Abstract
A technique of using a CO-3 calibration block for measuring not only the base of a piezoelectric transducer (PET) but also its incidence angle is proposed. This approach allows a single specimen to be used in an automated PET certification system for obtaining its main characteristics. The results of measuring the PET incidence angle in an automated mode on a CO-3 block are compared with the measurement results on CO-2 according to GOST (State Standard) 14782–86. It is shown that systematic error is absent. The use of a CO-3 block substantially reduces the time of obtaining the basic characteristics of a PET and simplifies the calibration procedure.
Similar content being viewed by others
References
Gurvich, A.K., Kuz’mina, L.G., and Nikolaev, S.V., Attention! Beam incidence Angle a = 70°, V Mire NK, 2006, no. 4(34), pp. 48–50.
GOST (State Standard) 14782-86, Nondestructive Testing. Welded Joints. Ultrasonic Methods.
GOST (State Standard) 26266-90, Ultrasonic Transducers. General Technical Requirements.
Bazulin, A.E., Bazulin, E.G., Vopilkin, A.Kh., and Tikhonov, D.S., AVGUR 5.4 Calibration System of Ultrasonic Piezoelectric Transducers. Part I. Description of the System and Calibratiob Techniques, Kontrol’. Diagnostika, 2009, no. 9, pp. 8–18.
Bazulin, A.E. and Bazulin, E.G., Measuring the Directivity Characteristics of Piezoelectric Transducers on a CO-3 Calibration Block, Akust. Zh. (in press).
Kachanov, E.I., Pigulevskii, E.D., and Yarichin, E.M., Metody i sredstva gidroakusticheskoi golografii (Methods and Facilities of Hydroacoustic Holography), Leningrad: Sudostroenie, 1989.
Goodman, J.W., Introduction to Fourier Optics, New York: McGraw-Hill, 1971.
GOST (State Standard) 23702-90 Ultrasonic Transducers. Testing Methods.
Author information
Authors and Affiliations
Corresponding author
Additional information
Original Russian Text © A.E. Bazulin, E.G. Bazulin, 2010, published in Defektoskopiya, 2010, Vol. 46, No. 2, pp. 56–62.
Rights and permissions
About this article
Cite this article
Bazulin, A.E., Bazulin, E.G. Measuring the incidence angle of a peizoelectric transducer using a CO-3 calibration block. Russ J Nondestruct Test 46, 119–124 (2010). https://doi.org/10.1134/S1061830910020063
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1134/S1061830910020063