Abstract
A procedure is described for choosing the optimum pulse amplitude interval for a proportional counter when the useful signal level is comparable with the natural background level. It is shown that the optimum interval for counter pulses having a Gaussian amplitude distribution function corresponds to 1.4σ. The application of the procedure is demonstrated in a test example with processing of the iron K α x-ray spectrum.
References
P. S. Antsiferov and V. G. Movshev, Z. Phys. D 21,Suppl., pp. 317–318 (1991).
P. S. Antsiferov, Zh. Tekh. Fiz. 65(5), 168 (1995) [Tech. Phys. 40, 499 (1995)].
E. Fünfert and H. Neuert, Zählröhre und Szintillationzähler: Messmethoden für die Strahlung der künstlichen und natürlich-radioaktiven Substanzen (Radiation and Scintillation Counters: Radiation Measurement Techniques for Synthetic and Naturally Radioactive Substances), 2nd ed. (G. Braun, Karlruhe, 1959; Gos. Izd. Lit. Obl. At. Nauki Tekh., Moscow, 1961).