angiography to the longitudinal quantification of blood flow volume in the fetal middle cerebral arteries, ascending aorta, descending aorta, and renal arteries during gestation. Am J Obstet Gynecol. 2000;182(2): 393–400.
Konje JC, Howarth ES, Kaufmann P, Taylor DJ. Longitudinal quantification of uterine artery volume blood flow changes during gestation in pregnancies complicated by intrauterine growth restriction. Br J Obstet Gynaecol. 2003;110(3):301–305.
Acharya G, Sitras V, Erkinaro T, et al. Experimental validation of uterine artery volume blood flow measurement by Doppler ultrasonography in pregnant sheep. Ultrasound Obstet Gynecol. 2007;29(4):401–406.
Lundell A, Bergqvist D, Mattsson E, Nilsson B. Volume blood flow measurements with a transit time flowmeter: an in vivo and in vitro variability and validation study. Clin Physiol. 1993;13(5):547–557.
Sokol GM, Liechty EA, Boyle DW. Comparison of steady-state diffusion and transit time ultrasonic measurements of umbilical blood flow in the chronic fetal sheep preparation. Am J Obstet Gynecol. 1996;174(5):1456–1460.
Taylor KJ, Holland S. Doppler US. Part 1. Basic principles, instrumentation, and pitfalls. Radiology. 1990;174(2): 297–307.
David AL, Torondel B, Zachary I, et al. Local delivery of VEGF adenovirus to the uterine artery increases vasorelaxation and uterine blood flow in the pregnant sheep. Gene Ther. 2008;15(19):1344–1350.
Abi Nader KN, Mehta V, Torondel B, et al. The effect of local over-expression of VEGF on the uterine arteries of pregnant sheep long term. Reprod Sci. 2009;16(3 suppl): 77A.
Lang U, Baker RS, Khoury J, Clark KE. Effects of chronic reduction in uterine blood flow on fetal and placental growth in the sheep. Am J Physiol Regul Integr Comp Physiol. 2000;279(1):R53-R59.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–310.
Oktar SO, Yu¨cel C, Karaosmanoglu D, et al. Blood-flow volume quantification in internal carotid and vertebral arteries: comparison of 3 different ultrasound techniques with phase-contrast MR imaging. AJNR Am J Neuroradiol. 2006;27(2): 363–369.
Palmer SK, Zamudio S, Coffin C, Parker S, Stamm E, Moore LG. Quantitative estimation of human uterine artery blood flow and pelvic blood flow redistribution in pregnancy. Obstet Gynecol. 1992;80(6):1000–1006.
Dickerson KS, Newhouse VL, Tortoli P, Guidi G. Comparison of conventional and transverse Doppler sonograms. J Ultrasound Med. 1993;12(9):497–506.
Ho SS, Chan YL, Yeung DK, Metreweli C. Blood flow quantification of cerebral ischemia: comparison of three noninvasive imaging techniques of carotid and vertebral arteries. AJR Am J Roentgenol. 2002;178(3):551–556.
Hoskins PR. Ultrasound techniques for measurement of blood flow and tissue motion. Biorheology. 2002;39(3–4): 451–459.
Weskott HP. B-flow: a new method for detecting blood flow [in German]. Ultraschall Med. 2000;21(2):59–65.
Guiot C, Gaglioti P, Oberto M, Piccoli E, Rosato R, Todros T. Is three-dimensional power Doppler ultrasound useful in the assessment of placental perfusion in normal and growth-restricted pregnancies? Ultrasound Obstet Gynecol. 2008;31(2): 171–176.
Author information
Authors and Affiliations
Corresponding author
Additional information
The authors declare that there are no financial, intellectual, or personal interests related to the material discussed in this study.
Rights and permissions
About this article
Cite this article
Abi-Nader, K.N., Mehta, V., Wigley, V. et al. Doppler Ultrasonography for the Noninvasive Measurement of Uterine Artery Volume Blood Flow Through Gestation in the Pregnant Sheep. Reprod. Sci. 17, 13–19 (2010). https://doi.org/10.1177/1933719109344772
Published:
Issue Date:
DOI: https://doi.org/10.1177/1933719109344772