Abstract
Using computer simulation, I have determined the parameters of a multichannel analyzer of charged particles of a simple design that I have proposed having the form of a cylindrical capacitor with a discrete outer cylinder and closed ends in a wide range of simultaneously recorded energies (E max/E min = 100). When introducing an additional cylindrical electrode of small dimensions near the front end of the system, it is possible to improve the resolution by more than an order of magnitude in the low-energy region. At the same time, the energy resolution of the analyzer in all the above energy range is ρ = (4–6) × 10–3.
Similar content being viewed by others
References
T. Ya. Fishkova, Patent No. 2016122283 (2016).
F. H. Read, Rev. Sci. Instrum. 73, 1129 (2002).
T. Ya. Fishkova, Prikl. Fiz., No. 2, 101 (2015).
Author information
Authors and Affiliations
Corresponding author
Additional information
Original Russian Text © T.Ya. Fishkova, 2017, published in Pis’ma v Zhurnal Tekhnicheskoi Fiziki, 2017, Vol. 43, No. 11, pp. 26–32.
Rights and permissions
About this article
Cite this article
Fishkova, T.Y. А two-electrode multichannel analyzer of charged particles with discrete outer cylindrical and flat end electrodes. Tech. Phys. Lett. 43, 514–516 (2017). https://doi.org/10.1134/S1063785017060086