Abstract
A procedure is proposed for the determination of the vapor concentration of 2,4,6-trinitrotoluene (TNT) in air at a level of 10–16 g/cm3 in the control of objects for the presence of TNT. The procedure includes preconcentration of TNT vapors and gas chromatographic determination with a limit of TNT detection at a level of 0.08 ± 0.02 pg in a sample. The procedure was applied to the control objects in cells of typical automatic luggage locker with sampling through sampling air ducts. A measure of sample losses as the breakthrough of vapors through sampling air ducts and gas dynamic conditions for the reduction of vapor losses as the ratio of diffusion coefficient to the air duct flow rate were proposed. The time constant of the saturation of the sampling system with TNT vapors and the time constant of its cleaning were determined. It was shown that TNT vapors can be detected virtually satisfactorily near TNT-containing objects with sampling through long channels.
Similar content being viewed by others
References
Gruznov, V.M., Baldin M.N., Makas’, A.L., and Titov, B.G., J. Anal. Chem., 2011, vol. 66, no. 11, p. 1121.
Baldin, M.N. and Gruznov, V.M., J. Anal. Chem., 2013, vol. 68, no. 11, p. 1002.
Buryakov, I.A., Int. J. Ion Mobility Spectrom., 2001, vol. 4, p. 112.
Buryakov, I.A. and Kolomiets, Yu.N., J. Anal. Chem., 2003, vol. 58, no. 10, p. 944.
Nadolinnyi, V.A., Kolomiets, Yu.N., Mardezhova, G.A., Danilenko, A.M., and Pronin, V.G., RF Patent 2410678, 2011.
Gruznov, V.M. and Filonenko, V.G., Skorostnoe kontsentrirovanie i vikhrevoi otbor prob vozdukha pri obnaruzhenii sledovykh kolichestv organicheskikh veshchestv (High-Speed Preconcentration and Vortex Sampling of Air in Detection of Trace Amounts of Organic Substances), Novosibirsk: Inst. Neftegas. Geol. Geofiz. Sib. Otd. Ross. Akad. Nauk, 2011.
Gruznov, V.M., Filonenko, V.G., and Shishmarev, A.T., J. Anal. Chem., 1999, vol. 54, no. 11, p. 1001.
Yashin, Ya.I., Fiziko-khimicheskie osnovy khromatograficheskogo razdeleniya (Physicochemical Fundamentals of Chromatographic Separation), Moscow: Khimiya, 1976.
Author information
Authors and Affiliations
Corresponding author
Additional information
Original Russian Text © V.M. Gruznov, M.N. Baldin, M.V. Pryamov, E.M. Maksimov, 2017, published in Zhurnal Analiticheskoi Khimii, 2017, Vol. 72, No. 11, pp. 1000–1005.
Rights and permissions
About this article
Cite this article
Gruznov, V.M., Baldin, M.N., Pryamov, M.V. et al. Determination of explosive vapor concentrations with remote sampling in the control of objects. J Anal Chem 72, 1155–1160 (2017). https://doi.org/10.1134/S1061934817110041
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1134/S1061934817110041