Abstract
Polymer compositions based on the ionic liquid tetradecylammonium triethylammonium-closododecaborate (TTCD) are proposed as the main components of membranes of potentiometric sensors (ion-selective electrodes) for determining ions [B12H11N(C2H5)3]−. Two types of polymer compositions are considered: conventional, polyvinyl chloride (PVC)-liquid ion-exchanger (solution of TTCD in o-nitrophenyl octyl ether) and another, PVC-TTCD (ionophore-plasticizer). The optimal composition of membranes for both types of electrodes are proposed, and their main electroanalytical parameters, such as selectivity, effect of pH, range of linear response, reproducibility, and stability of potential, were measured. A comparative analysis of the electroanalytical parameters of potentiometric sensors with membranes of two types is given. The detection limits for the electrodes of types I and II are 9 × 10−7 and 4 × 10−7 M. It is shown that [B12H11N(C2H5)3]− anions can be determined by potentiometric titration with indicator electrodes of different types.
Similar content being viewed by others
References
Morf, W.E., The Principles of Ion-Selective Electrodes and of Membrane Transport, Budapest: Akademiai Kiado, 1981.
Egorov, V.V., Ros. Khim. Zh., 2008, vol. 52, no. 2, p. 37.
Shvedene, N.V., Chernyshev, D.V., and Pletnev, I.V., Ros. Khim. Zh., 2008, vol. 52, no. 2, p. 80.
Sviridov, V.V., Avramenko, O.A., Raeva, A.A., Pletnev, I.V., Baulin, V.E., and Shvedene, N.V., Vest. Mosk. Univ., Ser. 2: Khim., 2007, vol. 48, no. 4, p. 245.
Kopityin, A.V., Gaborne Klatsmanyi Piroska, Izvekov, V.P., and Pungor Erno, Magyar Kemiai Folyoirat, 1982, vol. 88, p. 122.
Kopityin, A.V., Gabor-Klatsmanyi, P., Izvekov, V.P., Pungor, E., and Yagodin, G.A., Anal. Chim. Acta, 1983, vol. 148, p. 35.
Kopityin, A.V., Gáborné Klatsmányi Piroska, Pungor Ernö, and Jagodin, G.A., Magyar Kemiai Folyoirat, 1983, vol. 89, p. 42.
Kopityin, A.V., Gabor-Klatsmanyi, P., Izvekov, V.P., Pungor, E., and Yagodin, G.A., Anal. Chim. Acta, 1984, p. 123.
Kopytin, A.V., Pyatova, E.N., Zhukov, A.F., et al., J. Anal. Chem., 2008, vol. 63, no. 9, p. 888.
Agafonov, A.V., Butman, L.A., Vinokurov, A.A., et al., Zh. Neorg. Khim., 1982, vol. 27, no. 1, p. 63.
Petranek, J. and Ryba, O., Anal. Chim. Acta, 1974, vol. 72, p. 372.
Yosyio Umezawa, Kayoko Umezawa, Naoko Hamada, Hiroshi Aoki, Jun Nakanishi, and Yukiko Nishimura, Pure Appl. Chem., 2002, vol. 74, no. 6, p. 923.
Buck, R.P. and Lindner, E., Pure Appl. Chem., 1994, vol. 66, no. 12, p. 2527.
Kuznetsov, N.T., Kulikova, L.N., and Kanaeva, O.A., Zh. Anal. Khim., 1976, vol. 31, no. 7, p. 1382.
Kulikova, L.N. and Kuznetsov, N.T., Zh. Anal. Khim., 1976, vol. 31, no. 12, p. 2447.
Author information
Authors and Affiliations
Additional information
Original Russian Text © A.V. Kopytin, K.Yu. Zhizhin, Yu.I. Urusov, V.N. Mustyatsa, Yu.V. Kokunov, N.T. Kuznetsov, 2012, published in Zhurnal Analiticheskoi Khimii, 2012, Vol. 67, No. 2, pp. 203–206.